Capsure®Z Novus 5054-58Recalls & reported problems
Medtronic, Inc. · LEAD 5054-58 CAPSUREZ NOVUS USEN · Class 3 · Implant
All models of CapSureZ Novus (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Capsure Z Novus pacing lead has 272 reports, including 2 deaths and 180 injuries. Most describe the lead needing more electrical energy to pace the heart, plus readings that drift or misread heart signals. Some describe shortness of breath, dizziness or near-fainting in the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 273
- Deaths reported
- 2
- Injuries reported
- 181
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 273 adverse event reports that list this exact device (DI 00681490123228), received between Apr 10, 2013 and Sep 17, 2026 — classified as 2 deaths, 181 injuries, 90 malfunctions.
The most frequently reported specific device problem is “High Capture Threshold” (91 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Capture Threshold91
- High impedance50
- Over-Sensing49
- Failure to Capture45
- Impedance Problem35
- Signal Artifact/Noise29
- Under-Sensing28
- Capturing Problem24
Patient problems
- No Clinical Signs, Symptoms or Conditions126
- No Known Impact Or Consequence To Patient63
- Unspecified Infection12
- Bradycardia11
- Dizziness11
- No Consequences Or Impact To Patient9
- Syncope8
- Syncope/Fainting8
Recent adverse event reports
The 25 most recent of 273 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50811208 ↗InjuryAug 27, 2026Defibrillation/Stimulation Problem; Under-Sensing
It was reported that the patient experienced shortness of breath and the Right Ventricular (RV) lead exhibited undersensing resulting in delay of an episode and early termination. The lead remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Dyspnea
MAUDE report 50478809 ↗InjuryJul 6, 2026High impedance
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A DIAGNOSTIC RESET. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH UNDEFINED IMPEDANCE. THE IPG AND RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25716329 ↗MalfunctionJun 9, 2026Electrical /Electronic Property Problem; High impedance; Impedance Problem; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A POLARITY SWITCH, HIGH UNDEFINED IMPEDANCE AND HIGH THRESHOLDS. A GRADUAL RISE IN IMPEDANCE WAS NOTED. LEAD MINERALIZATION WAS SUSPECTED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25443888 ↗MalfunctionMay 8, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONIC HIGH THRESHOLDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25121413 ↗InjuryApr 23, 2026Intermittent Capture
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PRE-SYNCOPE. IT WAS FURTHER NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT CAPTURE. THE RV LEAD WAS CAPPED AND REPLACED WITH A LEADLESS IMPLANTABLE PULSE GENERATOR (IPG). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Presyncope
MAUDE report 24962766 ↗MalfunctionMar 30, 2026High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24726363 ↗MalfunctionMar 11, 2026Capturing Problem; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED AN INCREASE IN IMPEDANCE. IT WAS FURTHER NOTED THAT THE RV LEAD EXHIBITED AN INCREASE IN THRESHOLDS. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24569393 ↗MalfunctionMar 9, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING IN NON-PHYSIOLOGIC INTERVALS ON STORED EPISODES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24543572 ↗InjuryJan 22, 2026High Capture Threshold
IT WAS FURTHER REPORTED THAT PACEMAKER MEDIATED TACHYCARDIA WAS SUSPECTED. IT WAS FURTHER NOTED THAT THE RA AND RV LEAD WAS REPROGRA MMED.
