Versa™ VEDR01 (UDI 00643169708983)Recalls & reported problems
Medtronic, Inc. · IPG VEDR01 VERSA DR IS-1 US NS · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic Versa pacemaker has 122 adverse event reports, with 19 deaths and 54 injuries. Many reports describe pacing problems and difficulty communicating with the device, and several describe devices removed for unknown reasons. One open recall, started in January 2019, concerns a circuit error that may affect some pacemakers when programmed to dual chamber mode with atrial sensing.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 122
- Deaths reported
- 19
- Injuries reported
- 54
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 122 adverse event reports that list this exact device (DI 00643169708983), received between Feb 8, 2018 and May 16, 2025 — classified as 19 deaths, 54 injuries, 49 malfunctions.
The most frequently reported specific device problem is “Pacing Problem” (47 reports).
1 FDA recall record name this device's identifier, 1 still open.
15 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jan 17, 2019Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality
Z-0811-2019 · Root cause: Component design/selection · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available55
- Pacing Problem47
- Adverse Event Without Identified Device or Use Problem14
- Communication or Transmission Problem2
- Failure to Interrogate2
- Premature Discharge of Battery2
- Failure to Capture1
- Incorrect Measurement1
Patient problems
- No Known Impact Or Consequence To Patient69
- Death17
- No Clinical Signs, Symptoms or Conditions17
- Unspecified Infection10
- No Consequences Or Impact To Patient4
- Insufficient Information3
- Bradycardia1
- Cardiac Arrest1
Recent adverse event reports
The 25 most recent of 122 reports, in the reporter's own words as published by the FDA.
InjuryMay 16, 2025Premature Discharge of Battery; Failure to Interrogate; Premature Discharge of Battery; Failure to Interrogate
IT WAS REPORTED THAT THE PROGRAMMER WAS UNABLE TO INTERROGATE THE IMPLANTABLE PULSE GENERATOR (IPG). MULTIPLE INTERROGATION ATTEMPTS WERE MADE, INCLUDING MANUALLY SELECTING THE MODEL NUMBER; HOWEVER, THESE ATTEMPTS RESULTED IN ¿NOISY TELEMETRY/UNABLE TO INTERROGATE¿ AND ¿PACEMAKER UPDATE NEEDED¿ MESSAGES. A TRANSMISSION WAS SUBSEQUENTLY PERFORMED TO RETRIEVE DEVICE DATA. IT WAS NOTED THAT THE DEVICE DATA FROM FEBRUARY 2025 SHOWED AN ESTIMATED REMAINING LONGEVITY OF 1.6 YEARS. ON MARCH 25, 2025, A SUDDEN SPIKE IN THE RV THRESHOLD TRIGGERED THE DEVICE TO ADAPT ITS OUTPUT, AND THE DEVICE IMMEDIATELY REACHED RECOMMENDED REPLACEMENT TIME (RRT). DURING THE SUBSEQUENT DEVICE CHECK IN APRIL 2025, CLINIC STAFF NOTED NORMAL DEVICE FUNCTION. THE RV LEAD WAS TESTED THROUGH AN ANALYZER AND SHOWED ACCEPTABLE THRESHOLDS. THE IPG WAS REMOVED AND REPLACED. POST-IMPLANT TESTING CONFIRMED ADEQUATE LEAD PERFORMANCE, WITH RVCM (RIGHT VENTRICULAR CONDUCTION MONITORING) AND HIGH THRESHOLD ALERTS ENABLED ON THE NEW DEVICE. THE PATIENT WAS ALSO ENROLLED IN APP-BASED REMOTE MONITORING AS A PRECAUTIONARY MEASURE. THE PROGRAMMER REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS E
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22033800 ↗MalfunctionFeb 21, 2025Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED NO TELEMETRY. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21430295 ↗InjuryJan 17, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21177162 ↗InjuryNov 29, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20809175 ↗InjuryMar 8, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18865981 ↗MalfunctionDec 4, 2023Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED BY THE PATIENT THAT THE REMOTE MONITOR HAD TELEMETRY ISSUE AND THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED NO TELEMETRY. ADJUSTED THE READER POSITION AND RESET THE REMOTE MONITOR. THE REMOTE MONITOR RECONNECTED. THE REMOTE MONITOR AND IPG REMAINS IN USE. THIS EVENT IS BEING REPORTED IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18254738 ↗InjuryNov 9, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18098876 ↗MalfunctionMar 20, 2023Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED NO TELEMETRY. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16577064 ↗MalfunctionMar 9, 2023Pacing Problem; Pacing Problem
IT WAS REPORTED THE PATIENT EXPERIENCED BRADYCARDIA. THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE SET RATE. THE RA LEAD REMAINS IN USE. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 16511635 ↗InjuryNov 29, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPLACED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, GIVEN THE POTENTIAL FOR LOSS OF DEVICE FUNCTIONALITY, THE IPG WAS EXPLANTED AND REPLACED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15880119 ↗MalfunctionApr 15, 2022Pacing Problem; Pacing Problem
IT WAS REPORTED THAT AN INTERNAL REVIEW OF DATA TRANSMITTED FROM THE DEVICE SHOWED VENTRICULAR SENSE COUNTS IN THE LESS THAN 40 BPM BIN ON THE RATE HISTOGRAM. FOLLOW-UP WAS ATTEMPTED; HOWEVER, NO ADDITIONAL INFORMATION COULD BE OBTAINED. THIS INCIDENT WAS NOT REPORTED BY A HEALTH CARE PROVIDER; THEREFORE, NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14121554 ↗DeathDec 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: Insufficient Information
