MiniMed® 530G MMT-751LNAS (UDI 00643169507784)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-751LNAS PRDGM V3.0 SK EN PR LN · Class 3
All models of MiniMed 530G (71)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This MiniMed 530G insulin pump delivers insulin for people with diabetes. Many reports describe mechanical problems, difficulty programming, and incorrect messages, and some mention high blood sugar. The records include 1,923 reports, 608 injuries and 5 deaths. There is 1 recall, open since February 2026, because insulin delivery can be too high or too low when the pump changes height relative to the infusion site.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 1,923
- Deaths reported
- 5
- Injuries reported
- 608
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1,923 adverse event reports that list this exact device (DI 00643169507784), received between May 10, 2017 and Jul 20, 2021 — classified as 5 deaths, 608 injuries, 1,310 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (367 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (P120010); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Feb 13, 2026Class IIOpen, Classified
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Z-1751-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem367
- Device Difficult to Program or Calibrate298
- Device Displays Incorrect Message291
- Patient Device Interaction Problem257
- Obstruction of Flow249
- Power Problem200
- Adverse Event Without Identified Device or Use Problem192
- No Display/Image166
Patient problems
- No Consequences Or Impact To Patient1,185
- Hyperglycemia574
- Hypoglycemia179
- Elevated Ketones/ Diabetic Ketoacidosis25
- No Known Impact Or Consequence To Patient17
- Death6
- Nausea5
- Blood Loss4
Recent adverse event reports
The 25 most recent of 1,923 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 20, 2021Break; Battery Problem; Moisture or Humidity Problem; Patient Device Interaction Problem; Break; Battery Problem; Moisture or Humidity Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD CONDENSATION INSIDE THE SCREEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUP HAD BAD READINGS AT TIMES AND FREQUENCY OF THE BATTERY WAS CHANGED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Insufficient Information
MAUDE report 12194668 ↗MalfunctionAug 12, 2020Mechanical Problem; Obstruction of Flow; Battery Problem; Patient Device Interaction Problem; Mechanical Problem; Obstruction of Flow; Battery Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM, SCRATCHES ON SCREEN AND HAD UNEXPLAINED NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10395201 ↗MalfunctionAug 10, 2020Device Difficult to Program or Calibrate; Battery Problem; Material Integrity Problem; Power Problem; Device Difficult to Program or Calibrate; Battery Problem; Material Integrity Problem; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP'S BUTTONS WERE NOT RESPONDING. CUSTOMER STATED THE INSULIN PUMP HAD BATTERY LIFE LASTING UNDER A WEEK, REJECTED NEW BATTERIES AND BATTERY CAP STRIPPED AND HARD TO SECURE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10384189 ↗MalfunctionAug 5, 2020Display or Visual Feedback Problem; Mechanical Problem; Display or Visual Feedback Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD LOST SETTING AND HAD SEIZED, RAINBOW EFFECT ON THE SCREEN. CUSTOMER STATED THAT THE REWINDING WAS LOUD AND INSULIN PUMP REJECTED NEW BATTERIES. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10367080 ↗MalfunctionAug 5, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD REJECTED NEW BATTERY, BUTTONS WERE HARDER TO PUSH, REWIND/ PRIME WAS SLOWER, AND THE INSULIN PUMP RECEIVED NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10366320 ↗MalfunctionAug 4, 2020Display or Visual Feedback Problem; Obstruction of Flow; Patient Device Interaction Problem; Display or Visual Feedback Problem; Obstruction of Flow; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP SCREEN WAS DIM AND HARD TO READ. CUSTOMER STATED THAT THE INSULIN PUMP HAD NO DELIVERY ALARM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10361972 ↗InjuryAug 4, 2020Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 600 MG/DL. CUSTOMER STATED THAT THE INFORMATION OF THEIR METER WAS NOT SENDING TO THE INSULIN PUMP AFTER PULLING UP ACCOUNT. CUSTOMER ALSO STATED THAT THEY DID BOLUS DURING CALL. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10360767 ↗MalfunctionAug 4, 2020Failure to Cycle; Priming Problem; Failure to Cycle; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM ALARM AND IT STARTED ALARMING THEN THE INSULIN WAS SQUIRTING OUT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10360466 ↗MalfunctionJul 30, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Insufficient Device Problem Information; Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Insufficient Device Problem Information
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT BUTTONS OF INSULIN PUMP HAD TO PRESS MULTIPLE TIMES. CUSTOMER STATED THAT INSULIN PUMP HAD UNSPECIFIED INSULIN PUMP ERROR ALARM AND UNEXPLAINED RANDOM NO DELIVERY ALARM. CUSTOMER STATED THAT INSULIN PUMP WAS STOPPED IN MIDDLE OF REWIND AND IT TAKES LOUDER TO REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10343206 ↗MalfunctionJul 30, 2020Premature Discharge of Battery; Premature Discharge of Battery
