CapSure®SP Novus 4592-53Recalls & reported problems
Medtronic, Inc. · LEAD 4592-53 BI HYBR CAPSURE US PAS 53CM · Class 3 · Implant
All models of CapSureSP Novus (14)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic CapSure SP Novus pacemaker lead is an implanted wire for a pacemaker. Of its 164 adverse event reports, 63 are injuries and 100 are malfunctions, with 1 death. Many describe the lead sensing heartbeats wrongly, and some describe pocket infections and erosion. Reports fell from 18 in the prior 12 months to 5 in the last 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 165
- Deaths reported
- 1
- Injuries reported
- 64
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 165 adverse event reports that list this exact device (DI 00681490114974), received between Apr 6, 2017 and Jul 29, 2026 — classified as 1 death, 64 injuries, 100 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (63 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing63
- Under-Sensing49
- Signal Artifact/Noise24
- Adverse Event Without Identified Device or Use Problem21
- High impedance17
- Fracture15
- Electrical /Electronic Property Problem12
- Low impedance11
Patient problems
- No Clinical Signs, Symptoms or Conditions89
- No Known Impact Or Consequence To Patient35
- Unspecified Infection15
- No Consequences Or Impact To Patient4
- Erosion3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Bradycardia2
- Chest Pain2
Recent adverse event reports
The 25 most recent of 165 reports, in the reporter's own words as published by the FDA.
InjuryJul 29, 2026High Capture Threshold
It was reported that the patient experienced syncope. It was further reported that the Right Atrial (RA) lead exhibited high thresholds. The lead was reprogrammed and remains in use. No patient complications have been reported as a result of this event.
Patient problems: Syncope/Fainting
MAUDE report 50075296 ↗MalfunctionJun 29, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE RIGHT ATRIAL (RA) LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25653619 ↗MalfunctionMay 22, 2026Signal Artifact/Noise; Electrical /Electronic Property Problem; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING RESULTING IN MISCLASSIFICATION OF EPISODE TYPES. IT WAS ALSO REPORTED THAT THE RA LEAD EXHIBITED NOISE AND AN ATRIAL LEAD POSITION CHECK FAIL. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25252814 ↗MalfunctionApr 17, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24909927 ↗InjuryApr 16, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A POCKET INFECTION AND EROSION. THE PATIENT WAS TREATED WITH ANTIBIOTICS. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erosion; Medical device site infection
MAUDE report 24896258 ↗MalfunctionJan 21, 2026Electrical /Electronic Property Problem; Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING, INAPPROPRIATE PACING AND AN ATRIAL LEAD POSITION CHECK(ALPC) FAILURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24141181 ↗MalfunctionAug 22, 2025Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT IRREGULAR OVERSENSING AND NOISE WAS DETECTED ON THE ATRIAL LEAD DURING MONITORED ATRIAL TACHYCARDIA/ATRIAL FIBR ILLATION EPISODES. POCKET MANIPULATIONS COULD NOT REPRODUCE THE SENSING ISSUE, HOWEVER, INTERFERENCE WAS OBSERVED WITH THE PATIENT PRESSING THEIR HANDS TOGETHER. THE PATIENT IS BEING MONITORED WITH REGULAR FOLLOW-UP. THE RIGHT ATRIAL (RA) LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22872450 ↗InjuryAug 20, 2025Signal Artifact/Noise; Electrical /Electronic Property Problem; Over-Sensing; Pacing Problem; Signal Artifact/Noise; Electrical /Electronic Property Problem; Over-Sensing; Pacing Problem
IT WAS FURTHER REPORTED THAT THE RA LEAD CONTINUED TO EXHIBIT OVERSENSING AND NOISE AND HAD ANOTHER FAILED ATRIAL LEAD POSITION CHECK.
