Viva™ XT CRT-D DTBA1D4 (UDI 00643169530157)Recalls & reported problems
Medtronic, Inc. · CRT-D DTBA1D4 VIVA XT US DF4 · Class 3 · Implant
All models of Viva XT CRT-D (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic Viva XT CRT-D is an implanted heart device that delivers pacing and defibrillation shocks. The 58 reports mostly describe batteries running down sooner than expected, shocks felt by the patient, sensing or stimulation problems, and infections that led to removing the device. One death and 41 injuries are reported. Two recalls name this device: an open one from February 2021 about batteries reaching end of service earlier than expected, and one closed in December 2017 about a risk that therapy could be prevented.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 58
- Deaths reported
- 1
- Injuries reported
- 41
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 58 adverse event reports that list this exact device (DI 00643169530157), received between Dec 9, 2016 and Apr 8, 2026 — classified as 1 death, 41 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (14 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Feb 3, 2021Class IOpen, Classified
A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Z-1306-2021 · Root cause: Device Design · Recall details · FDA record ↗
- Dec 19, 2017Class ITerminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
Z-0584-2018 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem16
- Premature Discharge of Battery14
- Device Sensing Problem6
- Inappropriate/Inadequate Shock/Stimulation6
- Appropriate Device Problem Term/Code Not Available5
- Electrical /Electronic Property Problem3
- Electromagnetic Interference3
- Mechanical Problem3
Patient problems
- No Known Impact Or Consequence To Patient15
- Unspecified Infection12
- No Clinical Signs, Symptoms or Conditions11
- No Consequences Or Impact To Patient7
- Shock from Patient Lead(s)6
- Pain2
- Bacteremia1
- Cardiac Arrest1
Recent adverse event reports
The 25 most recent of 58 reports, in the reporter's own words as published by the FDA.
InjuryApr 8, 2026Data Problem
IT WAS REPORTED BY THE PATIENT THAT THEIR CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) HAS NOT BEEN WORKING ANYMORE BECAUSE OF THE BATTERY. IT WAS NOTED THAT THE CRT-D WAS INACTIVATED AND REMAINS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24816645 ↗InjuryFeb 27, 2026Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT THAT THEIR CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) ¿FAILED.¿ THE CRT-D WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24472882 ↗InjuryAug 7, 2025Signal Artifact/Noise; Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED ELECTROMAGNETIC INTERFERENCE. THE PATIENT RECEIVED A SHOCK FROM THE CRT-D WHILE IN A POOL. UPON REVIEW, NOISE CONSISTENT WITH ELECTROMAGNETIC INTERFERENCE WAS NOTED. THE POOL WATER IS IN CONTACT WITH ELECTRICITY EITHER THROUGH THE PUMPS, THE LIGHTS, OR PERHAPS POOR GROUNDING. WHEN THE PATIENT WAS IN THE WATER AND TOUCHING SOMETHING GROUNDED, THE ALTERNATING CURRENT (AC) POWER WAS SENSED BY THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND NOISE CONSISTENT WITH ELECTROMAGNETIC INTERFERENCE OCCURRED. THE CRT-D WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 22739407 ↗InjuryJun 11, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN INFECTION WITH PURULENT DISCHARGE/PUS. A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG) WAS IMPLANTED THE NEXT DAY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Purulent Discharge; Unspecified Infection
MAUDE report 22188990 ↗InjuryAug 21, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND RIGHT VENTRICULAR (RV) LEAD WERE EXPLANTED AND SUBSEQUENTLY REPLACED AND THE LEFT VENTRICULAR (LV) LEAD WAS INACTIVATED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20015280 ↗InjuryMar 15, 2024Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE PATIENT HAD CONCERNS REGARDING THE ESTIMATED REMAINING BATTERY LONGEVITY OF THEIR CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) ONCE IT REACHED ELECTIVE REPLACEMENT INDICATOR (ERI). THE PATIENT NOTED THEY WERE CONCERNED THE DEVICE WOULD NOT LAST THE EXPECTED AMOUNT OF TIME BETWEEN ERI AND END OF SERVICE (EOS). THE PHYSICIAN STATED THE BATTERY WAS "KNOWN TO DEPLETE FASTER ONCE THE DEVICE REACHED ERI" AND THE PATIENT WAS SCHEDULED FOR A REPLACEMENT. THE DEVICE WAS LATER EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18915003 ↗InjuryJan 31, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT WITH COMPLAINTS OF PELVIC PAIN AND DYSURIA. BLOOD CULTURES WERE POSITIVE FOR BACTEREMIA AND SEPTICEMIA AND ECHOCARDIOGRAM SHOWED A SMALL MOBILE VEGETATION ON THE RIGHT ATRIAL (RA) LEAD OF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM. THE ORGANISM IDENTIFIED VIA BLOOD CULTURE WAS ECHERICHIA COLI AND THE SOURCE WAS A BACTERIAL URINARY TRACT INFECTION. THE PATIENT WAS HOSPITALIZED AND TREATED WITH ANTIBIOTICS. THE PATIENT IS A PARTICIPANT IN A CLINICAL STUDY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Dysuria; Pain; Sepsis; Unspecified Infection; Urinary Tract Infection
