Skip to content

NA BB4B15R13Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK BB4B15R13 H/L · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jan 3, 2018Terminated

    Medtronic custom perfusion tubing packs contain Terumo OPS valves that were recalled due to a lack of flow through the valve.

    Z-0535-2018 · Root cause: Component change control · FDA record ↗

  2. Nov 16, 2016Terminated

    Medtronic is initiating a voluntary product recall for specific lot numbers of the Affinity Fusion Oxygenators with Balance1 Biosurface. These are distributed as stand-alone devices, or as a combination unit with the Affinity Fusion Cardiotomy/Venous Reservoir, or as part of tubing packs. Medtronic has identified an out-of-specification condition exhibiting additional plastic (flash) in the arterial sampling port adjacent to the arterial outlet port of the oxygenator.

    Z-0826-2017 · Root cause: Process control · FDA record ↗

Other Pump, blood, cardiopulmonary bypass, non-roller type devices

All Pump, blood, cardiopulmonary bypass, non-roller type devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page