Viva™ XT CRT-D DTBA1D1 (UDI 00643169929296)Recalls & reported problems

Medtronic, Inc. · CRT-D DTBA1D1 VIVA XT US DF1 · Class 3 · Implant

All models of Viva XT CRT-D (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

FDA's MAUDE database lists 65 adverse event reports for the Medtronic Viva XT CRT-D DTBA1D1, an implantable defibrillator that also delivers cardiac resynchronization therapy. Most describe early battery depletion, sensing problems, and inappropriate shocks or pacing; 41 injuries and 1 death are reported. One recall, begun in December 2017 over a design issue that could prevent high or low voltage therapy, was terminated.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
65
Deaths reported
1
Injuries reported
41
Recalls naming it
1

Summary

The FDA's MAUDE database contains 65 adverse event reports that list this exact device (DI 00643169929296), received between Oct 6, 2017 and Feb 5, 2026 — classified as 1 death, 41 injuries, 23 malfunctions.

The most frequently reported specific device problem is “Premature Discharge of Battery” (16 reports).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 19, 2017Class ITerminated

    Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).

    Z-0584-2018 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem17
  • Premature Discharge of Battery16
  • Device Sensing Problem10
  • Inappropriate/Inadequate Shock/Stimulation10
  • Pacing Problem7
  • Electrical /Electronic Property Problem6
  • Electromagnetic Interference4
  • Mechanical Problem4

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Unspecified Infection14
  • No Known Impact Or Consequence To Patient13
  • Shock from Patient Lead(s)10
  • No Patient Involvement4
  • Arrhythmia2
  • Hematoma2
  • Insufficient Information2

Recent adverse event reports

The 25 most recent of 65 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 5, 2026Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED BY THE HEALTH CARE PROFESSIONAL (HCP) THAT THE PATIENT RECEIVED INAPPROPRIATE SHOCKS, POSSIBLY DUE TO A DEVICE DETECTION ISSUE. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24269275 ↗
  • InjuryDec 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 23829997 ↗
  • InjuryOct 23, 2024Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE HIGH VOLTAGE THERAPY FOR SUSPECTED ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT WAS DETECTED AS VENTRICULAR TACHYCARDIA (VT). THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 20516737 ↗
  • InjuryJul 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 19732183 ↗
  • InjuryJun 4, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT WHEN THE PATIENT CAME IN FOR A GENERATOR CHANGE FOR REACHING RECOMMENDED REPLACEMENT TIME (RRT), THE PHYSICIAN NOTICED DISCHARGE AND DISCOLORATION INSIDE THE POCKET. THE PATIENT DID NOT HAVE ANY SYMPTOMS AND THERE WAS NO INDICATION OF AN INFECTION AT SIGHT. THE DEVICE WAS PLACED BACK IN POCKET AND SUTURED CLOSED. CULTURES WERE TAKEN. THE PATIENT WAS SENT TO ANOTHER FACILITY FOR EXTRACTION OF THE ENTIRE SYSTEM. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS PACEMAKER DUE TO THE INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Purulent Discharge; Unspecified Infection

    MAUDE report 19458339 ↗
  • InjuryMay 8, 2024Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE ANTI-TACHYCARDIA PACING AND HIGH VOLTAGE THERAPY FROM THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT WAS DETECTED AS VENTRICULAR TACHYCARDIA (VT) AND VENTRICULAR FIBRILLATION (VF). THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE CRT-D AND RA LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 19270702 ↗
  • InjuryDec 22, 2023Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED EARLY BATTERY DEPLETION. THE CRTD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18384346 ↗
  • MalfunctionDec 15, 2023Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT DURING USE OF THE LEFT VENTRICULAR (LV) LEAD IT WAS NOTED THE LEAD EXHIBITED VARIABLE LVTIP TO RV (RIGHT VENTRICULAR) COIL, LEAD IMPEDANCE. THE LV LEAD REMAINS IN USE.  ADDITIONALLY THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED VARIABLE R-WAVE AMPLITUDE.  THE RV LEAD REMAINS IN USE.  IT WAS ALSO REPORTED THAT THE PATIENT WAS HAVING A LOT OF ECTOPY AND PREMATURE VENTRICULAR CONTRACTIONS (PVC) AND THERE WAS CONCERN REGARDING THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D), VENTRICULAR SENSE RESPONSE PACING. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18334812 ↗
  • InjuryMar 2, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 16469486 ↗
  • MalfunctionJan 26, 2023Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THERE WAS CONCERNED REGARDING A DROP IN LONGEVITY ON THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16248622 ↗
  • MalfunctionJan 25, 2023Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED LOSS OF CAPTURE AT TIMES. THE LV LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT THE PATIENT HEART RATES ARE BELOW THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) PROGRAMMED LOWER RATE OF 50 BEATS PER MINUTE (BPM). THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED, AS A RESULT OF THIS EVENT.

