Skip to content

Protégé™ GPS™ SERB65-10-80-80 (UDI 00643169729094)Recalls & reported problems

Ev3, Inc · STENT SERB65-10-80-80 PROTEGE GPS V06 · Class 2 · Implant

All models of Protégé GPS (26)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169729094), received between Aug 15, 2024 and Jan 9, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P060001); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Fracture1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 9, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING USE OF THE STENT SERB65-10-80-80 PROTEGE GPS IN THE SUBCLAVIAN, CATHETER HANDLE SEPARATION OCCURRED (COMPLETELY DETACHED). THE LESION LENGTH WAS 60MM, AND A 6 FRENCH SHEATH AND NON MEDTRONIC GUIDEWIRE WERE USED. THE INSTRUCTIONS FOR USE WERE FOLLOWED WITHOUT ISSUE, AND NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. THE HANDLE WAS CRACKED TO AID IN DEPLOYMENT AND THE STENT WAS DEPLOYED IN THE TARGET LOCATION. THE DEVICE WAS SAFELY REMOVED FROM THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24017980 ↗
  • InjuryAug 15, 2024Fracture; Material Deformation; Fracture; Material Deformation

    ON (B)(6) 2024 PHYSICIAN IMPLANTED A PROTEGE GPS SELF-EXPANDING STENT DURING TREATMENT OF AN 80MM LESION IN THE PATIENTS LEFT PROXIMAL CEPHALIC SUBCLAVIAN VEIN. MODERATE VESSEL TORTUOSITY AND MINIMAL VESSEL CALCIFICATION WERE REPORTED. LESION EXHIBITED 70% PLUS STENOSIS. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO ANATOMY. THERE WAS NO DAMAGE NOTED TO PACKAGING, (I.E. SHELF CARTON, HOOP/TRAY). NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE IFU (INSTRUCTION FOR USE) WAS FOLLOWED DURING PREPARATION, PROCEDURE, POST-PROCEDURE WITHOUT ISSUE. EMBOLIC PROTECTION WAS NOT USED. NO RESISTANCE WAS NOTED DURING ADVANCEMENT OF THE DEVICE. ON (B)(6) 2024 DURING DEVICE FOLLOW UP A FRACTURE WAS NOTED TO THE DEPLOYED STENT. A REPLACEMENT 10X80X80 MEDTRONIC STENT WAS PLACED OVER THE PROTÉGÉ STENT. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19989438 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P060001. One submission often covers many models, so check the product and lot details in each record.

  • Z-0879-2015 · Dec 4, 2014 · Terminated
    ev3, Protg Rx Tapered Carotid Stent System, SEPX-8-6-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endarterect
  • Z-0878-2015 · Dec 4, 2014 · Terminated
    ev3 Protege Rx Tapered Carotid Stent System, SECX-10-7-40-135, Sterile EO, Rx only. Indicated for the treatment of patients at high risk for adverse events from carotid endartere

Other CATHETER, BILIARY, DIAGNOSTIC devices

All CATHETER, BILIARY, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page