Capsure®Z Novus 5054-52Recalls & reported problems
Medtronic, Inc. · LEAD 5054-52 CAPSUREZ NOVUS US EN · Class 3 · Implant
All models of CapSureZ Novus (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Capsure Z Novus is a pacemaker lead that carries electrical pulses between the heart and the device. The 308 reports mostly describe high pacing thresholds, high impedance, and noise or oversensing on the lead, with some reports of dizziness and slow heart rhythm. There are 6 deaths and 189 injuries reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 309
- Deaths reported
- 6
- Injuries reported
- 189
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 309 adverse event reports that list this exact device (DI 00681490123167), received between Feb 9, 2017 and Sep 14, 2026 — classified as 6 deaths, 189 injuries, 114 malfunctions.
The most frequently reported specific device problem is “High Capture Threshold” (105 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Capture Threshold105
- High impedance63
- Over-Sensing50
- Failure to Capture47
- Impedance Problem37
- Signal Artifact/Noise30
- Unstable Capture Threshold28
- Capturing Problem27
Patient problems
- No Clinical Signs, Symptoms or Conditions127
- No Known Impact Or Consequence To Patient70
- Bradycardia21
- Dizziness18
- No Consequences Or Impact To Patient12
- Syncope/Fainting12
- Unspecified Infection12
- Chest Pain9
Recent adverse event reports
The 25 most recent of 309 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 14, 2026High Capture Threshold
It was reported that the Right Ventricular (RV) lead exhibited high thresholds. The RV lead remains in use. It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times and resulting in termination of AT/AF detected event(s) in progress. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50751018 ↗InjurySep 3, 2026Signal Artifact/Noise; Failure to Capture; Over-Sensing; Impedance Problem
It was reported that the patient experienced syncope and Right Ventricular (RV) lead exhibited loss of capture, non-physiologic inte rvals, and noise. The lead remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Syncope/Fainting
MAUDE report 50583205 ↗MalfunctionJul 10, 2026Break; Over-Sensing; Unstable Capture Threshold
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25779207 ↗MalfunctionJun 23, 2026Failure to Capture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING, OVERSENSING AND OVERSENSING RESULTING IN MISCLASSIFICATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF). IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOSS OF CAPTURE. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25595133 ↗MalfunctionJun 18, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING . THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25543006 ↗MalfunctionMay 20, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED CHRONIC BIPOLAR AND UNIPOLAR LOW OUT OF RANGE IMPEDANCE WARNINGS ALONG WITH NOISE AND OVERSENSING ON STORED EPISODES RESULTING IN FALSE EPISODE DETECTIONS. ATRIAL LEAD POSITION CHECK FAILURE WAS OBSERVED AS WELL. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONIC HIGH THRESHOLDS. BOTH THE RV AND RA LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25226772 ↗MalfunctionApr 29, 2026Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT THERE WAS A HIGH V-V COUNT ON THE SENSING INTEGRITY COUNTER (SIC), NOISE AND OVERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25031865 ↗InjuryApr 3, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 24773676 ↗MalfunctionMar 31, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED SHORT V-V INTERVALS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24740072 ↗MalfunctionMar 20, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE PATIENT EXPERIENCED LETHARGY. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDER SENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS RESULTING IN UNNECESSARY PACING AT TIMES AND FALSE TERMINATION OF EPISODE. ADDITIONALLY, RIGHT VENTRICULAR (RV) LEAD DISPLAYED UNDER SENSING RESULTING IN INAPPROPRIATE PACING INTO VENTRICULAR TACHYCARDIA (VT) EPISODE OR T-WAVE. BOTH THE LEADS REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Lethargy
