INTERSTIM® 3889-28 (UDI 00643169864436)Recalls & reported problems
Medtronic, Inc. · LEAD 3889-28 ISTIM QUAD SNS 28CM US EMAN · Class 3 · Implant
All models of INTERSTIM (30)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic InterStim lead is a thin implanted wire used in nerve stimulators for bladder or bowel control. Most of the 56 reports describe the lead breaking, fraying, or having its insulation peel or tear away during handling or surgery, sometimes leaving it detached from the battery and requiring removal; 26 injuries and 30 malfunctions were reported, with no deaths. A few patients reported losing the stimulation effect. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 56
- Deaths reported
- 0
- Injuries reported
- 26
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 00643169864436), received between Jun 26, 2018 and Jul 9, 2025 — classified as 26 injuries, 30 malfunctions.
The most frequently reported specific device problem is “Break” (34 reports).
7 other recalls cover devices cleared under the same FDA submission (P080025, P970004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break34
- Peeled/Delaminated17
- Insufficient Device Problem Information8
- Device Damaged Prior to Receipt by User7
- Adverse Event Without Identified Device or Use Problem6
- Difficult to Insert4
- Material Twisted/Bent3
- Device Contamination with Chemical or Other Material2
Patient problems
- No Known Impact Or Consequence To Patient42
- Therapeutic Effects, Unexpected5
- No Clinical Signs, Symptoms or Conditions3
- Device Embedded In Tissue or Plaque2
- Apnea1
- Headache1
- Hypersensitivity/Allergic reaction1
- Itching Sensation1
Recent adverse event reports
The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.
InjuryJul 9, 2025Migration or Expulsion of Device; Migration or Expulsion of Device
ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE (REP). THE REP REPORTED THAT THE MOST LIKELY CAUSE OF THE PATIENT NO LONGER FEELING STIMULATION WAS THE RESULT OF THE BATTERY REACHING END OF SERVICE (EOS), AS DETERMINED BY THE PATIENT'S HCP. WHEN THE HCP INTERROGATED THE BATTERY, THE PATIENT DIDN'T MENTION ANYTHING TO THE HCP ABOUT BEING INVOLVED IN A CAR ACCIDENT. THE PATIENT STATED THAT PRIOR TO THE CAR ACCIDENT, THEIR SYMPTOMS HAD STARTED TO RETURN, WHICH WAS ATTRIBUTED TO THE BATTERY LIFE. THE REP STATED THE PATIENT HAD THEIR BATTERY REPLACED ON (B)(6) 2025, AND THE IMPEDANCES WERE WITHIN NORMAL LIMITS. THE PATIENT FELT NO SENSATION POST-OP WHEN THERAPY WAS TURNED ON (TESTED UP TO 8.0 MA). THE REP FOLLOWED UP WITH THE PATIENT 48 HOURS LATER AND THEY WERE STILL NOT FEELING SENSATION AND IT WAS AT THIS TIME THE PATIENT SHARED THEY WERE INVOLVED IN A CAR ACCIDENT IN (B)(6). THE REP STATED AN ANTEROPOSTERIOR (AP) AND LATERAL X-RAY WERE ORDERED BY THE PATIENT'S HCP, AND THE PATIENT HAD A FOLLOW-UP APPOINTMENT SCHEDULED WITH THEIR HCP ON (B)(6) 2025. THE REP STATED, THE CAUSE OF THERAPY NOT WORKING WAS THE RESULT OF THE BATTERY REACHING EOS, AND THE X-RAYS WER
