Paradigm® REAL-Time Revel™ MMT-523RNAH (UDI 00643169513068)Recalls & reported problems

Medtronic Minimed, Inc. · PUMP MMT-523RNAH PRDGM 3.0 PK EN RC · Class 3

All models of Paradigm REAL-Time Revel (180)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic MiniMed Paradigm Real-Time Revel insulin pump has 206 adverse event reports, with no deaths and 83 injuries. Many reports describe cracked or damaged reservoirs, motor error alarms and blank displays, and many list high blood sugar as a patient problem. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
206
Deaths reported
0
Injuries reported
83
Recalls naming it
0

Summary

The FDA's MAUDE database contains 206 adverse event reports that list this exact device (DI 00643169513068), received between Jun 2, 2017 and Sep 17, 2020 — classified as 83 injuries, 123 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (50 reports).

14 other recalls cover devices cleared under the same FDA submission (P980022); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Patient Device Interaction Problem50
  • Adverse Event Without Identified Device or Use Problem41
  • No Apparent Adverse Event19
  • No Display/Image17
  • Device Difficult to Program or Calibrate16
  • Device Displays Incorrect Message15
  • Mechanical Problem15
  • Connection Problem11

Patient problems

  • No Consequences Or Impact To Patient99
  • Hyperglycemia81
  • Hypoglycemia25
  • Elevated Ketones/ Diabetic Ketoacidosis5
  • No Known Impact Or Consequence To Patient4
  • Blood Loss3
  • Abdominal Pain1
  • Chronic Obstructive Pulmonary Disease (COPD)1

Recent adverse event reports

The 25 most recent of 206 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 17, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD CRACK AT THE BOTTOM. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10545683 ↗
  • MalfunctionAug 4, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR LIP RING WAS DAMAGED. THE CUSTOMER STATED THAT THE RESERVOIR DID LOCK INTO PLACE. CUSTOMER STATED THAT THE INSULIN PUMP HAD COSMETIC DAMAGE. CUSTOMER STATED THAT THE RESERVOIR AND INSULIN PUMP DOES NOT SHOW SIGNS OF DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10360786 ↗
  • MalfunctionJul 28, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. THE CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARM AND ABLE TO COMPLETE REWIND PROCESS. CUSTOMER REPORTED THAT INSULIN PUMP HAD AGAIN MOTOR ERROR ALARM AT NIGHT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10331197 ↗
  • MalfunctionJul 14, 2020Use of Device Problem; Battery Problem; Use of Device Problem; Battery Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MULTIPLE INSULIN PUMP ERROR. THE CUSTOMER WAS ABLE TO CLEAR THE ALARM AND ABLE TO REWIND THE INSULIN PUMP. CUSTOMER REPORTED LOW BATTERY ALARM OR SHORT BATTERY LIFETIME. ALARM OCCURRED SHORTLY AFTER INSERTING A NEW BATTERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10274343 ↗
  • MalfunctionJun 30, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR CHAMBER RIM IS GONE AND THE O-RING CAME OUT AND RESERVOIR WAS ABLE TO LOCK IN PLACE WHEN INSERTED INTO PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10210983 ↗
  • MalfunctionJun 23, 2020No Display/Image; No Display/Image

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A BLANK DISPLAY. THE CUSTOMER WAS ALSO REPORTED THAT BATTERY COMPARTMENT WAS NEITHER DAMAGED NOR CORRODED, BATTERY CAP WAS NEITHER MISSING NOR DAMAGED, AND THE SPRING WAS NEITHER DAMAGED NOR CORRODED. THE CUSTOMER WAS REPORTED THAT THE INSULIN PUMP DISPLAY DID NOT RETURN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10182694 ↗
  • MalfunctionJun 19, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE RESERVOIR KEPT FALLING FROM COMPARTMENT AND THAT RETAINER RING WAS BROKEN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10172891 ↗
  • MalfunctionJun 17, 2020Connection Problem; Patient Device Interaction Problem; Connection Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP WAS CRACK IN RESERVOIR COMPARTMENT AND RESERVOIR WAS NOT ABLE TO LOCK IN PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10161704 ↗
  • MalfunctionJun 1, 2020Failure to Cycle; Detachment of Device or Device Component; Failure to Cycle; Detachment of Device or Device Component

