Skip to content

Azure™ XT DR MRI SureScan™ W1DR01 (UDI 00643169871939)Recalls & reported problems

Medtronic, Inc. · IPG W1DR01 AZURE XT DR MRI WL USA BCP · Class 3 · Implant

All models of Medtronic Azure Pacemaker (12) · Azure XT DR MRI SureScan (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Azure XT DR MRI SureScan pacemaker is an implanted pacemaker. Its 654 reports mostly describe pacing and sensing problems, the battery draining early, and infections that led to device removal. The reports include 13 deaths and 333 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
654
Deaths reported
13
Injuries reported
333
Recalls naming it
0

Summary

The FDA's MAUDE database contains 654 adverse event reports that list this exact device (DI 00643169871939), received between Apr 2, 2018 and Sep 24, 2026 — classified as 13 deaths, 333 injuries, 308 malfunctions.

The most frequently reported specific device problem is “Pacing Problem” (92 reports).

16 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 3 reportsMar 2024: 5 reportsApr 2024: 2 reportsMay 2024: 4 reportsJun 2024: 4 reportsJul 2024: 6 reportsAug 2024: 2 reportsSep 2024: 4 reportsOct 2024: 5 reportsNov 2024: 3 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 5 reportsAug 2025: 0 reportsSep 2025: 4 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 6 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 5 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem215
  • Pacing Problem92
  • Unable to Obtain Readings71
  • Appropriate Device Problem Term/Code Not Available65
  • Premature Discharge of Battery41
  • Incorrect Measurement40
  • False Alarm36
  • Mechanical Problem24

Patient problems

  • Unspecified Infection159
  • No Clinical Signs, Symptoms or Conditions145
  • No Known Impact Or Consequence To Patient142
  • Bradycardia37
  • No Consequences Or Impact To Patient25
  • Syncope/Fainting21
  • Dizziness15
  • Dyspnea14

Recent adverse event reports

The 25 most recent of 654 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Unable to Obtain Readings

    It was reported that the Right Atrial (RA) lead and Right Ventricular (RV) lead exhibited rising and fluctuating unipolar impedance which appeared to be consistent with a Implantable Pulse Generator (IPG) or measurement-related issue. The IPG system remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50879957 ↗
  • MalfunctionAug 19, 2026Electromagnetic Interference

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50323046 ↗
  • InjuryAug 13, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced a Streptococcus mitis endocarditis infection with vegetation on the Right Atrial (RA) lead. The Implantable Pulse Generator (IPG) system was explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Bacterial Infection; Endocarditis; Unspecified Infection

    MAUDE report 50233420 ↗
  • MalfunctionJun 23, 2026Pacing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PACING SPIKES ON THE TELEMETRY. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE, UNDER SENSING RESULTING IN OVER PACING AND HIGH THRESHOLDS. THE IPG SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25592520 ↗
  • InjuryApr 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) AND RIGHT VENTRICULAR (RV) LEAD WERE REMOVED AND REPLACED. THE RA LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 25015530 ↗
  • MalfunctionApr 21, 2026False Alarm

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING, OVERSENSING, LOSS OF CAPTURE, ATRIAL LEAD POSITION CHECK FAILURE AND HIGH THRESHOLDS. THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A FALSE LOW IMPEDANCE WARNING. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24938827 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 24858561 ↗
  • InjuryApr 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INDUCED LEFT BUNDLE BRANCH BLOCK. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED WITH A CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 24844646 ↗
  • MalfunctionApr 3, 2026Premature Discharge of Battery

    IT WAS REPORTED BY A PATIENT REPRESENTATIVE THAT THE PATIENTS IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PREMATURE BATTERY DEPLETION. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24776241 ↗
  • InjuryFeb 18, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 24375529 ↗
  • InjuryJan 27, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE PATIENT THAT THEY WERE HAVING HEART ISSUES IN THE PAST AND THEY WERE "PASSING OUT". THIS WAS ADDRESSED WITH AN ADJUSTMENT ON THEIR IMPLANTABLE PULSE GENERATOR (IPG). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Syncope/Fainting

    MAUDE report 24188288 ↗
  • InjuryJan 23, 2026Signal Artifact/Noise; Electromagnetic Interference

