CapSure®Z Novus 5554-53Recalls & reported problems
Medtronic, Inc. · LEAD 5554-53 CAPSURE Z NOVUS US EN · Class 3 · Implant
All models of CapSureZ Novus (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic CapSure Z Novus pacemaker lead has 296 reports, including 106 injuries and 2 deaths. Many describe the lead sensing too much noise or failing to detect heartbeats, and some describe pacing pulses failing to make the heart beat. Some patients reported infections, including one bloodstream infection after which the device system was replaced. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 300
- Deaths reported
- 3
- Injuries reported
- 108
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 300 adverse event reports that list this exact device (DI 00681490135108), received between Dec 13, 2016 and Sep 24, 2026 — classified as 3 deaths, 108 injuries, 189 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (120 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Under-Sensing120
- Over-Sensing65
- High Capture Threshold43
- Signal Artifact/Noise40
- High impedance27
- Failure to Capture24
- Adverse Event Without Identified Device or Use Problem22
- Low impedance20
Patient problems
- No Clinical Signs, Symptoms or Conditions150
- No Known Impact Or Consequence To Patient98
- Unspecified Infection12
- No Consequences Or Impact To Patient11
- Chest Pain6
- Bacterial Infection5
- Sepsis5
- Arrhythmia3
Recent adverse event reports
The 25 most recent of 300 reports, in the reporter's own words as published by the FDA.
DeathSep 24, 2026Adverse Event Without Identified Device or Use Problem
It was reported by the patients representative that the patient implanted with a Cardiac Resynchronization Therapy Defibrillator (CRT-D) system "had a sepsis infection from "[the]" defibrillator ports" and "there was an infection on the lead from "[the]" defibrillator." The patient¿s representative reported the patient was "put on antibiotics instead of replacing as usually done." The patient expired approximately 8 weeks later.
Patient problems: Sepsis; Unspecified Infection
MAUDE report 50883752 ↗InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Unspecified Infection
MAUDE report 50811348 ↗MalfunctionSep 9, 2026Low impedance
It was reported that the Right Atrial (RA) lead triggered an alert for out of range low pacing impedance. The lead remains in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50686907 ↗InjuryAug 25, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced bacteremia. The Implantable Pulse Generator (IPG) system and previously capped Right Ventricular (RV) lead were explanted and replaced with a leadless IPG. No further patient complications have been reported as a result of this event.
Patient problems: Bacteremia
MAUDE report 50432541 ↗InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced a suspected pocket infection. The Implantable Pulse Generator (IPG) system was explante d. No further patient complications have been reported as a result of this event.
Patient problems: Medical device site infection
MAUDE report 50142970 ↗MalfunctionJul 6, 2026Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING/NOISE, RESULTING IN MISCLASSIFICATION OF EPISODE TYPE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25714097 ↗MalfunctionJun 17, 2026Signal Artifact/Noise
IT WAS REPORTED THAT NOISE ON THE RIGHT ATRIAL (RA) LEAD WAS OBSERVED. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25538232 ↗InjuryJun 11, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS CAPPED AND REPLACED FOR AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25467516 ↗MalfunctionJun 2, 2026Electrical /Electronic Property Problem; Low impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE AND A ATRIAL LEAD POSITION CHECK FAILURE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25357440 ↗MalfunctionMay 28, 2026Low impedance; Impedance Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD TRIGGERED AN ALERT FOR OUT OF RANGE (OOR) LOW IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dizziness; No Clinical Signs, Symptoms or Conditions; Presyncope
MAUDE report 25317388 ↗MalfunctionApr 14, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R WAVE OVERSENSING RECORDING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODES INCORRECTLY. THE LEAD REMAINS IN USE, NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24872061 ↗MalfunctionApr 7, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24799320 ↗MalfunctionMar 5, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEAD REMAINS IN USE. NO PATIENT C OMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24518918 ↗MalfunctionFeb 26, 2026Capturing Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A RECENT INCREASE IN CAPTURE THRESHOLDS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24454653 ↗MalfunctionFeb 17, 2026Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing
IT WAS FURTHER REPORTED THAT THE RA LEAD EXHIBITED ELECTROMAGNETIC INTERFERENCE (EMI) / NOISE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24365131 ↗MalfunctionFeb 12, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING ON CURRENT AND STORED ELECTROGRAMS (EGM) DURING VENTRICULAR TACHYCARDIA (VT) AND ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODE AND THE P WAVE IS VERY SMALL BUT STABLE. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24320407 ↗MalfunctionDec 17, 2025Failure to Capture; Low impedance; Failure to Capture; Low impedance
IT WAS REPORTED THAT A LEAD WARNING TRIGGERED DUE TO A ZERO IMPEDANCE MEASUREMENT ON THE RIGHT ATRIAL (RA) LEAD. IN ADDITION, THERE WAS A POSSIBLE FAILURE TO CAPTURE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23827267 ↗InjuryNov 28, 2025Failure to Capture; Pacing Problem; Failure to Capture; Pacing Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED AN EXIT BLOCK AND LOSS OF CAPTURE. THE LEAD WAS REPROGRAMMED TO INACTIVE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23671050 ↗MalfunctionNov 17, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING. IT APPEARS POTENTIAL ATRIAL UNDER SENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF AT/AF EVENT(S) AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23567531 ↗MalfunctionNov 7, 2025Failure to Capture; High impedance; Failure to Capture; High impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDEFINED HIGH IMPEDANCE. IT WAS NOTED THAT THE LEAD IMPEDANCE HAD BEEN HISTORICALLY HIGH. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23500625 ↗MalfunctionSep 18, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23083243 ↗InjurySep 17, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICAT IONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23070550 ↗InjurySep 9, 2025Fracture; High Capture Threshold; Fracture; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH AND RISING IMPEDANCES AND HIGH THRESHOLDS. THE RIGHT ATRIAL (RA) LEAD WAS SUSPECTED TO BE FRACTURED AND EXHIBITED HIGH THRESHOLDS. THE RV LEAD WAS CAPPED AND REPLACED AND THE RA LEAD WAS CAPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23002826 ↗MalfunctionAug 20, 2025Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22857266 ↗MalfunctionJul 30, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL HIGH RATE EPISODES, TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22664774 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page