ATTAIN STARFIX™ 419588Recalls & reported problems
Medtronic, Inc. · LEAD 419588 ATTAIN US DEPLOYABLE OTW · Class 3 · Implant
All models of ATTAIN STARFIX (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00643169590649), received between Dec 9, 2016 and May 27, 2021 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Failure to Capture” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Failure to Capture1
- High Capture Threshold1
- Low impedance1
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient3
- Unspecified Infection3
- Wound Dehiscence1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 27, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED DUE TO INFECTION. THE LEFT VENTRICULAR (LV) LEAD WAS PARTIALLY REMOVED WITH THE REMAINDER BEING CAPPED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 11897591 ↗InjurySep 24, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND LEADS WERE EXPLANTED AND LATER REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10582257 ↗InjuryJan 25, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT HAD A POCKET INFECTION. ANTIBIOTIC TREATMENT WAS NECESSARY. THE DEVICE AND LEADS WERE EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 8279319 ↗MalfunctionJan 23, 2019Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS UNDERSENSING DURING RECORDED ARRHYTHMIA EPISODES. THE LEFT VENTRICULAR (LV) LEAD HAD TRIGGERED A WARNING FOR OUT-OF-RANGE PACING IMPEDANCE WHICH WAS CURRENTLY NOTED TO BE LOW. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8271418 ↗MalfunctionApr 25, 2018Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE SUTURE SLEEVE ON THE ATTEMPTED LEFT VENTRICULAR (LV) LEAD BROKE IN HALF AFTER BEING TIED BY THE PHYSICIAN. THE LEAD WAS REMOVED AND A DIFFERENT LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7462094 ↗InjuryAug 9, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO DEVICE POCKET WOUND DEHISCENCE. THE DEVICE SYSTEM WILL BE REPLACED AT A LATER DATE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Wound Dehiscence
MAUDE report 6781934 ↗InjuryDec 9, 2016Failure to Capture; High Capture Threshold; Failure to Capture; High Capture Threshold
IT WAS REPORTED THAT DURING A DEVICE FOLLOW-UP, THE LEFT VENTRICULAR (LV) LEAD EXHIBITED LOSS OF CAPTURE, ALONG WITH A HIGH LV PACING THRESHOLD. THE LV LEAD WAS EXPLANTED AND REPLACED WITH AN EPICARDIAL LEAD. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6162439 ↗
Other Drug eluting permanent left ventricular (lv) pacemaker electrode devices
- Attain Stability™ Quad MRI SureScan™ 479888Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode28 deaths2,213 reports
- Attain Performa™ Straight MRI SureScan™ 439888Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode26 deaths1,595 reports
- Sentus ProMRI OTW QP L-85/49 408719BIOTRONIK SE & Co. KG · Drug eluting permanent left ventricular (lv) pacemaker electrode7 deaths1,385 reports
- Quartet™ 1457Q-86St. Jude Medical, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode21 deaths1,061 reports
- Quickflex™ 1258TSt. Jude Medical, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode26 deaths996 reports
- ATTAIN® BIPOLAR OTW 419488Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode15 deaths730 reports
- Attain Ability® Plus 429688Medtronic, Inc. · Drug eluting permanent left ventricular (lv) pacemaker electrode11 deaths529 reports
- Sentus ProMRI OTW QP S-85/49 406082BIOTRONIK SE & Co. KG · Drug eluting permanent left ventricular (lv) pacemaker electrode1 death511 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.