Skip to content

NA 3037Recalls & reported problems

Medtronic, Inc. · PROG 3037 INTRSTIM II PATIENT 60601-11 · Class 3

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169512955), received between Nov 8, 2016 and Jun 11, 2019 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Displays Incorrect Message” (2 reports).

5 other recalls cover devices cleared under the same FDA submission (P970004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Displays Incorrect Message2
  • Communication or Transmission Problem1
  • Failure to Power Up1
  • Inappropriate/Inadequate Shock/Stimulation1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Electric Shock1
  • Incontinence1
  • Therapeutic Response, Decreased1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 11, 2019Inappropriate/Inadequate Shock/Stimulation

    I GOT AN INTERSTIM ICON MODEL 3037 SACRAL NEUROMODULATOR IMPLANTED IN (B)(6) 2016. IT NEVER SEEMED TO HELP WITH MY BLADDER PROBLEMS. IN (B)(6) 2018, IT SHOCKED ME BADLY IN MY SPINE, HIPS, AND DOWN MY RIGHT LEG. LUCKILY MY PARTNER RECOGNIZED WHAT HAD HAPPENED AND FORCEABLY SHUT THE DEVICE DOWN. IT HAS BEEN TURNED OFF SINCE. I'VE ASKED SEVERAL DRS TO SEND A REFERRAL TO UROLOGISTS TO GET THE DEVICE REMOVED, BUT THEY ARE NOT COMFORTABLE ENOUGH TO REMOVE THE DEVICE SINCE THE STUDIES ARE LACKING. THEY CANNOT TAKE OUT THE ANCHORED LEAD WIRE IN MY SACRUM, BUT THEY REMOVE THE WIRES AND BATTERY PACK INSIDE OF MY SKIN. I JUST WANT THE THING TAKEN OUT OF ME SINCE IT'S PERMANENTLY TURNED OFF. IT CONSISTS OF A LEAD WIRE IN MY SACRUM WITH A CORD GOING TO A BATTERY PACK UNDER MY SKIN IN MY BACK. THERE IS AN EXTERNAL PROGRAMMER THAT DOES ELECTRO-SHOCK THERAPY. FDA SAFETY REPORT ID# (B)(4).

    Patient problems: Electric Shock

    MAUDE report 8691719 ↗
  • InjuryApr 24, 2018Device Displays Incorrect Message; Device Displays Incorrect Message

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIMULATION AND GASTROINTESTINAL/PELVIC FLOOR. THE HCP WAS ENCOUNTERING THE POOR COMMUNICATION SCREEN AND INDICATED THAT IT TOOK 10 MINUTES TO DECREASE FROM 1.8 TO 1.4. THE ISSUE APPEARED TO BE THAT THE PATIENT WAS NOT HOLDING THE PATIENT PROGRAMMER (PP) ON THE SKIN PRESSING THE MINUS BUTTON. THE CALLER INDICATED THAT THE PATIENT WAS IN THE HOSPITAL ABOUT TO HAVE SURGERY, BUT THE TYPE OF SURGERY OR IF THE SURGERY WAS RELATED TO THE PATIENT'S THERAPY WAS NOT REPORTED. PATIENT STATUS IS UNKNOWN AT THE TIME OF THIS REPORT. THERE WERE NO FURTHER COMPLICATION REPORTED OR ANTICIPATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7453690 ↗
  • InjuryMar 26, 2018Device Displays Incorrect Message; Communication or Transmission Problem; Device Displays Incorrect Message; Communication or Transmission Problem