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22460434 ↗MalfunctionNov 10, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT THERE WAS A REPORT ON LEAD INTEGRITY. THE HCP NOTICED MARKS ON LEAD WHEN IT WAS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15776366 ↗InjuryMar 11, 2022Break; Break
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR UNKNOWN INDICATIONS FOR USE. IT WAS REPORTED THAT THERE WAS A LEAD BREAKAGE. THE LEAD FRAYED AND NOT ATTACHED TO BATTERY. IT WAS EXPLANTED AND ISSUE RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13738833 ↗InjurySep 8, 2020Peeled/Delaminated; Peeled/Delaminated
ADDITIONAL INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MANUFACTURER REPRESENTATIVE (REP). IT WAS REPORTED THAT THE CAUSE OF HOW THE INSULATED COATING CAME APART WAS UNKNOWN, BUT THEY DID KNOW IT HAPPENED WHEN THE BOOT WAS PRESSED OVER THE CONNECTION PIECE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10502789 ↗MalfunctionJul 1, 2020Break; Difficult to Insert; Material Twisted/Bent; Break; Difficult to Insert; Material Twisted/Bent
ADDITIONAL INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP). THE REP REPORTED THAT IT WAS REPORTED THAT THE INITIAL LEAD HAD NEVER BEEN TUNNELED TO BEGIN WITH. THE SURGERY PROCEDURE TOOK LONGER THAN EXPECTED AS A RESULT OF THE ISSUE. NO FURTHER PATIENT COMPLICATIONS ARE ANTICIPATED OR EXPECTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10219098 ↗MalfunctionJun 17, 2020Break; Peeled/Delaminated; Human-Device Interface Problem; Break; Peeled/Delaminated; Human-Device Interface Problem
INFORMATION WAS RECEIVED BY A MANUFACTURE REPRESENTATIVE (REP) REGARDING AN EXTERNAL NEUROSTIMULATOR (ENS). IT WAS REPORTED THAT THE HEALTHCARE PROVIDER (HCP) APPLIED TO MUCH PRESSURE TO THE LEAD WHEN PUTTING ON THE BOOT DURING THE STAGE 1 PROTOCOL AND THE INSULATION RIPPED OFF OF THE LEAD. THE LEAD WAS REPLACED. NO PATIENT SYMPTOMS WERE REPORTED AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10162550 ↗MalfunctionJun 12, 2020Impedance Problem; Impedance Problem
ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE (REP). THE REP STATED THAT THERE WAS NO DEVICE ISSUE AND NO PATIENT/PROCEDURE ISSUE. THE REP STATED THAT WHEN CHECKED POST-OPERATIVELY, THE IMPEDANCE ISSUE WAS RESOLVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10149080 ↗MalfunctionJun 2, 2020Break; Positioning Problem; Break; Positioning Problem
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE (REP) REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). THE REP REPORTED THAT THE IMPLANTING PHYSICIAN NOTED RESISTANCE WHEN THE LEAD WAS ADVANCED THROUGH THE LEAD INTRODUCER SHEATH. THE RESISTANCE WAS NOTED WHEN THE FIRST DEPTH MARKER OF THE LEAD PASSED THROUGH THE TOP OF THE INTRODUCER SHEATH, HOWEVER THEY WERE UNABLE TO ADVANCE THE LEAD ANY FURTHER. THE LEAD WAS REMOVED FROM THE INTRODUCER SHEATH AND IT WAS NOTED THAT THE LEAD HAD BROKEN BETWEEN THE TINES. THERE WERE NO CONTRIBUTING FACTORS REPORTED. THE BROKEN LEAD WAS REMOVED FROM THE STERILE FIELD AND A NEW LEAD WAS OPENED TO FIELD AND IMPLANTED WITHOUT AN ISSUE. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED OR ANTICIPATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10109172 ↗MalfunctionMay 19, 2020Break; Peeled/Delaminated; Break; Peeled/Delaminated