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS FRACTURE. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT PRIOR TO EXPLANT THE RA LEAD EXHIBITED UNDEFINED AND HIGH IMPEDANCES. NOISE TRIGGERING EPISODE MISCLASSIFICATION WAS OBSERVED. IT WAS FURTHER REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED ELECTROMAGNETIC INTERFERENCE/NOISE. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24160337 ↗
  • InjuryJan 16, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 24093279 ↗
  • InjuryJan 16, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION AND BACTEREMIA. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 24091995 ↗
  • MalfunctionJan 15, 2026Signal Artifact/Noise; Electromagnetic Interference; Pacing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM EXHIBITED ELECTROMAGNETIC INTERFERENCE/NOISE WITH INHIBITION ATRIAL PACING AND FALSE CLASSIFICATION OF AN EPISODE. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24082428 ↗
  • MalfunctionJan 6, 2026Signal Artifact/Noise; Electromagnetic Interference

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED ELECTROMAGNETIC INTERFERENCE AND NOISE. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23974211 ↗
  • MalfunctionDec 12, 2025Battery Problem; Adverse Event Without Identified Device or Use Problem; Battery Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) HIS LEAD EXHIBITED HIGH THRESHOLD VALUES. THE RV HIS LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23788294 ↗
  • InjuryDec 2, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE PATIENT REPORTED THAT WHEN THEY HAD THEIR PACEMAKER FIRST IMPLANTED THEY "EXPERIENCED FAINTING EPISODES". THE PATIENT ALSO REPORTED THAT THEIR MD TOLD THEM THAT THEIR LEADS ARE "OVERLY SENSITIVE" AND THEY WILL ADJUST THE LEADS. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Syncope/Fainting

    MAUDE report 23693016 ↗
  • MalfunctionDec 1, 2025Electromagnetic Interference; Electromagnetic Interference

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED ELECTROMAGNETIC INTERFERENCE. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING , NOISE AND FAR FIELD R-WAVE (FFRW) OVERSENSING. THE IPG SYSTEM REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23680307 ↗
  • InjuryNov 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION AND BACTEREMIA. VEGETATIONS WAS ALSO NOTED ON THE RIGHT VENTRICULAR (RV) L EAD. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 23580687 ↗
  • MalfunctionOct 30, 2025False Alarm; False Alarm

    IT WAS REPORTED THAT THERE WAS A FALSE IMPEDANCE WARNING. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO PATIENT COMPLICAT IONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23432608 ↗
  • MalfunctionOct 28, 2025Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA. IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE LOWER PROGRAMMED LIMIT AND THE LEFT BUNDLE BRANCH (LBB) RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). BOTH THE IPG AND LBB RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bradycardia

    MAUDE report 23401242 ↗
  • InjurySep 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION AND EXPERIENCED DISCOMFORT, PAIN, REDNESS, PURULENT DISCHARGE WHICH LEAD TO POCKET EROSION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort; Erosion; Erythema; Pain; Purulent Discharge; Unspecified Infection

    MAUDE report 23093458 ↗
  • InjurySep 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED EROSION AND INFECTION. THE RIGHT VENTRICULAR (RV) LEAD AND RIGHT ATRIAL (RA) LEADS WERE EXPLANTED AND A TEMPORARY LEAD WAS ADDED TO THE IMPLANTABLE PULSE GENERATOR (IPG) WHICH ACTED AS AN EXTERNAL BRIDGE DEVICE. IT WAS ALSO REPORTED THAT 5 DAYS LATER, THE IPG AND TEMPORARY LEAD WERE EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Erosion; Unspecified Infection

    MAUDE report 23073221 ↗
  • InjurySep 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE ORGANISM WAS NOT IDENTIFIED. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED WITHOUT REPLACEMENT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 23038333 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.

  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-1924-2025 · Apr 29, 2025 · Open, Classified
    Azure S DR MRI SureScan, Product number W3DR01
  • Z-2510-2020 · Jun 4, 2020 · Open, Classified
    Azure S SR MRI, Model Number W3SR01
  • Z-2507-2020 · Jun 4, 2020 · Open, Classified
    Azure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
  • Z-2508-2020 · Jun 4, 2020 · Open, Classified
    Azure S DR MRI, Model Number W3DR01
  • Z-2509-2020 · Jun 4, 2020 · Open, Classified
    Azure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
  • Z-0809-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
  • Z-0807-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Numb
  • Z-0811-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
  • Z-0808-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
  • Z-0806-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (
  • Z-0810-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
  • Z-2120-2011 · Apr 4, 2011 · Terminated
    Medtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
  • Z-1440-2010 · Feb 11, 2010 · Terminated
    Medtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,

…and 1 more.

Other Implantable pulse generator, pacemaker (non-CRT) devices

All Implantable pulse generator, pacemaker (non-CRT) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page