    Patient problems: Ectopic Heartbeat; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16244247 ↗
  • InjuryDec 15, 2022Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE SHOCKS FROM THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT THE DEVICE DETECTED AS VENTRICULAR FIBRILLATION (VF). THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. IT WAS NOTED THAT THE CRT-D WAS IN MODE SWITCH AT THE TIME OF THE DETECTION. THE CRT-D AND RA LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 15991972 ↗
  • DeathNov 22, 2022Failure to Deliver Shock/Stimulation; Battery Problem; Failure to Deliver Shock/Stimulation; Battery Problem

    IT WAS REPORTED BY THE PATIENT'S SPOUSE THAT THE PATIENT'S CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DID NOT DELIVER A SHOCK WHEN THE PATIENT EXHIBITED AN IRREGULAR HEARTBEAT. THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Arrhythmia

    MAUDE report 15844491 ↗
  • InjurySep 20, 2022Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT POTENTIALLY RECEIVED INAPPROPRIATE THERAPY FROM THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE THAT WAS DETECTED AS FAST VENTRICULAR TACHYCARDIA. THERE WAS A POSSIBLE CAPTURE ISSUE NOTED ON THE RIGHT VENTRICULAR (RV) LEAD AS THE CAPTURE THRESHOLD WAS HIGHER THAN THE PROGRAMMED OUTPUT. THE RIGHT ATRIAL (RA) LEAD EXHIBITED INTERMITTENT UNDERSENSING. THE CRT-D, RV LEAD, AND RA LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 15455614 ↗
  • InjuryAug 12, 2022Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION. THE CRT-D WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15221699 ↗
  • MalfunctionJul 21, 2022Electromagnetic Interference; Electromagnetic Interference

    IT WAS REPORTED THAT ATRIAL FIBRILLATION (AF) ELECTROGRAMS (EGM) SHOWED NOISE ON THE RIGHT ATRIAL (RA) LEAD THAT WAS MOST LIKELY ELECTROMAGNETIC INTERFERENCE. THE DEVICE REMAINS IN USE. THE PATIENT IS A PARTICIPANT IN A CLINICAL STUDY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15075468 ↗
  • InjuryFeb 3, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 13446501 ↗
  • InjuryDec 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO INFECTION. IT WAS ALSO NOTED THAT DURING THE REMOVAL OF THE LEFT VENTRICULAR (LV) LEAD IT "SNAPPED" AND WAS ONLY PARTIALLY EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 13011255 ↗
  • InjuryNov 12, 2021Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing; Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing

    SENSITIVITY WAS DECREASED AND THE DEVICE REMAINS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12804585 ↗
  • InjuryNov 12, 2021Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION LIKELY DUE TO HIGH LEFT VENTRICULAR (LV) LEAD THRESHOLDS. THE DEVICE WAS EXPLANTED AND REPLACED, AND THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12804549 ↗
  • InjuryOct 19, 2021Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED PREMATURE BATTERY DEPLETION DUE TO HIGH RIGHT VENTRICULAR (RV) LEAD THRESHOLDS. THE DEVICE WAS EXPLANTED AND REPLACED AND THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12656976 ↗
  • MalfunctionOct 13, 2021Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVER SENSING WHICH LEAD TO INHIBITION OF PACING BY THE CARDIAC RESYNC HRONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE LEAD AND DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 12628845 ↗
  • InjuryAug 3, 2021Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    ADDITIONAL INFORMATION RECEIVED INDICATED THE PATIENT HAD BEEN ADMITTED TO THE HOSPITAL WITH VENTRICULAR TACHYCARDIA (VT) STORM, AND REPORTED THAT THE SHOCKS WERE APPROPRIATE.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 12266697 ↗
  • InjuryJun 4, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A POCKET REVISION PROCEDURE SIGNS OF AN INFECTION WERE NOTED. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS REMOVED DUE TO THE INFECTION. WHILE REMOVING THE LEFT VENTRICULAR (LV) LEAD THE PATIENT EXPERIENCED A PERFORATION. THE PATIENT SUBSEQUENTLY DEVELOPED TAMPONADE AND PLEURAL EFFUSION. A STERNOTOMY WAS PERFORMED AND THE PATIENT WAS PUT ON BYPASS AND HAD TO RECEIVED UNITS OF BLOOD. THE PATIENT WAS IN CRITICAL CONDITION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 11939271 ↗
  • InjuryJun 2, 2021Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) HAD EARLY BATTERY DEPLETION. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11921278 ↗

Other Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices

All Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D) devices →