CapSureFix Novus MRI™ SureScan® 5076-52 (UDI 00643169550766)Recalls & reported problems

Medtronic, Inc. · LEAD 5076-52 MRI US · Class 3 · Implant

All models of CapSureFix Novus MRI SureScan (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic CapSureFix Novus MRI heart lead appears in 8,236 adverse event reports, including 79 deaths and 3,651 injuries. The reports mostly describe the lead sensing noise or false signals, high pacing thresholds, and leads shifting out of place, and some mention infections around the implant. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
8,280
Deaths reported
80
Injuries reported
3,660
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8,280 adverse event reports that list this exact device (DI 00643169550766), received between Aug 9, 2016 and Sep 30, 2026 — classified as 80 deaths, 3,660 injuries, 4,540 malfunctions.

The most frequently reported specific device problem is “Under-Sensing” (2,835 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 139 reportsDec 2023: 53 reportsJan 2024: 76 reportsFeb 2024: 70 reportsMar 2024: 73 reportsApr 2024: 72 reportsMay 2024: 62 reportsJun 2024: 84 reportsJul 2024: 65 reportsAug 2024: 54 reportsSep 2024: 58 reportsOct 2024: 69 reportsNov 2024: 68 reportsDec 2024: 37 reportsJan 2025: 83 reportsFeb 2025: 49 reportsMar 2025: 63 reportsApr 2025: 35 reportsMay 2025: 73 reportsJun 2025: 56 reportsJul 2025: 67 reportsAug 2025: 64 reportsSep 2025: 50 reportsOct 2025: 52 reportsNov 2025: 63 reportsDec 2025: 46 reportsJan 2026: 66 reportsFeb 2026: 76 reportsMar 2026: 88 reportsApr 2026: 75 reportsMay 2026: 82 reportsJun 2026: 63 reportsJul 2026: 34 reportsAug 2026: 63 reportsSep 2026: 47 reportsOct 2026: 0 reports202420252026139
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Under-Sensing2,835
  • Over-Sensing1,801
  • High Capture Threshold1,209
  • Adverse Event Without Identified Device or Use Problem1,093
  • Device Dislodged or Dislocated793
  • Failure to Capture602
  • High impedance587
  • Pacing Problem541

Patient problems

  • No Clinical Signs, Symptoms or Conditions3,576
  • No Known Impact Or Consequence To Patient2,429
  • Unspecified Infection691
  • No Consequences Or Impact To Patient239
  • Dizziness177
  • Chest Pain170
  • Dyspnea141
  • Cardiac Perforation126

Recent adverse event reports

The 25 most recent of 8,280 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited oversensing and a mirrored artifact possibly due to an insulation breach on the Right Ventricular (RV) lead. This resulted in frequent mode switching to DDIR. Both leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946607 ↗
  • MalfunctionSep 30, 2026Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited frequent  Far Field R-Wave (FFRW) oversensingwhich was increasing the patient's atrial burden. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943384 ↗
  • MalfunctionSep 30, 2026Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing undersensing during Atrial Tachycardia/ Atrial Fibrillation (AT/AF) stored episodes resulting in termination of AT/AF detected event(s) in progress. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942845 ↗
  • MalfunctionSep 29, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938089 ↗
  • InjurySep 29, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936928 ↗
  • MalfunctionSep 29, 2026Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing triggering false AT/AF detections. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936612 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50935768 ↗
  • MalfunctionSep 29, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934213 ↗
  • MalfunctionSep 29, 2026Under-Sensing

    It was reported that the Implantable Pulse Generator (IPG) reached its End of Service (EOS). The Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) stored episodes resulting in termination of AT/AF detected events in progress. The lead and IPG remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930812 ↗
  • MalfunctionSep 28, 2026Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited intermittent undersensing. The RV lead remains in use. No patient com plications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912153 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Medical device site infection; Sepsis

    MAUDE report 50903137 ↗
  • MalfunctionSep 25, 2026Signal Artifact/Noise; Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902900 ↗
  • MalfunctionSep 25, 2026High Capture Threshold

    It was reported that that the Right Ventricular (RV) lead exhibited high thresholds. The RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50895059 ↗
  • MalfunctionSep 25, 2026Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited under sensing tiggering device classified false termination of Atrial Tachycardia (AT)/Atrial Fibrillation (AF) event(s) at times. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50894329 ↗
  • MalfunctionSep 24, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50891978 ↗
  • MalfunctionSep 23, 2026Positioning Problem

    It was reported by the patient that their pacing leads migrated from behind the Implantable Pulse Generator (IPG) up to their clavicle. The leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50872373 ↗
  • MalfunctionSep 23, 2026High impedance

    It was reported that the Right Atrial (RA) lead exhibited high impedance. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869625 ↗
  • MalfunctionSep 23, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times and resulting in termination of AT/AF detected events in progress. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867645 ↗
  • MalfunctionSep 18, 2026Under-Sensing

    It was reported that the patient experienced spasms around the Implantable Pulse Generator (IPG). The Right Ventricular (RV) lead exhibited undersensing and the Right Atrial (RA) lead exhibited undersensing resulting in overpacing. The IPG system remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50815889 ↗
  • MalfunctionSep 17, 2026Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited Far Field R-Wave (FFRW) oversensing. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50803300 ↗
  • DeathSep 17, 2026High Capture Threshold

    It was reported that the Right Atrial (RA) lead exhibited high thresholds. The patient later presented with chest pain. Approximately five days later, the patient expired. Additional information related to the cause of death was requested and not received.

    Patient problems: Chest Pain

    MAUDE report 50801328 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported by the patient that they had an infection and their Implantable Pulse Generator (IPG) system was explanted.. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50790341 ↗
  • MalfunctionSep 16, 2026High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds leading to battery depletion of the Implantable Pulse Generator (IPG). The lead and IPG remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50787853 ↗
  • MalfunctionSep 16, 2026Under-Sensing

    It was reported that undersensing on the Right Atrial (RA) lead resulted in misclassification of possible Supra Ventricular Tachycardia (SVT) as monitored Ventricular Tachycardia (VT). The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50785015 ↗
  • InjurySep 14, 2026Signal Artifact/Noise

    It was reported that the Right Atrial (RA) lead exhibited noise. The lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50752496 ↗

Other Permanent defibrillator electrodes devices