Sprint Quattro Secure S MRI™ SureScan™ 6935M62 (UDI 00643169356566)Recalls & reported problems

Medtronic, Inc. · LEAD 6935M62 QUATTRO SECURE S MRI US · Class 3 · Implant

All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S MRI SureScan (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Sprint Quattro Secure S MRI lead has 14,284 adverse event reports, including 221 deaths and 7,759 injuries. Many describe the lead sensing heartbeats wrongly, either over-sensing or under-sensing, or showing high impedance or a high capture threshold. Some patients report infections, bacteria in the blood, or shocks from the lead, and some reports describe the lead being replaced. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
14,405
Deaths reported
221
Injuries reported
7,808
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14,405 adverse event reports that list this exact device (DI 00643169356566), received between Dec 9, 2016 and Sep 30, 2026 — classified as 221 deaths, 7,808 injuries, 6,375 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (5,275 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 197 reportsDec 2023: 166 reportsJan 2024: 181 reportsFeb 2024: 198 reportsMar 2024: 197 reportsApr 2024: 207 reportsMay 2024: 160 reportsJun 2024: 186 reportsJul 2024: 160 reportsAug 2024: 169 reportsSep 2024: 189 reportsOct 2024: 258 reportsNov 2024: 158 reportsDec 2024: 157 reportsJan 2025: 260 reportsFeb 2025: 132 reportsMar 2025: 180 reportsApr 2025: 144 reportsMay 2025: 190 reportsJun 2025: 186 reportsJul 2025: 254 reportsAug 2025: 190 reportsSep 2025: 166 reportsOct 2025: 158 reportsNov 2025: 209 reportsDec 2025: 198 reportsJan 2026: 162 reportsFeb 2026: 199 reportsMar 2026: 192 reportsApr 2026: 179 reportsMay 2026: 159 reportsJun 2026: 120 reportsJul 2026: 63 reportsAug 2026: 122 reportsSep 2026: 142 reportsOct 2026: 0 reports202420252026260
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing5,275
  • Adverse Event Without Identified Device or Use Problem2,940
  • High impedance1,586
  • Under-Sensing1,554
  • High Capture Threshold1,481
  • Impedance Problem1,313
  • Signal Artifact/Noise1,255
  • Fracture1,004

Patient problems

  • No Clinical Signs, Symptoms or Conditions7,641
  • Unspecified Infection2,035
  • No Known Impact Or Consequence To Patient1,904
  • Shock from Patient Lead(s)562
  • No Consequences Or Impact To Patient422
  • Bacteremia215
  • Bacterial Infection199
  • Sepsis187

Recent adverse event reports

The 25 most recent of 14,405 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Endocarditis; Unspecified Infection

    MAUDE report 50953722 ↗
  • InjurySep 30, 2026Fracture; High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951647 ↗
  • MalfunctionSep 30, 2026Over-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited T-Wave Oversensing (TWOS) resulting in misclassification of arrhythmia episode type. The RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951242 ↗
  • MalfunctionSep 30, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949700 ↗
  • MalfunctionSep 30, 2026Fracture; Over-Sensing; Impedance Problem; Unstable Impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948625 ↗
  • MalfunctionSep 30, 2026High impedance

    It was reported that the Right Ventricular (RV) lead exhibited Out of Range (OOR) high defibrillation impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948569 ↗
  • MalfunctionSep 30, 2026Under-Sensing

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) experienced a partial electrical reset where diagnostic data such as pacing percentages and rate histograms were cleared, however parameter values remained unchanged. The CRT-D remains in use. It was reported that the Right Ventricular (RV) lead undersensed multiple beats on a treated Ventricular Fibrillation (VF) episode though there was no observed impact to overall device function or performance. The RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947090 ↗
  • InjurySep 30, 2026High impedance; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941471 ↗
  • MalfunctionSep 29, 2026Signal Artifact/Noise; Over-Sensing; Impedance Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940152 ↗
  • MalfunctionSep 29, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938297 ↗
  • MalfunctionSep 29, 2026Pacing Problem; Under-Sensing

    Patient problems: Syncope/Fainting

    MAUDE report 50937608 ↗
  • InjurySep 29, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936866 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50935723 ↗
  • InjurySep 29, 2026Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Pacing Problem

    It was reported that the Right Ventricular (RV) lead exhibited high, undefined defibrillation impedances and it was suspected that the lead had a high voltage fracture. The RV lead had triggered a Lead Integrity Alert (LIA) due to High-Rate Non-Sustained episodes and an elevated Sensing Integrity Counter (SIC). It was noted that the RV lead oversensing of non-physiologic intervals/noise resulted in pacing inhibition and the misclassification of High-Rate Non-Sustained episodes. The RV lead was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935008 ↗
  • InjurySep 29, 2026Signal Artifact/Noise; Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934692 ↗
  • MalfunctionSep 29, 2026High Capture Threshold

    It was reported that the patient experienced ongoing phrenic nerve stimulation during Left Ventricular (LV) lead pacing. The LV lead and Right Ventricular (RV) lead also exhibited high thresholds. The LV lead was reprogrammed, and the leads remain in use. No further patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50933224 ↗
  • MalfunctionSep 29, 2026High impedance; Impedance Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930495 ↗
  • MalfunctionSep 28, 2026High impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926039 ↗
  • MalfunctionSep 28, 2026Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50923838 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced a suspected Cardiac Resynchronization Therapy Defibrillator (CRT-D) system infection. The CRT-D system was explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50923728 ↗
  • MalfunctionSep 28, 2026Signal Artifact/Noise; Over-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited noise, oversensing, and T-Wave Oversensing (TWOS) resulting in miscla ssification of episode type. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920155 ↗
  • InjurySep 28, 2026Signal Artifact/Noise; Pacing Problem

    Patient problems: Asystole; Dizziness; Syncope/Fainting

    MAUDE report 50916228 ↗
  • MalfunctionSep 28, 2026Low impedance

    It was reported that the Right Ventricular (RV) lead exhibited Out of Range (OOR) low pacing impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915197 ↗
  • InjurySep 28, 2026Signal Artifact/Noise; Impedance Problem

    It was reported that the Right Ventricular (RV) lead triggered warning for RV defibrillation coil out of range multiple times that was noted to be rising and showed spike in impedance and noise noted on the electrograms (EGM). The RV lead was capped and remains in the patient. A new RV lead was implanted. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50914835 ↗
  • MalfunctionSep 28, 2026Low impedance

    It was reported that an alert triggered for low pacing impedance on the Right Ventricular (RV) lead. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912838 ↗

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →