Skip to content

Capsurefix® Novus 5076-58Recalls & reported problems

Medtronic, Inc. · LEAD 5076-58 CAPSUREFIX NOVUS US EN · Class 3 · Implant

All models of CAPSUREFIX NOVUS (17)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Capsurefix Novus lead is an implanted lead that connects a heart device to the heart. Its 4,568 adverse event reports mostly describe high electrical thresholds, high impedance, oversensing and fractured or breached leads, and some describe cardiac perforation and infection, with leads removed and replaced in several cases. The reports include 47 deaths and 3,066 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,583
Deaths reported
47
Injuries reported
3,075
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4,583 adverse event reports that list this exact device (DI 00681490124867), received between Aug 10, 2013 and Sep 30, 2026 — classified as 47 deaths, 3,075 injuries, 1,461 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (1,284 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 40 reportsJan 2024: 46 reportsFeb 2024: 47 reportsMar 2024: 34 reportsApr 2024: 51 reportsMay 2024: 36 reportsJun 2024: 34 reportsJul 2024: 40 reportsAug 2024: 28 reportsSep 2024: 40 reportsOct 2024: 56 reportsNov 2024: 45 reportsDec 2024: 24 reportsJan 2025: 47 reportsFeb 2025: 41 reportsMar 2025: 35 reportsApr 2025: 29 reportsMay 2025: 54 reportsJun 2025: 40 reportsJul 2025: 38 reportsAug 2025: 30 reportsSep 2025: 38 reportsOct 2025: 19 reportsNov 2025: 27 reportsDec 2025: 27 reportsJan 2026: 38 reportsFeb 2026: 29 reportsMar 2026: 40 reportsApr 2026: 39 reportsMay 2026: 20 reportsJun 2026: 22 reportsJul 2026: 14 reportsAug 2026: 27 reportsSep 2026: 18 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202656
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold1,284
  • High impedance1,042
  • Over-Sensing779
  • Adverse Event Without Identified Device or Use Problem753
  • Fracture702
  • Impedance Problem530
  • Failure to Capture522
  • Signal Artifact/Noise514

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,848
  • No Known Impact Or Consequence To Patient1,066
  • Unspecified Infection513
  • No Consequences Or Impact To Patient196
  • Bradycardia155
  • Dizziness152
  • Syncope/Fainting106
  • Syncope88

Recent adverse event reports

The 25 most recent of 4,583 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited undersensing resulting in inappropriate pacing. The lead remains in u se. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943950 ↗
  • InjurySep 29, 2026Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Pacing Problem; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds, high impedance, oversensing, pacing failure, noise and fracture. The lead was reprogrammed, capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50930259 ↗
  • MalfunctionSep 29, 2026Signal Artifact/Noise; Electrical /Electronic Property Problem; High impedance; Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928970 ↗
  • InjurySep 29, 2026High Capture Threshold

    It was reported that both the Right Atrial (RA) lead and the Right Ventricular (RV) lead exhibited high thresholds leading to premature battery depletion of the Implantable Pulse Generator (IPG). The Implantable Pulse Generator (IPG) system was reprogrammed VVI mode and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928921 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced infection. The Implantable Pulse Generator (IPG) and Right Ventricular (RV) lead were explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50928880 ↗
  • MalfunctionSep 26, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50908402 ↗
  • InjurySep 23, 2026Capturing Problem; Device Dislodged or Dislocated

    It was reported that on the pre-discharge device check post Implantable Pulse Generator (IPG) implant, the R waves had decreased and thresholds had increased since implant. A chest X-ray confirmed that the Implantable Pulse Generator (IPG) had moved downwards and the Right Ventricular (RV) lead dislodged as a result. The IPG and RV lead were repositioned and remain in use. It was considered that the cause was likely weak tissue in the pocket area. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866339 ↗
  • InjurySep 16, 2026Capturing Problem

    It was reported that the Right Atrial (RA) lead exhibited high and rising thresholds. It was further reported that the Right Ventricular (RV) lead exhibited rising thresholds. Both leads were explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50785831 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced an infection. The Right Atrial (RA) and Right Ventricular (RV) leads were explanted. The RV lead was replaced. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50771740 ↗
  • MalfunctionSep 11, 2026Signal Artifact/Noise; Over-Sensing; Low impedance; High Capture Threshold; Unstable Impedance

    Patient problems: Heart Failure/Congestive Heart Failure

    MAUDE report 50735222 ↗
  • InjurySep 10, 2026Capturing Problem; Impedance Problem; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited rising and high thresholds as well as rising impedance. The lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50714288 ↗
  • InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed a pocket infection. The leads and competitor Implantable Pulse Generator (IPG) were explanted and replaced with a Leadless implantable pulse generator (IPG). No further patient complications have been reported as a result of this event.

    Patient problems: Medical device site infection

    MAUDE report 50689025 ↗
  • MalfunctionSep 8, 2026High impedance; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds and elevated lead impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50668704 ↗
  • InjurySep 8, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that the Implantable Pulse Generator (IPG) system was removed for an unknown reason. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50666138 ↗
  • InjurySep 4, 2026High Capture Threshold

    It was reported that the patient experienced muscle stimulation. The Right Ventricular (RV) lead exhibited high thresholds. The lead remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50606601 ↗
  • InjurySep 2, 2026Fracture

    It was reported that the Right Ventricular (RV) lead triggered a warning for low RVtip to RVcoil lead impedance. It was also reported that the RV lead exhibited intermittent low RV Coil Impedance. The RV lead was explanted and replaced. It was also reported that the Right Ventricular (RV) pace/sense lead had been was noted to be broken. The RV pace/sense lead was capped, partially explanted and a new RV lead was implanted. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50572823 ↗
  • MalfunctionSep 1, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited under sensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times, misclassification of episodes and termination of AT/AF detected event(s) in progress. It was also reported that the Right Ventricular (RV) lead exhibited under sensing. Both, the RA and RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50539988 ↗
  • MalfunctionSep 1, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50537898 ↗
  • InjuryAug 31, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced palpitations, chest discomfort and a cardiac perforation from the Right Ventricular (RV) lead approximately eleven days post implant. The lead was explanted and replaced. No further patient complications have been reported as a result of this event.

    Patient problems: Cardiac Perforation; Discomfort; Palpitations

    MAUDE report 50525527 ↗
  • InjuryAug 28, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced an infection. Staphylococcus aureus noted. The Implantable Pulse Generator (IPG) system was explanted and replaced with a leadless IPG. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia; Bacterial Infection

    MAUDE report 50494570 ↗
  • MalfunctionAug 28, 2026Failure to Capture; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50493935 ↗
  • InjuryAug 27, 2026Break; Over-Sensing

    It was reported that the Right Atrial (RA) lead and the Right Ventricular (RV) lead exhibited insulation breach and oversensing. The leads were explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50473165 ↗
  • InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted due to infection with noted vegetation. A leadless Implantable Pulse Generator (IPG) was implanted. No further patient complications have been reported as a result of this event.

    Patient problems: Medical device site infection

    MAUDE report 50439797 ↗
  • InjuryAug 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed an infection. The Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted and replaced six days later. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50414164 ↗
  • InjuryAug 23, 2026Break; Failure to Capture

    It was reported that the Right Ventricular (RV) lead exhibited an insulation breach leading to loss of capture. The lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50397273 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P930039 on Mar 29, 1996. Submitted by Medtronic, Inc. as “Medtronic(R) Capsurefix Lead Model 4068,4067,4568”. The FDA review took 879 days. The FDA has since approved 267 supplements (changes to the device or its labeling), most recently on Jun 24, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

The FDA lists this type of device as life-sustaining or life-supporting.

FDA definition of this device type: Permanent defibrillator electrodes are used with compatible implantable cardioverter defibrillators (ICD) and cardiac rhythm therapy-defibrillators (CRT-D) to sense the cardiac rhythm and to deliver high energy therapy to the heart when tachycardia is detected. Some transvenous models can also be used for pacing the right ventricle.

Common questions about the Capsurefix® Novus 5076-58

Has the Capsurefix® Novus 5076-58 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00681490124867), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Capsurefix® Novus 5076-58?

The FDA's MAUDE database lists 4,583 adverse event reports for this device, received between Aug 10, 2013 and Sep 30, 2026. In those reports, the most common specific device problems are “High Capture Threshold” (1,284), “High impedance” (1,042) and “Over-Sensing” (779); the most common patient problems are “Unspecified Infection” (513), “Bradycardia” (155) and “Dizziness” (152). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Capsurefix® Novus 5076-58?

47 reports are classified as deaths, 3,075 as injuries and 1,461 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Capsurefix® Novus 5076-58 FDA approved?

It was approved through premarket approval (PMA) P930039 on Mar 29, 1996.

Is the Capsurefix® Novus 5076-58 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Capsurefix® Novus 5076-58, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Capsurefix® Novus, model 5076-58). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Permanent defibrillator electrodes devices

All Permanent defibrillator electrodes devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page