Claria MRI™ Quad CRT-D SureScan™ DTMA1Q1 (UDI 00643169837669)Recalls & reported problems
Medtronic, Inc. · CRTD DTMA1Q1 CLARIA MRI QUAD US DF1 · Class 3 · Implant
All models of Claria MRI Quad CRT-D SureScan (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic Claria MRI CRT-D is an implanted defibrillator that paces and shocks the heart. Many of the 289 reports describe electrical faults, sensing and pacing problems, inappropriate shocks, and batteries running down early. Reports fell from 16 in the prior 12 months to 8 in the last 12 months. The reports include 7 deaths and 153 injuries. The single open recall started in May 2023 and concerns a rare risk of reduced or no shock energy.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 289
- Deaths reported
- 7
- Injuries reported
- 153
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 289 adverse event reports that list this exact device (DI 00643169837669), received between Jun 12, 2018 and Sep 23, 2026 — classified as 7 deaths, 153 injuries, 129 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (65 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 10, 2023Class IOpen, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1755-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem69
- Electrical /Electronic Property Problem65
- Device Sensing Problem39
- Inappropriate/Inadequate Shock/Stimulation31
- Premature Discharge of Battery28
- Pacing Problem25
- Electromagnetic Interference19
- Over-Sensing15
Patient problems
- No Clinical Signs, Symptoms or Conditions90
- Unspecified Infection58
- No Patient Involvement33
- Shock from Patient Lead(s)29
- No Known Impact Or Consequence To Patient19
- No Consequences Or Impact To Patient10
- Tachycardia10
- Erosion9
Recent adverse event reports
The 25 most recent of 289 reports, in the reporter's own words as published by the FDA.
InjurySep 23, 2026Premature Discharge of Battery
It was reported that the Left Ventricular (LV) lead "thresholds "[were]" leading to increased battery drain" from the Cardiac Resynchronization Therapy Defibrillator (CRT-D). The lead was reprogrammed, and the lead and the device remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50873460 ↗MalfunctionMar 11, 2026Pacing Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXHIBITING OCCASIONAL INAPPROPRIATE ATRIAL PACING. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24570217 ↗MalfunctionMar 10, 2026Electromagnetic Interference; Over-Sensing
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DEVICE EXHIBITED ELECTROMAGNETIC INTERFERENCE RESULTING IN ATRIAL OVERSENSING THAT LEAD TO MISCLASSIFICATION OF EPISODE TYPE. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24560797 ↗InjuryMar 2, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS FURTHER REPORTED THE PATIENT UNDERWENT VENTRICULAR TACHYCARDIA (VT) ABLATION.
Patient problems: Shock from Patient Lead(s); Tachycardia
MAUDE report 24484998 ↗InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT COMPLAINED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS WARM. THE HEALTHCARE PROFESSIONAL IS HAVING LABS DRAWN TO ASSESS FOR INFECTION. THE HEALTHCARE PROFESSIONAL STATES THE DEVICE FEELS WARM COMPARED TO THE OTHER SIDE OF CHEST. THE DEVICE REMAINS IN USE. IT WAS FURTHER REPORTED THAT THE PATIENT FEELS HOT INSIDE THE CRT-D POCKET. THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT WITH THE DEVICE POCKET AREA RED AND SWOLLEN INDICATIVE OF CELLULITIS. THE PATIENT WAS TREATED WITH TWO ROUNDS OF ANTIBIOTICS GIVEN EMBOLIC INFECTION TO THE CRT-D POCKET REGION. THE FLUID UNDER THE CRT-D AREA FELT HOT. BLOOD CULTURES WERE POSITIVE FOR C-REACTIVE PROTEIN. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Blood Protein Abnormal; Cellulitis; Erythema; Inflammation; Medical device site infection
