Medtronic, Inc. · LEAD 5054-58 CAPSUREZ NOVUS USEN · Class 3 · Implant

All models of CapSureZ Novus (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Capsure Z Novus pacing lead has 272 reports, including 2 deaths and 180 injuries. Most describe the lead needing more electrical energy to pace the heart, plus readings that drift or misread heart signals. Some describe shortness of breath, dizziness or near-fainting in the patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
273
Deaths reported
2
Injuries reported
181
Recalls naming it
0

Summary

The FDA's MAUDE database contains 273 adverse event reports that list this exact device (DI 00681490123228), received between Apr 10, 2013 and Sep 17, 2026 — classified as 2 deaths, 181 injuries, 90 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (91 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 6 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 4 reportsMar 2024: 4 reportsApr 2024: 4 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 3 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 2 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 3 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 2 reportsNov 2025: 4 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold91
  • High impedance50
  • Over-Sensing49
  • Failure to Capture45
  • Impedance Problem35
  • Signal Artifact/Noise29
  • Under-Sensing28
  • Capturing Problem24

Patient problems

  • No Clinical Signs, Symptoms or Conditions126
  • No Known Impact Or Consequence To Patient63
  • Unspecified Infection12
  • Bradycardia11
  • Dizziness11
  • No Consequences Or Impact To Patient9
  • Syncope8
  • Syncope/Fainting8

Recent adverse event reports

The 25 most recent of 273 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50811208 ↗
  • InjuryAug 27, 2026Defibrillation/Stimulation Problem; Under-Sensing

    It was reported that the patient experienced shortness of breath and the Right Ventricular (RV) lead exhibited undersensing resulting in delay of an episode and early termination. The lead remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Dyspnea

    MAUDE report 50478809 ↗
  • InjuryJul 6, 2026High impedance

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A DIAGNOSTIC RESET. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH UNDEFINED IMPEDANCE. THE IPG AND RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25716329 ↗
  • MalfunctionJun 9, 2026Electrical /Electronic Property Problem; High impedance; Impedance Problem; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A POLARITY SWITCH, HIGH UNDEFINED IMPEDANCE AND HIGH THRESHOLDS. A GRADUAL RISE IN IMPEDANCE WAS NOTED. LEAD MINERALIZATION WAS SUSPECTED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25443888 ↗
  • MalfunctionMay 8, 2026High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONIC HIGH THRESHOLDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25121413 ↗
  • InjuryApr 23, 2026Intermittent Capture

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PRE-SYNCOPE. IT WAS FURTHER NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT CAPTURE. THE RV LEAD WAS CAPPED AND REPLACED WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Presyncope

    MAUDE report 24962766 ↗
  • MalfunctionMar 30, 2026High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24726363 ↗
  • MalfunctionMar 11, 2026Capturing Problem; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN INCREASE IN IMPEDANCE. IT WAS FURTHER NOTED THAT THE RV LEAD EXHIBITED AN INCREASE IN THRESHOLDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24569393 ↗
  • MalfunctionMar 9, 2026Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING IN NON-PHYSIOLOGIC INTERVALS ON STORED EPISODES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24543572 ↗
  • InjuryJan 22, 2026High Capture Threshold

    IT WAS FURTHER REPORTED THAT PACEMAKER MEDIATED TACHYCARDIA WAS SUSPECTED. IT WAS FURTHER NOTED THAT THE RA AND RV LEAD WAS REPROGRA MMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Tachycardia

    MAUDE report 24146390 ↗
  • InjuryJan 13, 2026Capturing Problem; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RISING IMPEDANCE AND RISING THRESHOLDS. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24050952 ↗
  • MalfunctionNov 17, 2025Pacing Problem; Under-Sensing; Decreased Sensitivity; Pacing Problem; Under-Sensing; Decreased Sensitivity

    IT WAS FURTHER REPORTED THAT THE RV LEAD EXHIBITED LOW AMPLITUDE R-WAVES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23568553 ↗
  • MalfunctionNov 11, 2025Over-Sensing; High Capture Threshold; Over-Sensing; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONICALLY HIGH THRESHOLDS AS WELL AS OVERSENSING ON THE VENTRICULAR LEAD. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23522293 ↗
  • MalfunctionNov 7, 2025Decreased Sensitivity; Decreased Sensitivity

