Skip to content

Visi-Pro™ PXB35-06-57-080Recalls & reported problems

Covidien LP · PXB35-06-57-080 STENT VISI PRO 035 V01 · Class 2 · Implant

All models of Visi-Pro (66)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169787728), received between Aug 29, 2023 and Aug 29, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Difficult to Advance1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 29, 2023Difficult to Advance; Device Dislodged or Dislocated; Difficult to Advance; Device Dislodged or Dislocated

    A PHYSICIAN WAS ATTEMPTING TO IMPLANT A VISIPRO STENT FOR THE TREATMENT OF A 55MM CALCIFIED LESION WITH 90% STENOSIS IN THE MID LEFT COMMON ILIAC ARTERY. THE ARTERY WAS 6MM WITH MODERATE CALCIFICATION A MILD TORTUOSITY. A 6FR NON MEDTRONIC SHEATH WAS USED AND A 0.035 X 260CM NON MEDTRONIC GUIDEWIRE. EMBOLIC PROTECTION WAS NOT USED. THE DEVICE WAS PREPPED AS PER THE IFU WITH NO ISSUES IDENTIFIED. THE LESION WAS PRE-DILATED WITH A 4MMX80MM NON MEDTRONIC PRE-DILATION DEVICE. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS MET WHEN ADVANCING THE DEVICE. THERE WAS A MODERATE AMOUNT OF CALCIUM IN THE VESSEL. THE PHYSICIAN DECIDED TO PROCEED WITHOUT ADVANCING THE SHEATH AND ATTEMPTED TO PLACE THE STENT. DURING THIS ATTEMPT IT WAS OBSERVED THAT THE STENT HAD BECOME DISLODGED FROM THE DELIVERY SYSTEM AND WHEN THE BALLOON CATHETER WAS REMOVED, THE STENT REMAINED INSIDE OF THE PATIENT. THE STENT APPEARED TO BE DEFORMED. THE DISLODGED STENT REMOVED WAS REMOVED VIA SURGERY. IT WAS REPORTED THAT THE PATIENT WAS STABLE WITHOUT ANY COMPLAINTS OR INJURY. THE PHYSICIAN DECIDED TO DO AN OPEN AORTO-BI-FEM PROCEDURE AS THE PATIENT HAD OTHER DISEASED VESSELS THAT NEEDED TO BE

    Patient problems: Insufficient Information

    MAUDE report 17642046 ↗

Other Stents, drains and dilators for the biliary ducts devices

All Stents, drains and dilators for the biliary ducts devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page