Skip to content

Paradigm® REAL-Time Revel™ MMT-723LNAS (UDI 00643169151369)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-723LNAS PRDGM INS V2.5 SK EN LN · Class 3

All models of Paradigm REAL-Time Revel (180)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This MiniMed Paradigm REAL-Time Revel insulin pump delivers insulin to people with diabetes. Its 461 reports mostly describe programming difficulties, incorrect messages on the screen, mechanical and power problems, and high blood sugar readings. The reports include 4 deaths and 130 injuries. There is 1 open recall, started in February 2026, over the pump delivering too much or too little insulin when it is raised or lowered relative to the infusion site.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
461
Deaths reported
4
Injuries reported
130
Recalls naming it
1

Summary

The FDA's MAUDE database contains 461 adverse event reports that list this exact device (DI 00643169151369), received between May 22, 2017 and May 20, 2024 — classified as 4 deaths, 130 injuries, 327 malfunctions.

The most frequently reported specific device problem is “Device Difficult to Program or Calibrate” (90 reports).

1 FDA recall record name this device's identifier, 1 still open.

13 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 13, 2026Class IIOpen, Classified

    All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

    Z-1750-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Difficult to Program or Calibrate90
  • Device Displays Incorrect Message90
  • Patient Device Interaction Problem65
  • Mechanical Problem60
  • Power Problem42
  • Adverse Event Without Identified Device or Use Problem38
  • Obstruction of Flow36
  • No Display/Image33

Patient problems

  • No Consequences Or Impact To Patient299
  • Hyperglycemia130
  • Hypoglycemia33
  • No Known Impact Or Consequence To Patient6
  • Death3
  • Elevated Ketones/ Diabetic Ketoacidosis3
  • Blood Loss1
  • Cardiac Arrest1

Recent adverse event reports

The 25 most recent of 461 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2024Break; Break

    IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED DAMAGE (PHYSICAL OR COSMETIC). THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-723LNAS. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-723LNAS WAS RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19351477 ↗
  • MalfunctionJan 30, 2024Battery Problem; Battery Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CUSTOMER REPORTS A HALF BATTERY LIFE AND DEAD BATTERY WITHOUT ALARM AND A HIGH BG OF 217 MG/DL. TROUBLESHOOTING WAS PERFORMED AND NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. ADVISED CUSTOMER TO DISCONTINUE PUMP AND REVERT BACK TO HCP¿S INSTRUCTIONS. THE INSULIN PUMP WILL BE REPLACED. THE INSULIN PUMP WAS RETURNED FOR PRODUCT ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 18604513 ↗
  • MalfunctionSep 29, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT BATTERY COMPARTMENT WAS CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10604107 ↗
  • MalfunctionAug 6, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR ALARM. THE TIME ON THE INSULIN PUMP ADVANCED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10371619 ↗
  • MalfunctionAug 4, 2020Mechanical Problem; Device Difficult to Program or Calibrate; No Apparent Adverse Event; Mechanical Problem; Device Difficult to Program or Calibrate; No Apparent Adverse Event

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BUTTON ERROR AND MOTOR ERROR ALARM. CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS CORRECT. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM. CUSTOMER WAS ABLE TO COMPLETE PUMP REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10360493 ↗
  • MalfunctionJul 31, 2020Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTONS GET STUCK. THE INSULIN PUMP HAD CRACKS IN INSULIN PUMP CASING NEAR BATTERY CAP, INSULIN PUMP WINDOW SCRATCHED AND RANDOM NO DELIVERY, REJECTED BATTERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10348889 ↗
  • MalfunctionJul 28, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Use of Device Problem; Moisture or Humidity Problem; Power Problem; Patient Device Interaction Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Use of Device Problem; Moisture or Humidity Problem; Power Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BUTTONS WERE UNRESPONSIVE WHEN AROUND MOISTURE. CUSTOMER STATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARMS. INSULIN PUMP WAS LOOSING ALL SETTING WHEN BATTERY CHANGED. INSULIN PUMP HAD FAILED BATTERY TEST. INSULIN PUMP HAD HAIRLINE CRACKS ON BATTERY PORT AND RESERVOIR PORT. CUSTOMER STATED THAT INSULIN PUMP WOULD NOT ALLOW THEM TO ADJUST BOLUS SETTINGS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10331132 ↗
  • MalfunctionJul 15, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ERROR ALARM OCCURRED. CUSTOMER REPORTED THAT THE DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER REPORTED THAT THE INSULIN PUMP WAS EXPOSED TO MOISTURE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10276030 ↗
  • MalfunctionJul 15, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10275050 ↗
  • MalfunctionJul 15, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A MOTOR ERROR ALARM FOUR TIMES. THE CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARM AND WERE ABLE TO COMPLETE THE REWIND PROCESS. CUSTOMER REPORTED THE DRIVE SUPPORT CAP APPEARED NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 10274750 ↗
  • MalfunctionJul 15, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR COMPARTMENT WAS CRACKED AND HAD A PIECE MISSING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10274686 ↗
  • MalfunctionJul 1, 2020Device Difficult to Program or Calibrate; Obstruction of Flow; Device Difficult to Program or Calibrate; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED A MOTOR ERROR AND BUTTON ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10216786 ↗
  • MalfunctionJun 26, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BUTTON ERROR ALARM. THE TIME ON THE INSULIN PUMP ADVANCED. THE CUSTOMER ALSO REPORTED THAT THEY EXPERIENCED LOW BLOOD GLUCOSE LEVEL AND WAS SWEATING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 10198962 ↗
  • MalfunctionJun 23, 2020Failure to Cycle; Failure to Cycle