MAUDE report 23649166 ↗InjuryNov 18, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DEVICE EXHIBITED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23586890 ↗MalfunctionNov 7, 2025Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED BY THE PATIENT TO BE IN ATRIAL FIBRILLATION (AF) FOR ABOUT A WEEK, SENT A MANUAL TRANSMISSION AND THE ON CALL TECHNICIAN LOOKED AT THE TRANSMISSION. ADDITIONALLY, THE PATIENT STATED, HAS HEARD THE ALARM FOR ABOUT FIVE SECONDS AND WAS TOLD BY THE PHYSICIAN TO CONTACT CARELINK TRANSMISSION SUPPORT BECAUSE THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) IS TRYING TO COMMUNICATE BUT IS NOT GOING THROUGH. THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Atrial Fibrillation
MAUDE report 23497346 ↗InjuryNov 3, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DEVICE EXPERIENCED PREMATURE BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23458074 ↗InjurySep 10, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS FURTHER REPORTED THAT THE CRT-D WAS EXPLANTED AND REPLACED AS A STANDARD ELECTIVE REPLACEMENT INDICATOR (ERI) GENERATOR CHANGEOUT. "NON-PROPHYLACTIC" WAS MISTAKENLY CHECKED ON THE PAPERWORK.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23017436 ↗InjuryAug 26, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED AND REPLACED FOR AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22896069 ↗InjuryAug 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED BY THE PATIENT THAT THEIR CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO A METH ICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Drug Resistant Bacterial Infection
MAUDE report 22708417 ↗InjuryJul 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED FOR A CARDIAC TRANSPLANTATION. IT WAS NOTED THAT THERE WAS SOME SUSPICION PREOPERATIVELY THAT THE PATIENT HAD LEAD RELATED ENDOCARDITIS. IT WAS REPORTED THAT PORTIONS OF THE RIGHT VENTRICULAR (RV) LEAD AND CORONARY SINUSLEFT VENTRICULAR (LV) LEAD HAD SOME THROMBUS OR VEGETATION ON THEM. THE ENTIRE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Endocarditis; Medical device site infection
MAUDE report 22649660 ↗MalfunctionJul 22, 2025Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH BRADYCARDIA, DIZZINESS AND NEAR SYNCOPE. A DEVICE INTERROGATION INDICATED OVERSENSING ON THE RIGHT ATRIAL (RA) LEAD. IT WAS REPORTED BY THE PATIENT'S SPOUSE THE FOLLOWING DAY THAT THE PATIENT'S HEART RATE WAS GOING DOWN TO FORTY BEATS PER MINUTE. THE PATIENT'S CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS PROGRAMMED TO A LOWER RATE OF SIXTY BEATS PER MINUTE. THE RA LEAD AND CRT-D REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 22566995 ↗InjuryJul 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN UNSPECIFIED INFECTION. IT WAS STATED THAT THE PATIENT WAS GOING TO GO THROUGH A SYSTEM EXTRACTION DUE TO THE INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) AND THE RIGHT VENTRICULAR (RV) LEAD WERE REPLACED. THE LEFT VENTRICULAR (LV) LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 22470609 ↗InjuryJul 7, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH CAPTURE THRESHOLDS WHICH RESULTED IN PREMATURE BATTERY DEPLETION. THE LEAD WAS EXPLANTED AND REPLACED. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22411504 ↗MalfunctionMay 16, 2025Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) PATIENT PRESENTED TO THE EMERGENCY ROOM AND WAS FEELING DIZZY. A HEART MONITOR SHOWED A BRADYCARDIC HEART RATE IN THE THIRTY AND FORTY BEATS PER MINUTE RANGE. IT WAS NOTED THAT THE PATIENT WAS PROGRAMMED TO A LOWER RATE OF SIXTY BEATS PER MINUTE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Dizziness
MAUDE report 22034217 ↗InjuryMar 17, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE HIGH VOLTAGE THERAPY AND ANTI-TACHYCARDIA PACING FOR SUSPECTED SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT WAS DETECTED AS VENTRICULAR TACHYCARDIA (VT). THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 21620801 ↗MalfunctionFeb 13, 2025Mechanical Problem; Misconnection; Mechanical Problem; Misconnection
