Skip to content

Paradigm® REAL-Time Revel™ MMT-723NAB (UDI 00643169202351)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-723NAB PRDGM INS V2.5 BL EN · Class 3

All models of Paradigm REAL-Time Revel (180)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic MiniMed Paradigm Revel insulin pump is used to deliver insulin. Its 328 reports describe incorrect messages on the screen, trouble programming or calibrating, blank displays, and mechanical problems such as cracks and lost settings after battery changes. The reports include 4 deaths and 96 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
328
Deaths reported
4
Injuries reported
96
Recalls naming it
0

Summary

The FDA's MAUDE database contains 328 adverse event reports that list this exact device (DI 00643169202351), received between Jun 1, 2017 and Feb 12, 2026 — classified as 4 deaths, 96 injuries, 228 malfunctions.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (75 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Displays Incorrect Message75
  • Device Difficult to Program or Calibrate44
  • Patient Device Interaction Problem42
  • No Display/Image38
  • Mechanical Problem36
  • Obstruction of Flow34
  • Adverse Event Without Identified Device or Use Problem26
  • Battery Problem24

Patient problems

  • No Consequences Or Impact To Patient202
  • Hyperglycemia103
  • Hypoglycemia17
  • Elevated Ketones/ Diabetic Ketoacidosis5
  • No Known Impact Or Consequence To Patient5
  • Death4
  • Vomiting3
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 25 most recent of 328 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Material Integrity Problem

    IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER HAD DAMAGE TO THE PUMP. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-723NAB. TROUBLESHOOTING WAS PERFORMED. CUSTOMER REPORTED PHYSICAL DAMAGE (CRACK OR SCRATCH) ON THE PUMP NOT RELATED TO THE RETAINER RING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE PUMP AND REVERT TO A BACKUP PLAN PER HEALTHCARE PROFESSIONAL INSTRUCTIONS. MMT-723NAB WAS REQUESTED AND CUSTOMER RESPONSE WAS THE DEVICE RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24320489 ↗
  • MalfunctionJul 11, 2023Material Integrity Problem; Material Integrity Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BATTERY CAP WAS WORN OUT. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE THE USE OF THE DEVICE. THE DEVICE WILL RETURN FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17295706 ↗
  • MalfunctionOct 8, 2020Obstruction of Flow; Battery Problem; Obstruction of Flow; Battery Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD UNEXPLAINED NO DELIVERY ALARMS. CUSTOMER STATED INSULIN PUMP LOSES SETTING WHEN CHANGING OUT THE BATTERIES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10647014 ↗
  • MalfunctionSep 22, 2020Material Integrity Problem; Patient Device Interaction Problem; Material Integrity Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS FALLEN APART ON BATTERY SIDE. THE INSULIN PUMP HAD DAMAGE TO BATTERY CAP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10565051 ↗
  • MalfunctionJul 28, 2020Power Problem; Insufficient Device Problem Information; Power Problem; Insufficient Device Problem Information

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A CODES AND BATTERY COUNT DOWN AND LOST SETTING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10331463 ↗
  • MalfunctionJul 23, 2020Use of Device Problem; Battery Problem; Insufficient Device Problem Information; Use of Device Problem; Battery Problem; Insufficient Device Problem Information

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD ERROR MESSAGE UPON BATTERY CHANGE. CUSTOMER STATED INSULIN PUMP HAD LOSES TIME AND DATE SETTINGS AT BATTERY CHANGE, DIMINISHED BATTERY LIFE, INSULIN PUMP NOT DOWNLOADABLE FOR BLOOD GLUCOSE READINGS, DID NOT RECOGNIZE INSULIN AMOUNT IN RESERVOIR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10313197 ↗
  • MalfunctionJul 20, 2020Premature Discharge of Battery; Obstruction of Flow; Premature Discharge of Battery; Obstruction of Flow

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP RECEIVED ALARM WHICH INDICATED THAT UNEXPECTED RESTART. A COLD RESET WAS PERFORMED. INSULIN PUMP ALSO RECEIVED NO DELIVERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10295072 ↗
  • MalfunctionJul 17, 2020Device Difficult to Program or Calibrate; Power Problem; Device Difficult to Program or Calibrate; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BUTTON ANOMALY. CUSTOMER REPORTED THAT BUTTONS WERE NOT RESPONDING AND SCROLLING WITHOUT TOUCHING. INSULIN PUMP REJECTED ALL BATTERIES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10287799 ↗
  • MalfunctionJul 13, 2020Device Alarm System; Patient Device Interaction Problem; Device Alarm System; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP¿S ALARM OR VIBRATIONS LASTED LONGER THAN THE NORMAL ONE. CUSTOMER REPORTED CRACKS AT THE TOP AND RIGHT, LEFT OF THE SCREEN. THE BATTERY SIDE COMPARTMENT, PIECES WERE MISSING. NO HARM REQUIRING MEDICAL INTERVENTIONS WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10262907 ↗
  • MalfunctionJul 9, 2020No Display/Image; Moisture or Humidity Problem; No Display/Image; Moisture or Humidity Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THEIR INSULIN PUMP DISPLAY WENT BLANK IMMEDIATELY AFTER INSULIN PUMP EXPOSE TO MOISTURE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10249253 ↗
  • MalfunctionJul 8, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10243487 ↗
  • MalfunctionJun 29, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD BUTTON ERROR. THE CUSTOMER STATED THAT, THEY CAN NOT GET RID FROM THE ERROR SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10205350 ↗
  • MalfunctionJun 23, 2020Premature Discharge of Battery; No Display/Image; Premature Discharge of Battery; No Display/Image

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. CUSTOMER STATED THAT THE BLANK DISPLAY DID NOT RETURN AFTER INSERTING NEW BATTERY. CUSTOMER STATED THAT INSULIN PUMP ALARMED UNEXPECTED RESTART ERROR TEST AND THEY WERE ABLE TO CLEAR THE ALARM. CUSTOMER STATED THAT HE WAS ABLE TO COMPLETE REWIND PROCESS. CUSTOMER STATED THAT HE WAS PERFORMED SELF TEST AND TEST WAS PASS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10183119 ↗
  • MalfunctionJun 18, 2020No Display/Image; Patient Device Interaction Problem; No Display/Image; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BLANK DISPLAY. CUSTOMER STATED THE CONTACT ON THE BATTERY CAP WAS NOT MISSING OR DAMAGE OR CORRODED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10166768 ↗
  • MalfunctionJun 15, 2020No Display/Image; Mechanical Problem; Battery Problem; No Display/Image; Mechanical Problem; Battery Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP SCREEN WENT BLANK AND NEVER CAME ON AGAIN. CUSTOMER REPORTED THAT THE PLUNGER WAS ACTING UP WHEN RETRACTING,STOP AND GO, THEN IT STOPPED HALFWAY IN KEPT PUSHING BUTTON. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10152297 ↗
  • MalfunctionJun 11, 2020No Display/Image; Battery Problem; No Display/Image; Battery Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP SHUTS DOWN RANDOMLY AND BATTERY DRAINS A LOT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10142220 ↗
  • MalfunctionMay 15, 2020Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. CUSTOMER STATED THAT THE INSULIN PUMP HAD SCRATCHES ON THE SCREEN AND THE BUTTONS WERE STUCK SOMETIMES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10060962 ↗
  • MalfunctionMay 14, 2020Use of Device Problem; Use of Device Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD INSULIN PUMP ERROR ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10055395 ↗
  • InjuryApr 28, 2020No Display/Image; Patient Device Interaction Problem; No Display/Image; Patient Device Interaction Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE EXPERIENCING HIGH BLOOD GLUCOSE. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 400 MG/DL. CUSTOMER STATED THAT THE INSULIN PUMP HAD CRACKED ON BATTERY COMPARTMENT AND HAD A BLANK DISPLAY. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10003976 ↗
  • MalfunctionApr 22, 2020Premature Discharge of Battery; Device Difficult to Program or Calibrate; Premature Discharge of Battery; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD UNEXPECTED RESTART AND BUTTON ERROR. CUSTOMER STATED THAT THEY WERE UNABLE TO CLEAR THE ALARM. CUSTOMER ALSO REPORTED HAT THE TIME CLOCK WAS ADVANCING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9985149 ↗
  • MalfunctionApr 10, 2020Failure to Cycle; Failure to Cycle

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM. THE DEVICE WAS NOT BUMPED OR DROPPED PRIOR TO THE ALARM AND THE DRIVE SUPPORT CAP WAS STICKING OUT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9949047 ↗
  • MalfunctionApr 9, 2020Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Power Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Power Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A FAILED BATTERY TEST, NO DELIVERY ALARM AND KEYPAD ANOMALY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9946658 ↗
  • MalfunctionApr 9, 2020Priming Problem; Priming Problem

    INFORMATION RECEIVED BY THE MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS NOT WORKING AND INSULIN PUMP HAD PRIME FILL ANOMALY. THE CUSTOMER ALSO REPORTED THAT THEY WERE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. CUSTOMER STATES THEY DID NOT RECEIVE 2ND SERIES OF BEEPS NOR DID NUMBERS APPEAR ON THE SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9942421 ↗
  • MalfunctionApr 9, 2020Premature Discharge of Battery; Premature Discharge of Battery

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD UNEXPECTED RESTART, A COLD RESET WAS PERFORMED ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER REPORTED THEY WERE ABLE TO CLEAR THE ALARM. CUSTOMER WAS NOT ABLE TO COMPLETE REWIND. ALARM OCCURRED SHORTLY AFTER INSERTING A NEW BATTERY. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9942382 ↗
  • MalfunctionApr 7, 2020Use of Device Problem; Obstruction of Flow; Power Problem; Insufficient Device Problem Information; Use of Device Problem; Obstruction of Flow; Power Problem; Insufficient Device Problem Information

    CUSTOMER REPORTED VIA PHONE CALL FAILED BATTERY TEST, NO RESERVOIR ALARM, E/A ERROR ALARM AND NO DELIVERY ALARM ON THE INSULIN PUMP. TROUBLESHOOTING WAS NOT PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9932190 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page