Paradigm® REAL-Time Revel™ MMT-523NAS (UDI 00643169446427)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-523NAS PRDGM V2.5B SK US PR · Class 3
All models of Paradigm REAL-Time Revel (180)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic MiniMed Paradigm REAL-Time Revel insulin pump has 222 adverse event reports, including 163 malfunctions and 57 injuries, and 2 deaths are listed. Many reports describe wrong messages or a blank screen, power and battery problems, and false no-delivery alarms. Some patients had high or low blood sugar, while most patient records note no consequences. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 222
- Deaths reported
- 2
- Injuries reported
- 57
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 222 adverse event reports that list this exact device (DI 00643169446427), received between Jun 6, 2017 and Sep 24, 2020 — classified as 2 deaths, 57 injuries, 163 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (44 reports).
14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Patient Device Interaction Problem44
- Device Displays Incorrect Message32
- No Display/Image24
- Adverse Event Without Identified Device or Use Problem22
- Power Problem21
- Device Difficult to Program or Calibrate20
- Break14
- Mechanical Problem14
Patient problems
- No Consequences Or Impact To Patient143
- Hyperglycemia60
- Hypoglycemia19
- Elevated Ketones/ Diabetic Ketoacidosis4
- Death2
- No Known Impact Or Consequence To Patient2
- Blood Loss1
- Fainting1
Recent adverse event reports
The 25 most recent of 222 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2020Obstruction of Flow; Battery Problem; Power Problem; Patient Device Interaction Problem; Obstruction of Flow; Battery Problem; Power Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD REJECTED NEW BATTERIES, HAD RECEIVED FALSE OR UNEXPLAINED NO DELIVERY ALARMS. THE INSULIN PUMP HAD RESERVOIR CAP MISSING, RESERVOIR TURNS OUT OF PLACE, BATTERY LIFE DIMINISHED NOT 7 DAYS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10578607 ↗MalfunctionAug 24, 2020Mechanical Problem; Battery Problem; Moisture or Humidity Problem; Mechanical Problem; Battery Problem; Moisture or Humidity Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE UPPER RIGHT HAND CORNER OF THE INSULIN PUMP HAD SOME CONDENSATION OR SOMETHING OBSCURING THE VIEW, THE BATTERY HAD DIMINISHED, AND THE INSULIN PUMP WAS SLOWER TO REWIND AND LOUDER. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10441272 ↗MalfunctionAug 10, 2020Connection Problem; Patient Device Interaction Problem; Connection Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR HAD FALLEN OUT. CUSTOMER REPORTED THAT THERE WAS PHYSICAL DAMAGE TO THE INSULIN PUMP'S RETAINER RING AND THE RESERVOIR WAS NOT ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10383849 ↗InjuryAug 7, 2020Power Problem; Power Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE EXPERIENCING HIGH BLOOD GLUCOSE LEVEL. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 600 MG/DL. CUSTOMER STATED INSULIN PUMP HAD FAILED BATTERY TEST. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10378023 ↗MalfunctionAug 4, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER STATED THA THE INSULIN PUMP HAD REWIND MULTIPLE TIMES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10362201 ↗MalfunctionJul 22, 2020Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. THE CUSTOMER STATED THAT, THE BUTTONS WERE HARDER TO PRESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10308001 ↗InjuryJul 21, 2020Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Display or Visual Feedback Problem; Device Difficult to Program or Calibrate
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED LOW BLOOD GLUCOSE. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 39 MG/DL. CUSTOMER REPORTED THAT THE INSULIN PUMP BUTTON DID NOT RESPONDING. CUSTOMER REPORTED THE TIME CLOCK DOES ADVANCE. NO ALARMS OCCURRED DURING OR AFTER THE TIME THAT THE BUTTONS WERE NOT RESPONDING. CUSTOMER DECLINED TROUBLESHOOTING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hypoglycemia