MAUDE report 13111098 ↗InjuryApr 6, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS FURTHER REPORTED THAT THE IPG WAS NOT EXPLANTED IN RESPONSE TO THE ADVISORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11618189 ↗InjuryMar 31, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPROGRAMMED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, THE DEVICE WAS FUNCTIONING IN A MODE SUSCEPTIBLE TO CIRCUIT ERROR AND WAS THEREFORE REPROGRAMMED TO PRECLUDE HARM TO THE PATIENT. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11594285 ↗DeathJan 28, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: Insufficient Information
MAUDE report 11247599 ↗MalfunctionJan 12, 2021Pacing Problem; Pacing Problem
IT WAS LATER REPORTED THAT DURING AN INTERNAL REVIEW OF DATA TRANSMITTED FROM THE DEVICE, AN ADDITIONAL VENTRICULAR SENSE COUNTS IN THE LESS THAN 40 BPM BIN ON THE RATE HISTOGRAM WAS NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11153747 ↗MalfunctionJan 11, 2021Pacing Problem; Pacing Problem
IT WAS REPORTED THAT AN INTERNAL REVIEW OF DATA TRANSMITTED FROM THE DEVICE SHOWED VENTRICULAR SENSE COUNTS IN THE LESS THAN 40 BPM BIN ON THE RATE HISTOGRAM. FOLLOW-UP WAS ATTEMPTED; HOWEVER, NO ADDITIONAL INFORMATION COULD BE OBTAINED. THIS INCIDENT WAS NOT REPORTED BY A HEALTH CARE PROVIDER; THEREFORE, NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11146675 ↗MalfunctionNov 26, 2020Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) DID NOT EXHIBIT A LONGEVITY ESTIMATOR VALUE. IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT CAPTURE. THE DEVICE AND LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10910201 ↗DeathSep 29, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: Death
MAUDE report 10598856 ↗InjuryAug 31, 2020Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10470292 ↗InjuryJul 22, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10306959 ↗MalfunctionJun 22, 2020Pacing Problem; Pacing Problem
IT WAS REPORTED THAT AN INTERNAL REVIEW OF DATA TRANSMITTED FROM THE DEVICE SHOWED VENTRICULAR SENSE COUNTS IN THE LESS THAN 40 BPM BIN ON THE RATE HISTOGRAM. FOLLOW UP REVEALED THAT NO INTERVENTION HAS BEEN CONDUCTED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10178380 ↗MalfunctionJun 22, 2020Pacing Problem; Pacing Problem
IT WAS REPORTED THAT AN INTERNAL REVIEW OF DATA TRANSMITTED FROM THE DEVICE SHOWED VENTRICULAR SENSE COUNTS IN THE LESS THAN 40 BPM BIN ON THE RATE HISTOGRAM. FOLLOW-UP WAS ATTEMPTED; HOWEVER, NO ADDITIONAL INFORMATION COULD BE OBTAINED. THIS INCIDENT WAS NOT REPORTED BY A HEALTH CARE PROVIDER; THEREFORE, NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10178324 ↗DeathJun 5, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: Death
MAUDE report 10125361 ↗DeathMay 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: Death
MAUDE report 10094104 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-1924-2025 · Apr 29, 2025 · Open, ClassifiedAzure S DR MRI SureScan, Product number W3DR01
- Z-2507-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
- Z-2509-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
- Z-2510-2020 · Jun 4, 2020 · Open, ClassifiedAzure S SR MRI, Model Number W3SR01
- Z-2508-2020 · Jun 4, 2020 · Open, ClassifiedAzure S DR MRI, Model Number W3DR01
- Z-0807-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
- Z-0810-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
- Z-0809-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
- Z-0806-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (e) Mo
- Z-0808-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
- Z-2120-2011 · Apr 4, 2011 · TerminatedMedtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
- Z-1440-2010 · Feb 11, 2010 · TerminatedMedtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
- Z-1515-2009 · May 18, 2009 · TerminatedMedtronic KAPPA 600/700/900 Series Implantable Pulse Generator (IPGs), Dual Chamber Rate Responsive Pacemaker, models KDR601, KDR603, KDR606, KDR651, KDR653, KDR701, KDR703, KDR706
Other Implantable pulse generator, pacemaker (non-CRT) devices
- Assurity MRI™ PM2272St. Jude Medical, Inc. · Implantable pulse generator, pacemaker (non-CRT)1 recall264 deaths16,996 reports
- ACCOLADE™ MRI EL DR L331Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls22 deaths7,984 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)187 deaths7,655 reports
- ACCOLADE™ MRI DR L311Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls59 deaths7,283 reports
- Azure™ XT DR MRI SureScan™ W1DR01Medtronic, Inc. · Implantable pulse generator, pacemaker (non-CRT)107 deaths5,971 reports
- ACCOLADE™ DR L301Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls6 deaths2,513 reports
- ACCOLADE™ EL DR L321Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)2 recalls2 deaths1,906 reports
- ACUITY™ X4 Spiral S 4674Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)16 deaths1,975 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.