MAUDE report 18620217 ↗InjuryMar 21, 2023Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION AND THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS. THE CRT-D WAS EXPLANTED AND REPLACED AND THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16585089 ↗MalfunctionJan 17, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS FUNCTIONING ABNORMALLY. THE SYSTEM REMAINS IN USE. THE PATIENT IS A PARTICIPANT IN A CLINICAL STUDY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16183169 ↗MalfunctionOct 25, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE HEALTHCARE PROFESSIONAL DIDN'T KNOW IF THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS WORKING APPROPRIATELY AND THEY NEEDED TO HAVE THE DEVICE INTERROGATED. THE PATIENT WAS CURRENTLY IN THE INTENSIVE CARE UNIT (ICU) AND ALARMS WERE REPORTEDLY GOING OFF, SIGNALING THAT THE PATIENT WAS IN VENTRICULAR TACHYCARDIA. THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Tachycardia
MAUDE report 15665687 ↗InjurySep 27, 2022Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED AND REPLACED DUE TO EARLY BATTERY DEPLETION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15491003 ↗InjuryApr 14, 2022Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT IT WAS SUSPECTED THE PATIENT RECEIVED A SHOCK FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT THE DEVICE CLASSIFIED AS VENTRICULAR FIBRILLATION (VF). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 14112513 ↗InjuryApr 6, 2022Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY FOR ATRIAL FIBRILLATION (AF) THAT WAS DETECTED AS VENTRICULAR FIBRIL LATION (VF) BY THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 14024441 ↗InjuryJan 19, 2022Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE HIGH VOLTAGE THERAPY FOR SUSPECTED ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT WAS DETECTED AS FAST VENTRICULAR TACHYCARDIA (VT). THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 13294369 ↗InjurySep 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A STAPH INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 12519822 ↗MalfunctionSep 21, 2021Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THERE WAS PREMATURE BATTERY DEPLETION ON THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12506966 ↗MalfunctionSep 1, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE PATIENT REPRESENTATIVE THAT THE PATIENT REPORTED CONCERNS THAT THEIR CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) IS NOT WORKING PROPERLY. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12408162 ↗InjuryJan 29, 2021Failure to Charge; Failure to Charge
IT WAS REPORTED THAT THE PATIENT HAD VENTRICULAR TACHYCARDIA (VT) STORM WITH APPROPRIATE SHOCKS. IT WAS NOTED THAT THE CARDIAC RESY NCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) ENCOUNTERED A CHARGE CIRCUIT TIMEOUT. THE CRT-D WAS EXPLANTED AND REPLACED APPROXIMATELY THREE DAYS LATER. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11251252 ↗InjuryDec 2, 2020Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset
IT WAS REPORTED THAT A PARTIAL ELECTRICAL RESET WAS OBSERVED ON THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) UPON INT ERROGATION, RESULTING IN CLEARING OF DIAGNOSTIC, HISTOGRAM, AND TRENDING DATA. PROGRAMMED PARAMETERS WERE UNCHANGED. THE DEVICE WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10936369 ↗InjuryOct 29, 2020Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE SHOCKS FOR ATRIAL FIBRILLATION WITH RAPID VENTRICULAR RESPONSE DETECTED AS SUPRA VENTRICULAR TACHYCARDIA (SVT). THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 10755542 ↗InjuryOct 26, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. PROGRAMMING ADJUSTMENTS WERE MADE AND THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10732738 ↗InjuryOct 23, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) MAY HAVE EXHIBITED PREMATURE BATTERY DEPLETION. THE CRT-D WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10727969 ↗InjuryOct 21, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10715589 ↗InjuryOct 8, 2020Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10648531 ↗InjuryAug 31, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED DUE TO LEAD VEGETATION EXCEPT FOR THE LEFT VENTRICULAR (LV) LEAD WHICH WAS CAPPED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10475161 ↗
Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices
- Gallant™ CDHFA500QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 recalls58 deaths8,401 reports
- Quartet™ 1458Q-86St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)207 deaths6,631 reports
- QUADRA ASSURA MP™ CD3369-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)103 deaths6,084 reports
- VIGILANT™ X4 CRT-D G247Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)17 deaths2,066 reports
- Quadra Assura™ CD3365-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)53 deaths2,056 reports
- Cobalt™ XT HF Quad CRT-D MRI SureScan™ DTPA2QQMedtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 recalls40 deaths1,639 reports
- Unify Assura™ CD3357-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)54 deaths1,784 reports
- Unify Assura™ CD3357-40CSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)56 deaths1,588 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page