MAUDE report 10306010 ↗MalfunctionJul 8, 2020Obstruction of Flow; Connection Problem; Patient Device Interaction Problem; Obstruction of Flow; Connection Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR WAS UNABLE TO LOCK IN PLACE. CUSTOMER REPORTED RANDOM NO DELIVERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10244211 ↗MalfunctionJun 29, 2020Mechanical Problem; Material Integrity Problem; Patient Device Interaction Problem; Priming Problem; Mechanical Problem; Material Integrity Problem; Patient Device Interaction Problem; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN SQUIRTS WHEN PRIMING. CUSTOMER REPORTED THAT THE PLASTIC CHIPS OFF AND REWIND WAS LOUDER AND SCREEN WAS BLOTCHY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10205096 ↗MalfunctionJun 18, 2020Device Difficult to Program or Calibrate; Visual Prompts will not Clear; Device Difficult to Program or Calibrate; Visual Prompts will not Clear
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. CUSTOMER STATED THAT THE BUTTONS WERE STICKING AND THE INSULIN PUMP WAS FROZEN UP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10167165 ↗InjuryJun 3, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE THEY WERE HOSPITALIZED DUE TO HYPERGLYCEMIA AS WELL AS DIABETIC KETOACIDOSIS ON (B)(6) 2020. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 400 MG/DL. THE CUSTOMER WAS TREATED WITH INSULIN DRIP. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis; Hyperglycemia
MAUDE report 10117367 ↗InjuryJun 1, 2020No Display/Image; No Apparent Adverse Event; No Display/Image; No Apparent Adverse Event
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE HOSPITALIZED DUE TO DIABETIC KETOACIDOSIS ON UNKNOWN DATE WITH UNKNOWN BLOOD GLUCOSE. THE CUSTOMER ALSO STATED THAT THE INSULIN PUMP WAS NOT WORKING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Elevated Ketones/ Diabetic Ketoacidosis
MAUDE report 10104439 ↗MalfunctionJun 1, 2020No Display/Image; No Display/Image
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP STOPPED WORKING. THE CUSTOMER WAS REPORTED THAT INSULIN PUMP WAS COMPLETELY BLANK. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10104309 ↗MalfunctionMay 29, 2020No Display/Image; Patient Device Interaction Problem; No Display/Image; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY AND THERE WAS A SCREEN SHOWING UP. CUSTOMER STATED THAT THERE WAS CRACKED ON INSULIN PUMP AT UPPER LEFT HAND SIDE RIGHT ABOVE BATTERY SYMBOL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10100919 ↗MalfunctionMay 29, 2020No Display/Image; No Display/Image
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP SCREEN WAS BLANK. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10100117 ↗MalfunctionMay 19, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE ARROW DOWN BUTTON OF INSULIN PUMP WAS NOT RESPONDING. THE CUSTOMER STATED THAT TIME WAS ADVANCING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10068480 ↗MalfunctionMay 15, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT TOP OF THE RESERVOIR COMPARTMENT HAD PLASTIC CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER STATED THAT RESERVOIR ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10060327 ↗MalfunctionMay 14, 2020Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS STICKING OUT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10055047 ↗MalfunctionMay 7, 2020Moisture or Humidity Problem; Power Problem; Patient Device Interaction Problem; Moisture or Humidity Problem; Power Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP WAS BROKEN AND IT WAS EXPOSED TO MOISTURE. CUSTOMER WAS REPORTED THAT INSULIN PUMP WAS STOPPED WORKING. CUSTOMER STATED THEY SAW FLUID UNDER THE LCD. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10033069 ↗MalfunctionApr 21, 2020Device Difficult to Program or Calibrate; Battery Problem; Power Problem; Device Difficult to Program or Calibrate; Battery Problem; Power Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. CUSTOMER BLOOD GLUCOSE AT THE TIME OF INCIDENT UNKNOWN. TROUBLESHOOT WAS NOT PERFORMED. INSULIN PUMP WILL NOT BE RETURNING FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9980580 ↗InjuryMar 13, 2020Device Difficult to Program or Calibrate; Power Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Power Problem; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE AND FLU ON UNKNOWN DATE. CUSTOMER¿S BLOOD GLUCOSE VALUE WAS UNKNOWN. CUSTOMER ALSO STATED THAT THEY EXPERIENCED A LOW BLOOD GLUCOSE IN THE MORNING BUT THE VALUE WAS UNKNOWN. CUSTOMER STATED THAT HE BUTTONS WERE HARDER TO PRESS. CUSTOMER STATED THAT THE BATTERY CAP AREA WAS CHIPPED, NEW BATTERY WAS REJECTED. CUSTOMER STATED THAT THE INSULIN PUMP HAD AN UNEXPLAINED NO DELIVERY ALARM AT ONE TIME. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 9828650 ↗MalfunctionFeb 26, 2020Computer Software Problem; Device Difficult to Program or Calibrate; Battery Problem; Power Problem; Computer Software Problem; Device Difficult to Program or Calibrate; Battery Problem; Power Problem
INFORMATION RECEIVED BY THE MEDTRONIC INDICATED THAT THE BUTTON OF INSULIN PUMP WAS HARDER TO PRESS. CUSTOMER STATED THAT THE WHEN THEY GOT ERROR ALARM IT COULD LOCK THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9754435 ↗MalfunctionFeb 12, 2020Mechanical Problem; Obstruction of Flow; Battery Problem; Patient Device Interaction Problem; Mechanical Problem; Obstruction of Flow; Battery Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER REPORTED THAT INSULIN PUMP HAD SCRATCHES AND REJECTED BATTERY. CUSTOMER STATED INSULIN PUMP HAD NO DELIVERY ALERTS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9698762 ↗MalfunctionDec 6, 2019Display or Visual Feedback Problem; Obstruction of Flow; Patient Device Interaction Problem; Display or Visual Feedback Problem; Obstruction of Flow; Patient Device Interaction Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD SCRATCHES ON THE SCREEN AND HAD TO TILT AROUND TO SEE THE SCREEN. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAD NO DELIVERY ALARM AND A PIECE CHIPPED OFF AROUND THE BATTERY COMPARTMENT. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9427817 ↗MalfunctionDec 2, 2019Premature Discharge of Battery; Premature Discharge of Battery
CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD UNEXPECTED RESTART. CUSTOMER'S BLOOD GLUCOSE VALUE WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9399621 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.
- Z-1750-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
- Z-1749-2026 · Feb 13, 2026 · Open, ClassifiedParadigm REAL-Time Insulin Pump (MMT-522, MMT-722)
- Z-1256-2025 · Jan 31, 2025 · Open, ClassifiedParadigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
- Z-1593-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723K
- Z-1590-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722K
- Z-1588-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-722
- Z-1589-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522K
- Z-1594-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-723
- Z-1592-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523K
- Z-1587-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-522
- Z-1591-2020 · Jun 27, 2019 · Open, ClassifiedMedtronic MiniMed Paradigm Insulin pump, model number MMT-523
- Z-2442-2015 · Jul 22, 2015 · TerminatedGuardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
- Z-2179-2014 · Jul 21, 2014 · TerminatedMedtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Sensor, glucose, invasive devices
- Dexcom G4 Platinum Sensor 9500-27Dexcom, Inc. · Sensor, Glucose, Invasive2 recalls2 deaths99,426 reports
- Dexcom G5 Transmitter Kit STT-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 death58,932 reports
- Dexcom G5 Mobile IOS App SW10611Dexcom, Inc. · Sensor, Glucose, Invasive54,544 reports
- Dexcom G5 Transmitter 9438-06Dexcom, Inc. · Sensor, Glucose, Invasive15,360 reports
- Dexcom G5 Mobile Receiver MT22719Dexcom, Inc. · Sensor, Glucose, Invasive1 death12,594 reports
- Dexcom G4 SBT Transmitter 9438-05Dexcom, Inc. · Sensor, Glucose, Invasive1 death9,041 reports
- Dexcom G5 Mobile Receiver MT22719-BLUDexcom, Inc. · Sensor, Glucose, Invasive7,685 reports
- Dexcom G5 Mobile Receiver Kit STK-GF-001Dexcom, Inc. · Sensor, Glucose, Invasive1 recall6,345 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page