INFORMATION WAS RECEIVED FROM AN HEALTHCARE PROFESSIONAL VIA A MANUFACTURE REPRESENTATIVE REGARDING A PATIENT WHO WAS USING WITH AN EXTERNAL NEUROSTIMULATOR (ENS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIMULATION AND GASTROINTESTINAL/PELVIC FLOOR. IT WAS REPORTED THAT THE PUTTER SHEATH IN THE LEAD WAS PULLED AWAY A SMALL DISTANCE AT THE END THAT GOES INTO THE INS. IT WAS FOUND WHEN MOVING FROM STAGE 1 TO STAGE 2. ONCE THE LEAD WAS PLACED INTO THE INS WITH THE SET SCREW THE ENTIRE PORTION WAS COVERED AND HAD PERFECT IMPEDANCE. NO SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED. 2020-MAY-19 E1 (REP): NO NEW INFO FOR EVENT DESCRIPTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10070679 ↗MalfunctionMay 8, 2020Device Damaged Prior to Receipt by User; Component Missing; Device Damaged Prior to Receipt by User; Component Missing
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MANUFACTURER¿S REPRESENTATIVE (REP) REGARDING A PATIENT UNDERGOING AN IMPLANT FOR NEUROSTIMULATOR (INS) FOR URINARY DYSFUNCTION/SACRAL NERVE STIMULATION AND GASTROINTESTINAL /PELVIC FLOOR THERAPY. IT WAS REPORTED THE REP WAS AT A CASE TODAY AND THE LEAD KIT WAS MISSING A TUNNELING TOOL WEDGE TIP, SO THE PHYSICIAN USED ANOTHER LEAD KIT. NO SYMPTOMS WERE REPORTED AND NO FURTHER COMPLICATIONS WERE REPORTED OR ARE ANTICIPATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10038728 ↗MalfunctionMar 20, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
NO NEW INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9860231 ↗InjuryMar 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ADDITIONAL INFORMATION WAS RECEIVED FROM A HEALTH CARE PROFESSIONAL. IT WAS REPORTED THAT THE PATIENT¿S ALLERGIC REACTION DID RESOLVE. NO FURTHER PATIENT COMPLICATIONS ARE ANTICIPATED OR EXPECTED AS A RESULT OF THIS EVENT.
Patient problems: Hypersensitivity/Allergic reaction; Itching Sensation; Swelling
MAUDE report 9844924 ↗MalfunctionMar 17, 2020Peeled/Delaminated; Peeled/Delaminated
INFORMATION WAS RECEIVED BY A MANUFACTURE REPRESENTATIVE (REP) REGARDING A TRIAL EXTERNAL NEUROSTIMULATOR (ENS). IT WAS REPORTED THAT DURING AN LEAD PLACEMENT FOR TRIAL, THE HEALTHCARE PROVIDE ATTEMPTED TO SLIDE BOOT OVER THE CONNECTED PERCUTANEOUS EXTENSION, AND LEAD WAS SEPARATED FROM THE ELECTRODES EXPOSING THE INSULATED PORTION OF THE LEAD. THE DAMAGED LEAD WAS REMOVED AND REPLACED. THE DAMAGED LEAD WAS DISCONNECTED AND REMOVED AND ENTIRE LEAD WAS REPLACED. NO PATIENT SYMPTOMS WERE REPORTED AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9844591 ↗InjuryFeb 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED FROM A CONSUMER VIA THE MANUFACTURER¿S REPRESENTATIVE (REP) REGARDING A PATIENT WITH AN IMPLANTED NEUROSTIMULATOR (INS) FOR GASTROINTESTINAL/PELVIC FLOOR THERAPY. IT WAS REPORTED THERE WAS AN INFECTION FOUND AT THE POCKET SITE WHEN THE STAGE 1 EVALUATION WAS FINISHED. THE PHYSICIAN CLEANED THE POCKET, IRRIGATED POCKET WITH ANTIBIOTICS AND PROCEEDED WITH STAGE 2 IMPLANT AND THE ISSUE WAS RESOLVED AT THE TIME OF THE REPORT. NO FURTHER COMPLICATIONS WERE REPORTED OR ARE ANTICIPATED.
Patient problems: Unspecified Infection
MAUDE report 9735071 ↗InjuryFeb 20, 2020Break; Break
NO NEW INFORMATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9733495 ↗InjuryJan 31, 2020Break; Peeled/Delaminated; Break; Peeled/Delaminated
THE REP REPORTED INFORMATION IDENTIFYING THE PATIENT AND DID NOT KNOW THE WEIGHT OF THE PATIENT. INFORMATION WAS ALSO REPORTED IN MANUFACTURER REPORT NUMBER 3007566237-2020-00169 AND WAS DIRECTED TO FOLLOW UP HERE GOING FORWARD.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9653326 ↗InjuryJan 27, 2020Peeled/Delaminated; Peeled/Delaminated
INFORMATION WAS RECEIVED BY A MANUFACTURE REPRESENTATIVE (REP) VIA A HEALTHCARE PROVIDER (HCP) REGARDING A LEAD. REP REPORTED A LEAD BREAK DURING THE PROCEDURE; NO FAULT WAS BLAMED ON THE HCP. THEY NOTED THAT THE PLASTIC WAS PULLED BACK FORM THE WIRE. NO ENVIRONMENTAL/EXTERNAL/PATIENT FACTORS THAT MAY HAVE LED OR CONTRIBUTED TO THE ISSUE ARE KNOWN. THE ACTION/INTERVENTION TAKEN TO RESOLVE THE ISSUE WAS THE BROKEN LEAD WAS SURGICALLY REMOVED AND AN INTACT ONE WAS IMPLANTED. THE ISSUE WAS RESOLVED AT THE TIME OF THE REPORT. NO PATIENT SYMPTOMS WERE REPORTED AND NO FURTHER COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9633067 ↗MalfunctionJan 27, 2020Peeled/Delaminated; Peeled/Delaminated
INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL (HCP), VIA THE MANUFACTURER¿S REPRESENTATIVE, REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT THE COATING OF THE LEAD CAME BACK, WHICH EXPOSED THE COILED WIRES. AS AN FACTOR THAT MAY HAVE LED TO THE ISSUE, THE PHYSICIAN REMOVED AND REPOSITIONED THE LEAD SEVERAL TIMES TRYING TO FIND OPTIMAL PLACEMENT. NO TROUBLESHOOTING WAS PERFORMED. THE LEAD WAS THEN REMOVED AND REPLACED. NO SURGICAL INTERVENTION HAD OCCURRED AND NONE WERE PLANNED. THE ISSUE WAS REPORTED AS RESOLVED. THERE WERE NO FURTHER COMPLICATIONS REPORTED OR ANTICIPATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9632948 ↗InjuryJan 23, 2020Break; Break
INFORMATION WAS RECEIVED BY A MANUFACTURE REPRESENTATIVE (REP) VIA A HEALTHCARE PROVIDER (HCP) REGARDING A LEAD. A REP REPORTED A LEAD BREAK UPON REMOVAL. THEY NOTED THAT THEY WEREN'T PRESENT IN THE CASE AND WAS INFORMED THAT THE HCP WASN'T PUTTING A LOT OF TENSION ON THE LEAD WHILE REMOVING IT, BUT IT SEPARATED. THE INTERVENTIONS/ACTIONS TAKEN TO RESOLVE THE ISSUE INCLUDED ASKING THE HCP TO SAVE THE LEAD FOR THE REP SO IT COULD BE RETURNED FOR ANALYSIS. THE HCP ALSO NOTED THAT THEY HAD TO DISSECT DOWN IN ORDER TO REMOVE THE REMAINING PORTION. THE ISSUE WAS RESOLVED AT THE TIME OF THE REPORT. NO PATIENT SYMPTOMS WERE REPORTED AND NO FURTHER COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9622348 ↗InjuryJan 21, 2020Break; Break
ADDITIONAL INFORMATION WAS RECEIVED. IT WAS CONFIRMED THAT THE OUTER INSULATION SEPARATED AT THE BUTT JOINT OF THE PROXIMAL END. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9613024 ↗MalfunctionJan 10, 2020Break; Difficult to Remove; Adverse Event Without Identified Device or Use Problem; Break; Difficult to Remove; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED BY A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURE REPRESENTATIVE (REP) REGARDING AN IMPLANTABLE NEUROSTIMULATOR (INS) FOR AN UNKNOWN PELVIC HEALTH THERAPY. IT WAS REPORTED THAT THE LEAD BROKE AT THE FORAMEN DURING EXPLANT AND ALL 4 ELECTRODES REMAIN IN S3. THE HCP TRIED TO EXPOSE TIMES AND WAS UNABLE TO VISUALIZE DUE TO THE SIZE OF THE PATIENT. UTILIZED FLUOROSCOPY TO IDENTIFY LOCATION OF ELECTRODES. A NEW LEAD WAS PLACED IN THE FORAMEN AND MOTOR RESPONSE WAS ACHIEVED ON ALL FOUR CONTACTS UNDER 1.0MA. NO FURTHER DEVICE ISSUE ALLEGED/IDENTIFIED. NO FURTHER PATIENT SYMPTOMS OR COMPLICATIONS WERE REPORTED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9577503 ↗InjuryJan 8, 2020Break; Break
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT THE LEAD WAS READY TO BE CONNECTED TO THE INS DURING A REVISION/ REPLACEMENT. THE SHEATH OF THE LEAD WAS FOUND TO BE LOOSE, EXPOSING THE BRAIDED WIRES NEAR THE ELECTRODES/ BELOW THE BLUE ELECTRODE. THE LEAD WAS REMOVED WITHOUT DIFFICULTY AND A NEW LEAD WAS PLACED. THERE WERE NO KNOWN CONTRIBUTING FACTORS. IT WAS REPORTED THAT THE ISSUE WAS RESOLVED AT THE TIME OF THE REPORT. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED OR ANTICIPATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9564291 ↗InjuryJan 7, 2020Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA A MANUFACTURER REPRESENTATIVE (REP) REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS). IT WAS REPORTED THAT THE LEAD APPEARED TO NOT LOOK RIGHT ACCORDING TO THE PHYSICIAN, BUT IT WAS TUNNELED TO THE POCKET ANYWAY. AFTER RETRIEVING THE LEAD FROM THE TUNNELING TOOL SHEATH, THE ELECTRODES APPEARED TO BE PULLING APART FROM THE LEAD WIRE. THE LEAD WAS REMOVED AND A NEW LEAD WAS PLACED WITH NO ISSUES. THE PATIENT WAS UNHARMED. THERE WERE NO KNOWN CONTRIBUTING FACTORS. IT WAS REPORTED THE ISSUE WAS RESOLVED AT THE TIME OF THE REPORT. NO FURTHER COMPLICATIONS WERE REPORTED OR ANTICIPATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9557291 ↗MalfunctionDec 19, 2019Peeled/Delaminated; Peeled/Delaminated
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WHO WAS IMPLANTED WITH A NEUROSTIMULATOR (INS) FOR GASTROINTESTINAL/PELVIC FLOOR AND URINARY DYSFUNCTION/SACRAL NERVE STIM. IT WAS REPORTED THAT THE LEAD WAS DEFECTIVE AND WAS NEVER USED ON THE PATIENT. WHEN THE SCRUB TECHNICIAN OPENED THE LEAD AND HANDED IT TO THE PHYSICIAN. THE PHYSICIAN NOTICED THAT THE PLASTIC CASING BETWEEN THE TINES AND THE ELECTRODES WERE NOT ATTACHED LEAVING RAW WIRES EXPOSED BETWEEN THE TINES AND ELECTRODES. THE NURSE OPENED A NEW LEAD WITH THE SAME LOT NUMBER AND THERE WAS NO DEFECT OBSERVED. THE NEW LEAD WAS USED FOR THE PROCEDURE AND THE DEFECTIVE LEAD WAS PLACED IN A BIOHAZARD BAD AND WAS GIVEN TO THE MANUFACTURER REPRESENTATIVE. THE PATIENT IS ALIVE AND NO INJURY WAS REPORTED AT THE TIME OF THIS REPORT. NO FURTHER PATIENT COMPLICATIONS ARE ANTICIPATED OR EXPECTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9495920 ↗MalfunctionDec 19, 2019Break; Peeled/Delaminated; Break; Peeled/Delaminated
INFORMATION WAS RECEIVED BY A MANUFACTURE REPRESENTATIVE (REP) REGARDING AN IMPLANTABLE NEUROSTIMULATOR (INS). THE REPRESENTATIVE STATED THAT LEAD SHEATH SHEARED AT THE DISTAL CONNECTION POINT WITH THE ELECTRODE WHEN CONNECTING IT TO THE PERCUTANEOUS EXTENSION AND SLIDING THE BOOT OVER THE WIRE. THE PRODUCT WAS NEVER IMPLANTED. NO PATIENT ISSUES, STRICTLY DEVICE. LEAD WAS REMOVED AND REPLACED WITH ANOTHER PRODUCT. NO TROUBLESHOOTING POSSIBLE. NO EXTERNAL FACTORS, NORMAL PROCEDURAL STEPS FOLLOWED. THERE WERE NO FURTHER PATIENT COMPLICATIONS ARE ANTICIPATED OR EXPECTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9492275 ↗
FDA clearance and classification
Approved through premarket approval (PMA) P080025 on Mar 14, 2011. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Sacral Nerve Stimulation Therapy System”. The FDA review took 899 days. The FDA has since approved 350 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗
Approved through premarket approval (PMA) P970004 on Sep 29, 1997. Submitted by Medtronic Neuromodulation as “Medtronic Interstim Therapy System For Urinary Control”. The FDA review took 242 days. The FDA has since approved 449 supplements (changes to the device or its labeling), most recently on Oct 29, 2026. FDA record ↗
The FDA regulates product code EZW (stimulator, electrical, implantable, for incontinence) under 21 CFR 876.5270 as a Class III device. 21 CFR 876.5270 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under P080025, P970004. One submission often covers many models, so check the product and lot details in each record.
- Z-1704-2025 · Apr 3, 2025 · Open, ClassifiedEnhanced Verify Evaluation Handset (CFN HH90130FA)
- Z-0372-2025 · Sep 18, 2024 · Open, ClassifiedMedtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
- Z-2365-2023 · Jul 6, 2023 · Open, ClassifiedMedtronic InterStim Ground Pad REF 041826
- Z-0998-2022 · Feb 28, 2022 · TerminatedMEDTRONIC INTERSTIM THERAPY SYSTEM FOR URINARY CONTROL, which includes the following: a) Handset with Communicator, Model Numbers: 1) TH90Q01, 2) TH90QFA, 3) TH90PFA, 4) TH90GFA,
- Z-1147-2015 · Jan 9, 2015 · TerminatedThe Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
- Z-0040-2008 · Jun 25, 2007 · TerminatedMedtronic InterStim INS (Model 3023) with N'Vision Inter Stim-B software supplied on the Model 8870 Version MMB_01/NNB_01 with N'Vision Clinician Programmer Model 8840 or InterStim
- Z-0037-2008 · Jun 25, 2007 · TerminatedMedtronic InterStim II INS Model 3058 with N'Vision InterStim-B Software supplied on the Model 8870 Version MMB_01/NNB_01 N'Vision with a Clinic an Programmer Model 8840. Implanta
Common questions about the INTERSTIM® 3889-28
Has the INTERSTIM® 3889-28 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00643169864436), based on the FDA recall file dated Nov 27, 2026. However, 7 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the INTERSTIM® 3889-28?
The FDA's MAUDE database lists 56 adverse event reports for this device, received between Jun 26, 2018 and Jul 9, 2025. In those reports, the most common specific device problems are “Break” (34), “Peeled/Delaminated” (17) and “Device Damaged Prior to Receipt by User” (7); the most common patient problems are “Therapeutic Effects, Unexpected” (5), “Device Embedded In Tissue or Plaque” (2) and “Apnea” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the INTERSTIM® 3889-28?
None of the reports are classified as deaths, 26 as injuries and 30 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the INTERSTIM® 3889-28 FDA approved?
It was approved through premarket approval (PMA) P970004 on Sep 29, 1997; its record lists 2 FDA submissions in total.
Is the INTERSTIM® 3889-28 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the INTERSTIM® 3889-28, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand INTERSTIM®, model 3889-28). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stimulator, electrical, implantable, for incontinence devices
- INTERSTIM® 3058Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence6 deaths23,076 reports
- InterStim™ Micro SureScan™ 97810Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence2,762 reports
- InterStim™ X 97800Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1,840 reports
- Axonics 4101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence1 death818 reports
- INTERSTIM® 3058Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence818 reports
- Axonics 1101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence1 death399 reports
- Axonics 5101Axonics Modulation Technologies, Inc. · Stimulator, Electrical, Implantable, For Incontinence162 reports
- INTERSTIM® 3023Medtronic, Inc. · Stimulator, electrical, implantable, for incontinence1 death147 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page