Patient problems: Syncope/Fainting
MAUDE report 22851838 ↗InjuryJun 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 22319186 ↗InjuryJun 13, 2025Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Impedance Problem; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Impedance Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD FRACTURED RESULTED IN OVERSENSING OF NOISE DURING STORED EPISODES. HIGH AND RISING IMPEDANCE WAS NOTED AS WELL. THE RA LEAD WAS REPROGRAMMED OFF. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22201610 ↗InjuryMay 21, 2025High impedance; Over-Sensing; High impedance; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH IMPEDANCE AND OVERSENSING. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22059423 ↗MalfunctionMay 12, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING ON STORED OR CURRENT ELECTROGRAMS (EGM). DOES NOT APPEAR TO BE TRUE ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF)EVENT. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22000678 ↗InjuryApr 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS REMOVED WITHOUT REPLACEMENT DUE TO INFECTION. THE ORGANISM WAS IDENTIFIED AS BACTEREMIA. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia
MAUDE report 21907059 ↗InjuryMar 28, 2025Signal Artifact/Noise; Fracture; High impedance; Signal Artifact/Noise; Fracture; High impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH IMPEDANCE AND NOISE. IT WAS NOTED THE LEAD HAD A POSSIBLE FRACTURE ALSO. IT WAS FURTHER NOTED THE LEAD WAS REPROGRAMMED. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21717948 ↗MalfunctionMar 12, 2025Impedance Problem; Impedance Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN ABRUPT CHANGE IN ATRIAL LEAD IMPEDANCE MEASUREMENTS NOTED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21582854 ↗InjuryFeb 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED EROSION AND WOUND DEHISCENCE AND THE DEVICE MAIN UNIT WAS EXPOSED IN THE POCKET. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONICALLY ELEVATED THRESHOLDS. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM REMAINS IN USE. SURGICAL INTERVENTION WAS REQUIRED DUE TO THE EVENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Erosion; Wound Dehiscence
MAUDE report 21344363 ↗InjuryFeb 4, 2025Signal Artifact/Noise; Electrical /Electronic Property Problem; Signal Artifact/Noise; Electrical /Electronic Property Problem
IT WAS REPORTED THAT NOISE WAS OBSERVED ON THE RIGHT ATRIAL (RA) LEAD AND THE RA LEAD EXHIBITED AN ATRIAL LEAD POSITION CHECK FAILURE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21303039 ↗InjuryJan 13, 2025Break; Break
IT WAS REPORTED THAT THE INSULATION OF THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD WAS DAMAGED DURING IMPLANTABLE PULSE GENERATOR (IPG) CHANGE. THE RA LEAD AND RV LEAD WERE KINKED AND CUT. THE RA LEAD AND RV LEAD WERE CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21134970 ↗InjuryJan 8, 2025Fracture; Fracture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A CONFIRMED FRACTURE. THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21103157 ↗MalfunctionDec 5, 2024Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING ON STORED ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20851272 ↗MalfunctionDec 2, 2024Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED INTERMITTENT OVERSENSING NOTED ON STORED ELECTROGRAMS (EGM). THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20821311 ↗MalfunctionNov 22, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A POLARITY SWITCH. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20756401 ↗InjuryNov 11, 2024Mechanical Problem; Over-Sensing; Mechanical Problem; Over-Sensing
IT WAS REPORTED THAT POST IMPLANTABLE PULSE GENERATOR (IPG) CHANGEOUT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING. THE LEAD WAS REPROGRAMMED. APPROXIMATELY TWO WEEKS LATER, A SHOULDER ROTATION AND A STRESS TEST ON THE POCKET SITE WERE PERFORMED IN ORDER TO CHECK THE LEAD CONDITION AGAIN AT THE OUTPATIENT VISIT AND FURTHER OVERSENSING OCCURRED. ALSO, NO LOOSE PIN WAS OBSERVED IN THE PHOTOS WHEN CHECKING THE A CHEST X-RAY HOWEVER, A LOOSE PIN COULD NOT BE RULED OUT. THE SURGEON DECIDED NOT TO CHANGE THE SETTINGS AND CONTINUED TO FOLLOW-UP WITH THE LEAD REMAINING IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20650618 ↗MalfunctionNov 1, 2024Signal Artifact/Noise; Electrical /Electronic Property Problem; Over-Sensing; Signal Artifact/Noise; Electrical /Electronic Property Problem; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED NOISE ON ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODES WITH MIRROR DEFLECTIONS OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. IN ADDITION, THE ATRIAL LEAD POSITION CHECK FAILED. THE RA LEAD AND THE RV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20587071 ↗MalfunctionJul 16, 2024Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THERE WAS INTERMITTENT ATRIAL OVERSENSING OBSERVED ON STORED ELECTROGRAMS. THE RIGHT ATRIAL (RA) LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19751132 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page