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10183350 ↗
  • MalfunctionJun 19, 2020Obstruction of Flow; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10173096 ↗
  • MalfunctionJun 5, 2020Detachment of Device or Device Component; Patient Device Interaction Problem; Detachment of Device or Device Component; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER ALSO STATED THAT THE BOTTOM OF INSULIN PUMP FELL OFF. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10122166 ↗
  • MalfunctionJun 5, 2020Detachment of Device or Device Component; Patient Device Interaction Problem; Detachment of Device or Device Component; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. THE CUSTOMER STATED THAT THE INSULIN PUMP HAD FALLEN OFF AND BROKEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10122042 ↗
  • MalfunctionMay 27, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10091278 ↗
  • MalfunctionMay 12, 2020Defective Device

    THIS IS A CONVERTED RECORD. NO EVENT NARRATIVE WAS CREATED. PUMP HAD INTERMITTENT BUTTONS RESPONSE DUE TO FLATTENED (CREASED) ESC BUTTON DOME SWITCH.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10047845 ↗
  • MalfunctionMay 7, 2020Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTONS OF THE INSULIN PUMP WERE STICKY AND HARD TO PRESS. THE CUSTOMER STATED THAT THE INSULIN PUMP HAD UNEXPLAINED NO DELIVERY ALARMS AND REJECTED A NEW BATTERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10031909 ↗
  • MalfunctionApr 22, 2020Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ERROR ALARM AND NO DELIVERY ALARM. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS NORMAL. THE CUSTOMER WAS NOT ABLE TO REWIND THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9985170 ↗
  • MalfunctionApr 8, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9936428 ↗
  • MalfunctionApr 6, 2020Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem; Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD SCRATCHES ON THE DISPLAY SCREEN AND PIECES ON THE RESERVOIR RING WAS BROKEN OFF. CUSTOMER STATED RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. CUSTOMER STATED INSULIN PUMP HAD PROBLEM WITH ESC BUTTON AND THE DISPLAY SCREEN WAS FOGGY AND SCREEN WAS HARD TO READ. CUSTOMER STATED TIME CLOCK WAS ADVANCED BY ONE MINUTE. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9927653 ↗
  • MalfunctionMar 27, 2020Mechanical Problem; Patient Device Interaction Problem; Mechanical Problem; Patient Device Interaction Problem

    CUSTOMER REPORTED VIA PHONE CALL THE INSULIN PUMP ALARMED MOTOR ERROR. CUSTOMER SAID BLOOD GLUCOSE AT THE TIME OF INCIDENT 180 MG/DL. TROUBLESHOOT WAS NOT PERFORMED FOR MOTOR ERROR. THE CUSTOMER DOES NOT HAVE THE PUMP WITH THEM TO TROUBLESHOOT. THE INSULIN PUMP WILL BE RETURNING FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9891887 ↗
  • MalfunctionMar 26, 2020Obstruction of Flow; Battery Problem; Material Integrity Problem; Obstruction of Flow; Battery Problem; Material Integrity Problem

    THE CUSTOMER REPORTED NO DELIVERY ISSUE. CUSTOMER DECLINED TO TROUBLESHOOT. CUSTOMER DID NOT PROVIDED BLOOD GLUCOSE READING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9887106 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page