Patient problems: No Clinical Signs, Symptoms or Conditions; Tachycardia
MAUDE report 24146390 ↗InjuryJan 13, 2026Capturing Problem; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RISING IMPEDANCE AND RISING THRESHOLDS. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24050952 ↗MalfunctionNov 17, 2025Pacing Problem; Under-Sensing; Decreased Sensitivity; Pacing Problem; Under-Sensing; Decreased Sensitivity
IT WAS FURTHER REPORTED THAT THE RV LEAD EXHIBITED LOW AMPLITUDE R-WAVES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23568553 ↗MalfunctionNov 11, 2025Over-Sensing; High Capture Threshold; Over-Sensing; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED CHRONICALLY HIGH THRESHOLDS AS WELL AS OVERSENSING ON THE VENTRICULAR LEAD. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23522293 ↗MalfunctionNov 7, 2025Decreased Sensitivity; Decreased Sensitivity
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED DECREASING R WAVES. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23496214 ↗MalfunctionNov 3, 2025Break; Break
THE PATIENT REPORTED THAT THEY WERE SHORT OF BREATH ALL THE TIME AND THAT ONE OF THEIR PACING LEADS WAS BROKEN. THE PACING LEAD REMAINS IN USE. ;NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dyspnea
MAUDE report 23449353 ↗InjuryOct 16, 2025Signal Artifact/Noise; Fracture; High impedance; High Capture Threshold; Unstable Capture Threshold; Signal Artifact/Noise; Fracture; High impedance; High Capture Threshold; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED SUSPECTED FRACTURE, HIGH IMPEDANCE, HIGH THRESHOLD, UNSTABLE THRESHOLD AND NOISE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23310800 ↗MalfunctionOct 15, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING ON CURRENT AND STORED ELECTROGRAMS (EGM) AND SHORT V-V IN TERVALS. THE RIGHT ATRIAL (RA) LEAD EXHIBITED ATRIAL LEAD POSITION CHECK FAILURE AND HIGH THRESHOLDS. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23298586 ↗InjurySep 15, 2025Signal Artifact/Noise; Fracture; High impedance; Over-Sensing; Signal Artifact/Noise; Fracture; High impedance; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS FRACTURED. HIGH IMPEDANCE, NOISE AND OVERSENSING WERE ALSO OBSERVED. THE LEAD WAS REPROGRAMMED AND LATER CAPPED WITH REPLACEMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23058663 ↗MalfunctionJul 18, 2025Failure to Capture; Over-Sensing; Capturing Problem; Unstable Capture Threshold; Failure to Capture; Over-Sensing; Capturing Problem; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT NON-CAPTURE, VARIABLE THRESHOLDS AND LARGE INCREASES IN THRESHOLDS. A HIGH SENSING INTEGRITY COUNTER (SIC) WAS ALSO OBSERVED. IT WAS NOTED "PATIENT HAD BEEN SYMPTOMATIC". THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 22540041 ↗MalfunctionJun 23, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED NO CAPTURE, NO SENSING AND LOW IMPEDANCE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NO SENSING ALSO. THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22307751 ↗MalfunctionJun 5, 2025Electrical /Electronic Property Problem; High impedance; Impedance Problem; Electrical /Electronic Property Problem; High impedance; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A RISE IN IMPEDANCE AND HIGH OUT OF RANGE IMPEDANCE RESULTING POLARITY SWITCH. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22148848 ↗InjuryJun 3, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED BY THE PATIENT¿S SPOUSE THAT A LEAD ¿NEEDS TO BE REPLACED AND IT WAS TURNED OFF AT SOME POINT.¿ IT WAS SUBSEQUENTLY NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS CAPPED AND REPLACED DUE TO AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22133436 ↗InjuryMay 26, 2025Signal Artifact/Noise; Signal Artifact/Noise
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LEAD NOISE. THE LEADS WERE CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22084378 ↗MalfunctionFeb 19, 2025Failure to Capture; Failure to Capture
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NO CAPTURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21408497 ↗InjuryFeb 10, 2025Fracture; High impedance; High Capture Threshold; Fracture; High impedance; High Capture Threshold
IT WAS REPORTED THAT APPROXIMATELY FIVE AND A HALF YEARS POST RIGHT VENTRICULAR (RV) LEAD IMPLANTATION, THE LEAD FRACTURED AND WAS REPLACED WITH ANOTHER RV LEAD. IT WAS FURTHER REPORTED THAT APPROXIMATELY FIFTEEN YEARS LATER, THE REPLACEMENT RV LEAD HAD A SUSPECTED FRACTURE IN THE VOLTAGE PORTION AND EXHIBITED OUT OF RANGE (OOR) HIGH PACING IMPEDANCE AND HIGH THRESHOLDS. THE TWO LEADS WERE PARTIALLY CLIPPED AND CAPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21344139 ↗
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page