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED DECREASING R WAVES. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23496214 ↗
  • MalfunctionNov 3, 2025Break; Break

    THE PATIENT REPORTED THAT THEY WERE SHORT OF BREATH ALL THE TIME AND THAT ONE OF THEIR PACING LEADS WAS BROKEN. THE PACING LEAD REMAINS IN USE. ;NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dyspnea

    MAUDE report 23449353 ↗
  • InjuryOct 16, 2025Signal Artifact/Noise; Fracture; High impedance; High Capture Threshold; Unstable Capture Threshold; Signal Artifact/Noise; Fracture; High impedance; High Capture Threshold; Unstable Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED SUSPECTED FRACTURE, HIGH IMPEDANCE, HIGH THRESHOLD, UNSTABLE THRESHOLD AND NOISE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23310800 ↗
  • MalfunctionOct 15, 2025Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING ON CURRENT AND STORED ELECTROGRAMS (EGM) AND SHORT V-V IN TERVALS. THE RIGHT ATRIAL (RA) LEAD EXHIBITED ATRIAL LEAD POSITION CHECK FAILURE AND HIGH THRESHOLDS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23298586 ↗
  • InjurySep 15, 2025Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS FRACTURED. HIGH IMPEDANCE, NOISE AND OVERSENSING WERE ALSO OBSERVED. THE LEAD WAS REPROGRAMMED AND LATER CAPPED WITH REPLACEMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23058663 ↗
  • MalfunctionJul 18, 2025Failure to Capture; Over-Sensing; Capturing Problem; Unstable Capture Threshold; Failure to Capture; Over-Sensing; Capturing Problem; Unstable Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT NON-CAPTURE, VARIABLE THRESHOLDS AND LARGE INCREASES IN THRESHOLDS. A HIGH SENSING INTEGRITY COUNTER (SIC) WAS ALSO OBSERVED. IT WAS NOTED "PATIENT HAD BEEN SYMPTOMATIC". THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 22540041 ↗
  • MalfunctionJun 23, 2025Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED NO CAPTURE, NO SENSING AND LOW IMPEDANCE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NO SENSING ALSO. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22307751 ↗
  • MalfunctionJun 5, 2025Electrical /Electronic Property Problem; High impedance; Impedance Problem; Electrical /Electronic Property Problem; High impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A RISE IN IMPEDANCE AND HIGH OUT OF RANGE IMPEDANCE RESULTING POLARITY SWITCH. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22148848 ↗
  • InjuryJun 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED BY THE PATIENT¿S SPOUSE THAT A LEAD ¿NEEDS TO BE REPLACED AND IT WAS TURNED OFF AT SOME POINT.¿ IT WAS SUBSEQUENTLY NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS CAPPED AND REPLACED DUE TO AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22133436 ↗
  • InjuryMay 26, 2025Signal Artifact/Noise; Signal Artifact/Noise

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LEAD NOISE. THE LEADS WERE CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22084378 ↗
  • MalfunctionFeb 19, 2025Failure to Capture; Failure to Capture

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NO CAPTURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21408497 ↗
  • InjuryFeb 10, 2025Fracture; High impedance; High Capture Threshold; Fracture; High impedance; High Capture Threshold

    IT WAS REPORTED THAT APPROXIMATELY FIVE AND A HALF YEARS POST RIGHT VENTRICULAR (RV) LEAD IMPLANTATION, THE LEAD FRACTURED AND WAS REPLACED WITH ANOTHER RV LEAD. IT WAS FURTHER REPORTED THAT APPROXIMATELY FIFTEEN YEARS LATER, THE REPLACEMENT RV LEAD HAD A SUSPECTED FRACTURE IN THE VOLTAGE PORTION AND EXHIBITED OUT OF RANGE (OOR) HIGH PACING IMPEDANCE AND HIGH THRESHOLDS. THE TWO LEADS WERE PARTIALLY CLIPPED AND CAPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21344139 ↗