EverCross™ AB35W07080080 (UDI 00643169974135)Recalls & reported problems
Ev3, Inc · PTA AB35W07080080 EVERCROSS 035 V08 · Class 2
All models of EverCross (114)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169974135), received between Feb 25, 2021 and Oct 5, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Burst Container or Vessel2
- Detachment of Device or Device Component1
- Difficult to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Unspecified Tissue Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 5, 2021Burst Container or Vessel; Detachment of Device or Device Component; Difficult to Advance; Burst Container or Vessel; Detachment of Device or Device Component; Difficult to Advance
PHYSICIAN WAS ATTEMPTING TO USE AN EVERCROSS PTA BALLOON DURING PROCEDURE TO TREAT A CALCIFIED LESION IN THE PATIENT¿S PROXIMAL AND MID LEFT COMMON ILIAC ARTERY. SLIGHT VESSEL TORTUOSITY AND SEVERE CALCIFICATION ARE REPORTED. AN INFLATION DEVICE WAS USED FOR BALLOON INFLATION. THERE WAS NO DAMAGE NOTED TO PACKAGING. THERE WAS NO ISSUE NOTED WHEN REMOVING THE DEVICE FROM HOOP/TRAY. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES IDENTIFIED. THE DEVICE WAS PASSED THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOTED DURING ADVANCEMENT, BUT NO EXCESSIVE FORCE WAS USED. IT IS REPORTED THAT AFTER BALLOON INFLATION THE PHYSICIAN REALIZED THE BALLOON HAD RUPTURED. WHEN PHYSICIAN TRIED TO REMOVE THE BALLOON, HE FELT THE BALLOON BREAK. THE PHYSICIAN WAS USING A NON-MEDTRONIC GUIDEWIRE AT THE TIME. THE PROXIMAL PART OF THE BALLOON AND THE PROXIMAL AND DISTAL BALLOON MARKERS DID NOT COME OUT WHEN BALLOON WAS RETRIEVED. A SECOND .035 WIRE WAS INSERTED, AND A SELF-EXPANDING STENT WAS ADVANCED TO JAIL OFF THE BALLOON FRAGMENT. TWO STENTS WERE USED TO COMPLETELY JAIL OFF THE BALLOON AND MARKERS. A SHEATH WAS USED TO GET UP AND OVER AND TREAT THE OTHER LESIONS IN THE POPLITEAL (POP) AND SUPERFIC
Patient problems: Unspecified Tissue Injury
MAUDE report 12575913 ↗MalfunctionFeb 25, 2021Burst Container or Vessel; Burst Container or Vessel
PHYSICIAN ATTEMPTED TO USE AN EVERCROSS PTA BALLOON WITH A 7FR SHEATH AND 0.035 NON-MEDTRONIC GUIDEWIRE DURING TREATMENT OF AN AV FISTULA. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING PRIOR TO USE. NO ISSUES WERE NOTED DURING REMOVAL OF THE DEVICE FROM THE PACKAGING. SALINE AND CONTRAST WERE USED FOR BALLOON INFLATION. A BALLOON BURST IS REPORTED TO HAVE OCCURRED AT INFLATION TO APPROXIMATELY 14ATM. THE BALLOON WAS REMOVED, AND ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11377355 ↗
Other Catheter, angioplasty, peripheral, transluminal devices
- Coyote™ ES H74939135152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal879 reports
- Coyote™ ES H74939135204010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal585 reports
- Coyote™ ES H74939135202010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal437 reports
- Coyote™ ES H74939135304010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal384 reports
- Coyote™ ES H74939135404010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal293 reports
- Coyote™ ES H74939134152010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal279 reports
- Coyote™ H74939185202210Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal200 reports
- Coyote™ ES H74939135254010Boston Scientific Corporation · Catheter, angioplasty, peripheral, transluminal127 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