IT WAS REPORTED THAT DURING A GENERATOR CHANGEOUT PROCEDURE, THE LEFT VENTRICULAR (LV) LEAD WAS STUCK IN THE HEADER OF THE EXISTING CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) PORT. REMOVING THE SET SCREW, APPLYING MINERAL OIL, FLUSHING WITH SALINE, AND USING FORCE WERE UNSUCCESSFUL AT REMOVING THE LV LEAD FROM THE CRTD PORT. THE PATIENT REPORTED THAT A GRINDER WAS USED TO REMOVE THE SET SCREW. ULTIMATELY, THE LV LEAD WAS REMOVED FROM THE PORT AND SUCCESSFULLY CONNECTED TO THE REPLACEMENT CRTD AFTER TESTING NORMALLY. THE CRTD WAS ELECTIVELY REPLACED AND THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21379360 ↗InjuryFeb 3, 2025Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. THE CRT-D WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21300045 ↗InjuryDec 27, 2024Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION, AND THE LEFT VENTRICULAR (LV) LEAD THRESHOLDS WERE HIGH. THE DEVICE WAS EXPLANTED AND REPLACED, AND THE LV LEAD WAS REPROGRAMMED AND LATER CAPPED AND REPLACED WITH A NEW LV LEAD IN A DIFFERENT BRANCH. IT WAS ALSO REPORTED THAT WHEN PREPPING FOR THE REPLACEMENT PROCEDURE THEY WERE UNABLE TO MAKE ANY PROGRAMMING CHANGES TO THE LV LEAD PACE POLARITY AS THE DEVICE HAD REACHED RECOMMENDED REPLACEMENT TIME (RRT). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21030592 ↗InjuryDec 13, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO METHICILLIN-SENSITIVE STAPHYLOCOCCUS AUREUS BACTEREMIA. A WEEK LATER, THE SYSTEM WAS REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Bacterial Infection
MAUDE report 20929483 ↗MalfunctionNov 18, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT HAD A CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DEVICE CHECK DUE TO ELEVATED HEART RATE. THE PHYSICIAN REPORTED THE DEVICE/CRT-D WAS MALFUNCTIONING. THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20708098 ↗InjuryNov 7, 2024Mechanical Problem; Misconnection; Mechanical Problem; Misconnection
IT WAS REPORTED THAT DURING A CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) CHANGEOUT PROCEDURE, THE RIGHT ATRIAL (RA) LEAD COULD NOT BE REMOVED FROM THE CRT-D HEADER DESPITE MULTIPLE ATTEMPTS WITH THE WRENCH TO UNSCREW THE SET SCREW. FINALLY, THE WRENCH ENGAGED, AND THE SET SCREW WAS COMPLETELY REMOVED; HOWEVER, THE LEAD REMAINED STUCK IN THE HEADER. AN RA LEAD WAS NOT NECESSARY FOR THE PATIENT, SO THE LEAD WAS CUT TO FREE IT FROM THE DEVICE AND WAS PARTIALLY EXPLANTED AND NOT REPLACED. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD AN INSULATION BREACH AND THERE WAS ¿OLD¿ BLOOD IN THE LEAD. THE LEAD WAS CAPPED AND REPLACED. THE CRT-D WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20636648 ↗InjuryNov 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. VEGETATION WAS NOTED ON THE LEADS. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED. EIGHT DAYS LATER THE DEVICE SYSTEM WAS REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20632006 ↗MalfunctionOct 29, 2024Capturing Problem; Capturing Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR CAPTURE MANAGEMENT WAS POTENTIALLY CAUSING VENTRICULAR TACHYCARDIA (VT). THE CARDIAC RESY NCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Tachycardia
MAUDE report 20559906 ↗
Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices
- Gallant™ CDHFA500QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)2 recalls58 deaths8,401 reports
- Quartet™ 1458Q-86St. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)207 deaths6,631 reports
- QUADRA ASSURA MP™ CD3369-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)103 deaths6,084 reports
- VIGILANT™ X4 CRT-D G247Boston Scientific Corporation · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)17 deaths2,066 reports
- Quadra Assura™ CD3365-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)53 deaths2,056 reports
- Cobalt™ XT HF Quad CRT-D MRI SureScan™ DTPA2QQMedtronic, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 recalls40 deaths1,639 reports
- Unify Assura™ CD3357-40QSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)54 deaths1,784 reports
- Unify Assura™ CD3357-40CSt. Jude Medical, Inc. · Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)56 deaths1,588 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page