    ON (B)(6) 2018 (B)(4) (CON): INFORMATION WAS RECEIVED FROM A CONSUMER REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIMULATION AND GASTROINTESTINAL/PELVIC FLOOR. PATIENT REPORTED SEEING POOR COMMUNICATION WHEN THEY WENT TO USE THEIR PROGRAMMER TO CHANGE SETTINGS. THE PATIENT PROGRAMMER (PP) WOULD NOT DO TELEMETRY WITH OR WITHOUT ANTENNA. IT WAS NOTED THAT THE PATIENT USED AN ANTENNA. IT WAS REVIEWED FOR PATIENT TO CONFIRM ANTENNA WAS SECURE AND SYNC WITH INS BUT THAT DID NOT RESOLVE THE ISSUE. IT WAS REVIEWED FOR PATIENT TO UNPLUG THE ANTENNA, AND PLACE THE PP DIRECTLY OVER INS, ON SKIN AND SYNC AND THIS DID NOT RESOLVE THE ISSUE EITHER. PATIENT WAS WARM TRANSFERRED TO REPAIR DEPARTMENT. NO PATIENT SYMPTOMS WERE REPORTED. PATIENT'S MEDICAL HISTORY INCLUDED PARKINSON'S DISEASE WHICH WAS THERE PRIOR TO IMPLANT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED/ ANTICIPATED AS A RESULT OF THIS EVENT. ON (B)(6) 2018 (B)(4) (CON): PATIENT REPORTED THE SAME PROGRAMMER ISSUES AND NOTED THAT THE PP WAS REPLACED. PATIENT ALSO REPORTED THE NEED TO GET IN CONTACT WITH THE LOCAL REP BEFORE GETTING DISCHARGED FROM THE HOSPITA

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7371621 ↗
  • MalfunctionNov 8, 2016Failure to Power Up; Failure to Power Up

    A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIMULATION AND GASTROINTESTINAL/PELVIC FLOOR REPORTED A LOSS OF THERAPY. THE PATIENT STATED THAT THE PATIENT PROGRAMMER (PP) HAS NOT BEEN WORKING/NOT TURNING ON FOR ABOUT ONE WEEK AND THAT ABOUT A WEEK AGO THE PATIENT BEGAN TO HAVE ACCIDENTS AND WANTED TO USE THE PP TO ADJUST THE SETTING. REPLACEMENT OF THE PP BATTERIES RESOLVED THE PP ISSUE AND THE PATIENT STATED THAT SHE WOULD ADJUST HER STIMULATION. TROUBLESHOOTING RESOLVED THE PP ISSUE.

    Patient problems: Incontinence; Therapeutic Response, Decreased

    MAUDE report 6087031 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P970004 on Sep 29, 1997. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Therapy System For Urinary Control”. The FDA review took 242 days. The FDA has since approved 449 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗

The FDA regulates product code EZW (stimulator, electrical, implantable, for incontinence) under 21 CFR 876.5270 as a Class III device. 21 CFR 876.5270 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under P970004. One submission often covers many models, so check the product and lot details in each record.

  • Z-0372-2025 · Sep 18, 2024 · Open, Classified
    Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
  • Z-0998-2022 · Feb 28, 2022 · Terminated
    MEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA,
  • Z-1147-2015 · Jan 9, 2015 · Terminated
    The Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
  • Z-0037-2008 · Jun 25, 2007 · Terminated
    Medtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implanta
  • Z-0040-2008 · Jun 25, 2007 · Terminated
    Medtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim

Common questions about the NA 3037

Has the NA 3037 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00643169512955), based on the FDA recall file dated Nov 27, 2026. However, 5 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the NA 3037?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Nov 8, 2016 and Jun 11, 2019. In those reports, the most common specific device problems are “Device Displays Incorrect Message” (2), “Communication or Transmission Problem” (1) and “Failure to Power Up” (1); the most common patient problems are “Electric Shock” (1), “Incontinence” (1) and “Therapeutic Response, Decreased” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 3037?

None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 3037 FDA approved?

It was approved through premarket approval (PMA) P970004 on Sep 29, 1997.

Who makes the NA 3037, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model 3037). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Stimulator, electrical, implantable, for incontinence devices

All Stimulator, electrical, implantable, for incontinence devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page