Sensia™ SESR01 (UDI 00643169708945)Recalls & reported problems
Medtronic, Inc. · IPG SESR01 SENSIA SR IS-1 US NS · Class 3 · Implant
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00643169708945), received between Dec 7, 2017 and Apr 24, 2026 — classified as 9 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Failure to Interrogate” (2 reports).
16 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Appropriate Device Problem Term/Code Not Available2
- Failure to Interrogate2
- Pacing Problem2
- Premature Discharge of Battery2
- Battery Problem: High Impedance1
- Device Sensing Problem1
- Failure to Capture1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Discomfort1
- Dyspnea1
- Medical device site infection1
- No Known Impact Or Consequence To Patient1
- Non specific EKG/ECG Changes1
- Pain1
- Pocket Erosion1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjuryApr 24, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A POCKET EROSION AND AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) AND RIGHT ATRIAL (RA) LEAD WERE REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection; Pocket Erosion
MAUDE report 24979396 ↗MalfunctionNov 27, 2025Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE PROGRAMMER AND RADIOFREQUENCY (RF) HEAD WERE UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). IT WAS NOTED THAT A PATIENT MAGNET HAD BEEN PLACED OVER THE PATIENT¿S DEVICE. NO PACING WAS OBSERVED, AND NO CHANGE WAS OBSERVED IN THE PATIENT¿S HEART RATE. THE PROGRAMMER AND RF HEAD ARE EXPECTED TO BE RETURNED FOR EVALUATION. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23664806 ↗InjuryMay 17, 2024High impedance; Battery Problem: High Impedance; High Capture Threshold; High impedance; Battery Problem: High Impedance; High Capture Threshold
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED HIGH IMPEDANCE AND HIGH BATTERY THRESHOLDS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19338840 ↗InjuryJan 15, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A STROKE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW IMPEDANCE. THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A SUSPECTED FAILURE AND APPEARED TO BE INTERMITTENTLY PACING. THE RV LEAD AND IPG REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Stroke/CVA
MAUDE report 18510364 ↗MalfunctionMar 1, 2023Unable to Obtain Readings; Unable to Obtain Readings
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A DATA COLLECTION ISSUE FOR THE LEAD IMPEDANCE DATA. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16458148 ↗InjuryMay 27, 2022Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SHORTNESS OF BREATH (SOB). THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED ELEVATED THRESHOLDS IN BIPOLAR CONFIGURATION. INITIALLY, THE RV LEAD WAS REPROGRAMMED TO UNIPOLAR CONFIGURATION, AND CAPTURE MANAGEMENT INCREASED UNIPOLAR OUTPUTS TO APPROXIMATELY MAXIMUM, TRIGGERING RECOMMENDED REPLACEMENT TIME (RRT) IN THE IMPLANTABLE PULSE GENERATOR (IPG), TRIGGERING MODE CHANGE, CAUSING THE PATIENT'S SOB. APPROXIMATELY SIX WEEKS LATER, THE RV LEAD WAS CAPPED AND REPLACED AND THE IPG WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dyspnea
MAUDE report 14521802 ↗InjurySep 14, 2021Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. PREMATURE BATTERY DEPLETION WAS ALSO NOTED ON THE IM PLANTABLE PULSE GENERATOR (IPG). BOTH THE RV LEAD AND IPG WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12468402 ↗InjuryMar 9, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED CHRONIC PAIN AT IMPLANT SITE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT UNDERSENSING AND HIGH THRESHOLDS. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED. THE RV LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Pain
MAUDE report 11442287 ↗InjuryJan 13, 2021Failure to Capture; Failure to Interrogate; Pacing Problem; Device Sensing Problem; Impedance Problem; Failure to Capture; Failure to Interrogate; Pacing Problem; Device Sensing Problem; Impedance Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED THREE SECOND PAUSES ON TELEMETRY WITH NO PACING AND POSSIBLE LOSS OF CAPTURE,SENSING ISSUES AND DECLINED IMPEDANCE. THE IPG WAS REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 11162092 ↗MalfunctionSep 1, 2020Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT FEELS UNCOMFORTABLE DUE TO DEVICE MOVING OR INTEGRATION FROM ABDOMEN TO LOWER BREAST. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Discomfort
MAUDE report 10475786 ↗InjuryOct 25, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPAL EPISODES. IT WAS THOUGHT THAT THE IMPLANTED SYSTEM MAY HAVE EXHIBITED NON-CAPTURE. HOWEVER, IT WAS DETERMINED THAT THE PATIENT WAS IN BIGEMINY WHICH GAVE THE ILLUSION OF NON-CAPTURE. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope
MAUDE report 8004937 ↗InjuryDec 7, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PHYSICIAN INADVERTENTLY PLACED A SINGLE CHAMBER IMPLANTABLE PULSE GENERATOR (IPG) INSTEAD OF A DUAL CHAMBER IPG. THE PHYSICIAN SUBSEQUENTLY EXPLANTED AND REPLACED THE SINGLE CHAMBER DEVICE WITH A DUAL CHAMBER DEVICE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7092659 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-1924-2025 · Apr 29, 2025 · Open, ClassifiedAzure S DR MRI SureScan, Product number W3DR01
- Z-2510-2020 · Jun 4, 2020 · Open, ClassifiedAzure S SR MRI, Model Number W3SR01
- Z-2509-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
- Z-2508-2020 · Jun 4, 2020 · Open, ClassifiedAzure S DR MRI, Model Number W3DR01
- Z-2507-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
- Z-0810-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
- Z-0811-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
- Z-0807-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
- Z-0808-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
- Z-0806-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (e) Mo
- Z-0809-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
- Z-2120-2011 · Apr 4, 2011 · TerminatedMedtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
- Z-1440-2010 · Feb 11, 2010 · TerminatedMedtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
…and 1 more.
Other Implantable pulse generator, pacemaker (non-CRT) devices
- Assurity MRI™ PM2272St. Jude Medical, Inc. · Implantable pulse generator, pacemaker (non-CRT)1 recall264 deaths16,996 reports
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths7,984 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,655 reports
- ACCOLADE™ MRI DR L311Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls59 deaths7,283 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,971 reports
- ACCOLADE™ DR L301Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls6 deaths2,513 reports
- ACCOLADE™ EL DR L321Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls2 deaths1,906 reports
- ACUITY™ X4 Spiral S 4674Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)16 deaths1,975 reports