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM DURING PRIME. CUSTOMER STATED THAT DRIVE SUPPORT CAP WAS PROTRUDED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10107466 ↗
  • MalfunctionMay 18, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER STATED THEY WERE ABLE TO REWIND INSULIN PUMP. CUSTOMER STATED DRIVE SUPPORT CAP APPEARS NORMAL. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 10064904 ↗
  • MalfunctionMay 1, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ERROR ALARM. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM AND ABLE TO REWIND THE INSULIN PUMP. CUSTOMER REPORTED THAT DRIVE SUPPORT CAP WAS NORMAL AND INSULIN PUMP HAD NOT BEEN EXPOSED TO HIGH MAGNETIC FIELD. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10016666 ↗
  • MalfunctionApr 14, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT ACT BUTTON WAS STICKING, INSULIN PUMP WAS SLOWER TO REWIND. CUSTOMER ALSO REPORTED THAT INSULIN PUMP HAD UNEXPECTED NO DELIVERY ALARM AND CHANGING BATTERIES FREQUENTLY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9956874 ↗
  • MalfunctionApr 13, 2020Failure to Cycle; Failure to Cycle

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. CUSTOMER STATED DRIVE SUPPORT CAP WAS STICKING OUT. CUSTOMER STATED THEY WERE UNABLE TO EXIT THE PREPARING TO PRIME LOOP. TROUBLESHOOTING WAS PERFORMED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9952168 ↗
  • MalfunctionApr 10, 2020Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR ALARM. CUSTOMER ALSO REPORTED THAT DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM BUT UNABLE TO REWIND THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9950857 ↗
  • MalfunctionApr 8, 2020Patient Device Interaction Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP NOT GIVING ALERT WHEN INSULIN WAS LOW. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9936634 ↗
  • MalfunctionApr 7, 2020Premature Discharge of Battery; Mechanical Problem; Loss of Data; Patient Device Interaction Problem; Premature Discharge of Battery; Mechanical Problem; Loss of Data; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER REPORTED THE INSULIN PUMP HAD CRACKED AT THE CORNER OF BATTERY COMPARTMENT. THERE WAS NO DAMAGED TO THE RESERVOIR LIP RING. NO HARM REQUIRING MEDICAL INTERVENTIONS WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 9931987 ↗
  • MalfunctionApr 7, 2020Premature Discharge of Battery; Battery Problem; Loss of Data; Power Problem; Premature Discharge of Battery; Battery Problem; Loss of Data; Power Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD UNEXPECTED RESTART. CUSTOMER REPORTED THAT THEY WERE ABLE TO REWIND THE INSULIN PUMP. CUSTOMER WAS ABLE TO CLEAR THE ALARM. CUSTOMER REPORTED THAT THE INSULIN PUMP DID NOT RECEIVE DURING NORMAL USE. NO HARM REQUIRING MEDICAL INTERVENTIONS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9931927 ↗
  • InjuryApr 2, 2020Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL AT THE TIME OF INCIDENT WAS 465 MG/DL AND CURRENT BLOOD GLUCOSE 191 MG/DL. THE CUSTOMER WAS TREATED FOR HIGH BLOOD GLUCOSE WITH INSULIN PUMP. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP APPEARED NORMAL. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 9915192 ↗
  • MalfunctionMar 27, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD ISSUE ON BUTTONS. CUSTOMER STATES UP AND DOWN BUTTONS HAD ISSUE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9887583 ↗
  • MalfunctionMar 20, 2020Connection Problem; Patient Device Interaction Problem; Connection Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT RESERVOIR CHAMBER AND BELT CLIP NOTCH WAS CRACK. THE CUSTOMER ALSO REPORTED THAT THE RESERVOIR WAS NOT ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9859161 ↗
  • InjuryMar 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CALLER REPORTED VIA PHONE CALL THAT A CUSTOMER HAD SYMPTOMS INDICATIVE OF DIABETIC KETOACIDOSIS WITH A BLOOD GLUCOSE LEVELS OF 343 MG/DL. THE CUSTOMER DID NOT MENTION HOW THEY TREATED THEIR BLOOD GLUCOSE LEVELS. THE CALLER REPORTED THAT THE CUSTOMER HAD ISSUES WITH THE SCREEN GOING BLANK DUE TO A BATTERY CAP ISSUE. THE CUSTOMER DID NOT RETURN THE INSULIN PUMP FOR ANALYSIS.

