CapSureZ NovusRecalls & reported problems across all models
Medtronic, Inc. · 4 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1,202
- Deaths reported
- 12
- Injuries reported
- 623
- Recalls
- 0
We found 4 models of CapSureZ Novus by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 1,202 adverse event reports in the FDA's MAUDE database, received between Apr 10, 2013 and Sep 24, 2026 — classified as 12 deaths, 623 injuries, 567 malfunctions.
The most frequently reported specific device problem is “High Capture Threshold” (299 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CapSureZ Novus problems
Device problems
- High Capture Threshold299
- Under-Sensing257
- Over-Sensing252
- High impedance177
- Signal Artifact/Noise148
- Failure to Capture132
- Impedance Problem103
- Adverse Event Without Identified Device or Use Problem92
Patient problems
- No Clinical Signs, Symptoms or Conditions560
- No Known Impact Or Consequence To Patient312
- Unspecified Infection57
- No Consequences Or Impact To Patient51
- Bradycardia40
- Dizziness37
- Chest Pain25
- Syncope/Fainting25
CapSureZ Novus models
4 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CapSure®Z Novus 5554-45Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes1 death320 reports
- Capsure®Z Novus 5054-52permanent pacemaker Electrode6 deaths309 reports
- CapSure®Z Novus 5554-53permanent pacemaker Electrode3 deaths300 reports
- Capsure®Z Novus 5054-58Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes2 deaths273 reports
Recent CapSureZ Novus adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
DeathSep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported by the patients representative that the patient implanted with a Cardiac Resynchronization Therapy Defibrillator (CRT-D) system "had a sepsis infection from "[the]" defibrillator ports" and "there was an infection on the lead from "[the]" defibrillator." The patient¿s representative reported the patient was "put on antibiotics instead of replacing as usually done." The patient expired approximately 8 weeks later.
Patient problems: Sepsis; Unspecified Infection
MAUDE report 50883752 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50811348 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50811208 ↗MalfunctionSep 14, 2026High Capture Threshold
It was reported that the Right Ventricular (RV) lead exhibited high thresholds. The RV lead remains in use. It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times and resulting in termination of AT/AF detected event(s) in progress. The RA lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50751018 ↗MalfunctionSep 9, 2026Low impedance
It was reported that the Right Atrial (RA) lead triggered an alert for out of range low pacing impedance. The lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50686907 ↗InjurySep 8, 2026High impedance
It was reported that the patient experienced lethargy. The Right Atrial (RA) lead exhibited high impedance. The Implantable Pulse Generator (IPG) had reached its End of Service (EOS). The RA lead was capped and replaced. No further patient complications have been reported as a result of this event.
Patient problems: Lethargy
MAUDE report 50663384 ↗InjurySep 3, 2026Signal Artifact/Noise; Failure to Capture; Over-Sensing; Impedance Problem
It was reported that the patient experienced syncope and Right Ventricular (RV) lead exhibited loss of capture, non-physiologic inte rvals, and noise. The lead remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Syncope/Fainting
MAUDE report 50583205 ↗InjuryAug 27, 2026Defibrillation/Stimulation Problem; Under-Sensing
It was reported that the patient experienced shortness of breath and the Right Ventricular (RV) lead exhibited undersensing resulting in delay of an episode and early termination. The lead remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Dyspnea
MAUDE report 50478809 ↗MalfunctionAug 26, 2026High Capture Threshold
It was reported that the patient experienced chest pain. Upon device assessment it was noted that the Right Ventricular (RV) lead exhibited under sensing resulting in inappropriate pacing. It was also reported that the Right Atrial (RA) lead exhibited high thresholds. Both leads remain in use. No further patient complications have been reported as a result of this event.
Patient problems: Chest Pain
MAUDE report 50449556 ↗InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced bacteremia. The Implantable Pulse Generator (IPG) system and previously capped Right Ventricular (RV) lead were explanted and replaced with a leadless IPG. No further patient complications have been reported as a result of this event.
Patient problems: Bacteremia
MAUDE report 50432541 ↗MalfunctionAug 10, 2026High impedance; High Capture Threshold
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50191466 ↗InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced a suspected pocket infection. The Implantable Pulse Generator (IPG) system was explante d. No further patient complications have been reported as a result of this event.
Patient problems: Medical device site infection
MAUDE report 50142970 ↗MalfunctionJul 10, 2026Break; Over-Sensing; Unstable Capture Threshold
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25779207 ↗InjuryJul 6, 2026High impedance
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A DIAGNOSTIC RESET. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH UNDEFINED IMPEDANCE. THE IPG AND RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25716329 ↗MalfunctionJul 6, 2026Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING/NOISE, RESULTING IN MISCLASSIFICATION OF EPISODE TYPE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25714097 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.