MAUDE report 24647580 ↗MalfunctionMar 18, 2026Impedance Problem; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN INCREASE IN IMPEDANCE AND VARIABLE THRESHOLDS. THE RA AND RV LEADS BOTH REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24634708 ↗MalfunctionMar 5, 2026Signal Artifact/Noise; Fracture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD HAD SUSPECTED FRACTURES AND EXHIBITED NOISE. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24518466 ↗MalfunctionJan 29, 2026Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED SLIGHT VARIABILITY IN LEAD TREND PARAMETERS, IT WAS NOTED THERE WERE "QUES TIONABLE" ATRIAL INTERVALS. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING WITH SUSPECTED LEAD NOISE. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24214916 ↗MalfunctionJan 28, 2026Failure to Capture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING RESULTED IN UNDER PACING ALONG WITH THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NON-CAPTURE. BOTH THE RA AND RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24197200 ↗MalfunctionJan 19, 2026Over-Sensing; Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING AND UNDERSENSING RESULTING IN INAPPROPRIATE PACING. THE R IGHT ATRIAL (RA) LEAD EXHIBITED AN OUT OF RANGE (OOR) LOW IMPEDANCE WARNING AND UNDERSENSING DURING ATRIAL TACHYCARDIA/ ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24108564 ↗MalfunctionJan 12, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT UNDERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24034128 ↗InjuryNov 21, 2025Intermittent Capture; Unstable Capture Threshold; Intermittent Capture; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNSTABLE THRESHOLDS WITH INTERMITTENT CAPTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23615823 ↗InjuryNov 4, 2025Over-Sensing; Low impedance; Device Sensing Problem; Unstable Capture Threshold; Over-Sensing; Low impedance; Device Sensing Problem; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW IMPEDANCE AND UNSTABLE THRESHOLDS. THE LEAD WAS EXPLANTED AND REP LACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23463148 ↗InjuryAug 11, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS CAUSING EARLY BATTERY DEPLETION OF THE IMPLANTABLE PULSE GENERATOR (IPG). THE RV LEAD WAS CAPPED AND REPLACED. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22770818 ↗MalfunctionJul 16, 2025High impedance; High impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR OUT OF RANGE (OOR) HIGH IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22517803 ↗InjuryJul 11, 2025Fracture; High impedance; Fracture; High impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDEFINED HIGH IMPEDANCE AND THE LEAD WAS NOTED TO BE FRACTURED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22478896 ↗InjuryJul 8, 2025Pacing Problem; Extracardiac stimulation; Impedance Problem; Adverse Event Without Identified Device or Use Problem; High Capture Threshold; Pacing Problem; Extracardiac stimulation; Impedance Problem; Adverse Event Without Identified Device or Use Problem; High Capture Threshold
IT WAS FURTHER REPORTED THAT THERE WAS NO PERFORMANCE ISSUE WITH THE IPG. THE RV LEAD OUTPUTS WERE REPROGRAMMED. IT WAS ALSO NOTED THAT THE IMPEDANCE HAD STABILIZED.
Patient problems: Device Overstimulation of Tissue; Discomfort; No Clinical Signs, Symptoms or Conditions
MAUDE report 22434446 ↗InjuryJul 4, 2025Break; Failure to Capture; Electrical /Electronic Property Problem; Fracture; High impedance; Impedance Problem; High Capture Threshold; Break; Failure to Capture; Electrical /Electronic Property Problem; Fracture; High impedance; Impedance Problem; High Capture Threshold
IT WAS REPORTED THAT THE PATIENT BECAME SYMPTOMATIC WITH LOSS OF CAPTURE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED SUSPECTED FRACTURE, HIGH IMPEDANCE, POLARITY SWITCH, RISING IMPEDANCE, AND HIGH THRESHOLDS. IT WAS NOTED THAT THE LEAD EXHIBITED INSULATION BREACH AT THE LEAD TIP OR SEVERE ENCAPSULATION/MINERALIZATION OF THE LEAD TIP. THE LEAD WAS REPROGRAMMED AND LATER EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22403400 ↗MalfunctionJun 18, 2025Electrical /Electronic Property Problem; High impedance; Impedance Problem; Electrical /Electronic Property Problem; High impedance; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RISING AND HIGH BIPOLAR AND UNIPOLAR IMPEDANCE AND A POLARITY SWITCH. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22271237 ↗InjuryJun 13, 2025Impedance Problem; Impedance Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. THE RIGHT VENTRICULAR (RV) LEAD HAD VARYING IMPEDANCE. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 22201138 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page