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD INSULIN PUMP ERROR AND WHEN CHANGING BATTERIES HAD TO RESET DATE AND TIME. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10342792 ↗MalfunctionJul 30, 2020Device Difficult to Program or Calibrate; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BUTTONS WERE HARD TO PRESS. CUSTOMER STATED THAT THE SCREEN WAS HARD TO READ. CUSTOMER STATED THAT THE TIME WAS ADVANCING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10342672 ↗MalfunctionJul 30, 2020Detachment of Device or Device Component; Patient Device Interaction Problem; Detachment of Device or Device Component; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE DRIVE SUPPORT CAP WAS DAMAGED. CUSTOMER STATED THAT THE INSULIN PUMP HAD COSMETIC DAMAGE. CUSTOMER STATED THAT THE RESERVOIR AND INSULIN PUMP DOES NOT SHOW SIGNS OF DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10342669 ↗InjuryJul 30, 2020Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH AND LOW BLOOD GLUCOSE. THE CUSTOMER¿S LOW BLOOD GLUCOSE WAS 45 MG/DL AND HIGH BLOOD GLUCOSE WAS 300 MG/DL. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 10342635 ↗MalfunctionJul 29, 2020Mechanical Problem; Priming Problem; Mechanical Problem; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A MOTOR ERROR ALARM DURING THE PRIMING PROCESS. THE CUSTOMER STATED THAT SOMETIMES THE PUMP STOPS THE PRIMING PROCESS AND THE CUSTOMER HAS TO COMPLETELY REDO THE PRIMING PROCESS. CUSTOMER DECLINED TO PROVIDE THEIR BLOOD GLUCOSE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10337486 ↗MalfunctionJul 24, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD CRACK BUT WORKING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10318601 ↗MalfunctionJul 23, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Mechanical Problem; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. CUSTOMER STATED THE INSULIN PUMP'S ACT BUTTON WAS NOT REGISTERING PROPERLY WHEN PRESSING DOWN AND HAD TO PRESS TWICE AND SLOWER TO REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10313875 ↗MalfunctionJul 22, 2020Display or Visual Feedback Problem; Overheating of Device; Display or Visual Feedback Problem; Overheating of Device
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BACKLIGHT OF INSULIN PUMP SCREEN WAS NO LONGER WORKING AND IT WAS NOT BRIGHT ENOUGH TO SAW. THE CUSTOMER STATED THAT INSULIN PUMP BECOME HOT WHILE REWINDING. CUSTOMER PERFORMED SELF TEST AND IT WAS PASSED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10307707 ↗MalfunctionJul 22, 2020Device Difficult to Program or Calibrate; Obstruction of Flow; Device Difficult to Program or Calibrate; Obstruction of Flow
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD ACT, UP AND DOWN BUTTONS ARE HARDER TO PRESSED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER STATED THAT THEY RECEIVED UNEXPLAINED NO DELIVERY ALARM. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10307038 ↗MalfunctionJul 22, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR AND INSULIN DELIVERY TEMPORARILY STOPPED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10306690 ↗MalfunctionJul 21, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP BUTTONS WERE LESS RESPONSIVE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10306016 ↗InjuryJul 21, 2020Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE IN EMERGENCY ROOM DUE TO HIGH BLOOD GLUCOSE ON UNKNOWN DATE WITH UNKNOWN BLOOD GLUCOSE LEVEL. CUSTOMER REPORTED THAT THE INSULIN PUMP HAD MOTOR ERRORS AND UNEXPLAINED NO DELIVERY ALARMS. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 10300710 ↗MalfunctionJul 17, 2020No Display/Image; Mechanical Problem; Insufficient Flow or Under Infusion; Power Problem; No Display/Image; Mechanical Problem; Insufficient Flow or Under Infusion; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS NOT WORKING AT ALL AND NOT GIVING INSULIN. ALSO HAD UNEXPLAINED NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE AND RESERVOIR WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10288299 ↗MalfunctionJul 14, 2020Mechanical Problem; Power Problem; Patient Device Interaction Problem; Mechanical Problem; Power Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. INSULIN PUMP HAD NO DELIVERY ALARM. INSULIN PUMP FAILED BATTERY TEST ALARM. CUSTOMER STATED THAT THE BATTERY CAP WAS CRACKED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10269942 ↗MalfunctionJul 14, 2020Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Priming Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY AND INSULIN WAS SQUIRTING OUT DURING PRIME. THE CUSTOMER REPORTED THAT, THE BUTTONS CAN BE STICKY AND NON RESPONSIVE. THE CUSTOMER STATED THAT, THE SCREEN WAS BLANK HOWEVER BATTERY CAP WAS DAMAGED AND ALSO INSULIN PUMP HAD UNEXPLAINED NO DELIVERY ALARMS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10269683 ↗MalfunctionJul 13, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Power Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP'S BUTTONS WERE HARDER TO PRESS. CUSTOMER REPORTED THAT THE INSULIN PUMP RECEIVED UNEXPLAINED OR RANDOM NO DELIVERY ALARMS REGULARLY AND ENDS UP CHANGING SET. CUSTOMER HAD TO REWOUND THE INSULIN PUMP PER RESERVOIR CHANGE ALSO RECEIVED FAILED BATTERY TEST AND BATTERIES WERE REJECTED. CUSTOMER DECLINED TO TROUBLESHOOTING WITH 24 HOURS HELPLINE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10262618 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010. One submission often covers many models, so check the product and lot details in each record.
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Artificial pancreas device system, threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page