    Patient problems: Hyperglycemia

    MAUDE report 9846119 ↗
  • InjuryMar 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THAT CUSTOMER WERE EXPERIENCING LOW BLOOD GLUCOSE. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS 42 MG/DL, 40 MG/DL. CUSTOMER WAS USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED LOW BLOOD GLUCOSE EVENT. CUSTOMER TREATED WITH FOOD. CUSTOMER WAS NEITHER IN EMERGENCY ROOM NOR ADMITTED INTO HOSPITAL. BASED ON CUSTOMER REPORT CUSTOMER DID NOT ALLEGE INSULIN PUMP WAS OVER DELIVERING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 9841112 ↗
  • InjuryMar 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CUSTOMER REPORTED VIA PHONE CALL THE THEY WERE EXPERIENCING LOW BLOOD GLUCOSE. BLOOD GLUCOSE LEVEL WAS 44 MG/DL AT TIME OF INCIDENT. CUSTOMER HAD USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED LOW BLOOD GLUCOSE EVENT. CUSTOMER WAS NEITHER IN EMERGENCY ROOM, NOR ADMITTED INTO HOSPITAL. CUSTOMER TREATED WITH FOOD. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP APPEARS TOO NORMAL. CUSTOMER REPORTED THAT THE INSULIN DRIPPING OR SQUIRTING FROM END OF SET BEFORE CONNECTING AT THEIR SITE. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 9839172 ↗
  • MalfunctionMar 16, 2020No Display/Image; Patient Device Interaction Problem; No Display/Image; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP SCREEN WAS BLANK. THE CUSTOMER WAS REPORTED THAT THE BATTERY COMPARTMENT WAS DAMAGED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

    Patient problems: Hypoglycemia

    MAUDE report 9833835 ↗
  • MalfunctionFeb 28, 2020Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem

    INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD BUTTON HARDER TO PRESSED, BUT HAD ALWAYS BEEN HARDER TO PRESS. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. CUSTOMER STATED THAT BATTERIES DOWN TO LASTING ONLY 28 DAYS AND SCRATCHED ON THE SCREEN, BUT THEY DID READ THE SCREEN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9768111 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980022. One submission often covers many models, so check the product and lot details in each record.

  • Z-1749-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
  • Z-1750-2026 · Feb 13, 2026 · Open, Classified
    Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)
  • Z-1256-2025 · Jan 31, 2025 · Open, Classified
    Paradigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
  • Z-1588-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722
  • Z-1587-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522
  • Z-1591-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523
  • Z-1593-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723K
  • Z-1592-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-523K
  • Z-1589-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-522K
  • Z-1594-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-723
  • Z-1590-2020 · Jun 27, 2019 · Open, Classified
    Medtronic MiniMed Paradigm Insulin pump, model number MMT-722K
  • Z-2442-2015 · Jul 22, 2015 · Terminated
    Guardian Real-Time Monitor, Model No. CSS7100. Indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, in adults (18 years and older) with d
  • Z-2179-2014 · Jul 21, 2014 · Terminated
    Medtronic MiniMed Guardian Monitor, Model No. CSS7100, CSS7100K Product Usage: The Medtronic MiniMed model CSS7100 and CSS7100K Guardian REAL-Time monitors are used in combinatio
  • Z-2544-2014 · Mar 13, 2014 · Terminated
    Medtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K

Other Sensor, glucose, invasive devices

All Sensor, glucose, invasive devices →