Skip to content

CapSure®SP Novus 5592-45 (UDI 00681490135696)Recalls & reported problems

Medtronic, Inc. · LEAD 5592-45 CAPSURE SP NOVUS US · Class 3 · Implant

All models of CapSureSP Novus (14)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic CapSure SP Novus pacing lead has 246 reports on file, with no deaths and 70 injuries. Many describe the lead missing real heartbeats or picking up noise as heartbeats, and some describe high pacing thresholds, a confirmed fracture, or infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
246
Deaths reported
0
Injuries reported
70
Recalls naming it
0

Summary

The FDA's MAUDE database contains 246 adverse event reports that list this exact device (DI 00681490135696), received between Feb 9, 2017 and Aug 18, 2026 — classified as 70 injuries, 176 malfunctions.

The most frequently reported specific device problem is “Under-Sensing” (111 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 4 reportsJan 2024: 0 reportsFeb 2024: 3 reportsMar 2024: 2 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 5 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 5 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 5 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 3 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 4 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 2 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Under-Sensing111
  • Over-Sensing86
  • High Capture Threshold26
  • Adverse Event Without Identified Device or Use Problem21
  • Signal Artifact/Noise18
  • Electrical /Electronic Property Problem16
  • Pacing Problem16
  • Low impedance14

Patient problems

  • No Clinical Signs, Symptoms or Conditions126
  • No Known Impact Or Consequence To Patient54
  • No Consequences Or Impact To Patient19
  • Unspecified Infection15
  • Dizziness5
  • Dyspnea5
  • Arrhythmia4
  • Chest Pain4

Recent adverse event reports

The 25 most recent of 246 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50296150 ↗
  • MalfunctionAug 6, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50155222 ↗
  • MalfunctionJul 7, 2026Over-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING ON STORED EPISODES RESULTING IN MISCLASSIFICATION OF EPISODE TYPE. IT WAS ALSO REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED NOISE. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS, WITH CURRENT THRESHOLDS REMAINING WITHIN NORMAL LIMITS AND ARE CONSISTENT. THE IPG WAS EXPLANTED DUE TO NORMAL ELECTIVE REPLACEMENT INDICATOR (ERI). THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25734295 ↗
  • MalfunctionJul 1, 2026Over-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING AND FAR FIELD R-WAVE (FFRW) OVERSENSING ON STORED ELECTROGRAMS (EGM) TRIGGERING FALSE ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) DETECTIONS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25676022 ↗
  • MalfunctionJun 23, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS FURTHER REPORTED THAT NO ACTION WAS TAKEN ON THE RA LEAD, DUE TO UNABLE TO DEMONSTRATE ANY ISSUES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25595255 ↗
  • MalfunctionMay 14, 2026Signal Artifact/Noise; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED NOISE/OVERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25170529 ↗
  • MalfunctionApr 17, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911602 ↗
  • MalfunctionMar 20, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS RESULTING IN UNNECESSARY PACING AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649123 ↗
  • MalfunctionMar 16, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24603038 ↗
  • MalfunctionMar 10, 2026Signal Artifact/Noise

    IT WAS REPORTED THAT NOISE/ ELECTROMAGNETIC INTERFERENCE (EMI) WAS OBSERVED ON A STORED ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODE, WHICH TRIGGERED EPISODE MISCLASSIFICATION. THE CARDIAC RESYNCHRONIZATION THERAPY PACEMAKER (CRT-P) AND RIGHT ATRIAL (RA) LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557202 ↗
  • InjuryMar 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED A POCKET INFECTION. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Medical device site infection

    MAUDE report 24549882 ↗
  • MalfunctionFeb 18, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS RESULTING IN UNNECESSARY PACING AT TIMES AND FALSE TERMINATION OF DEVICE CLASSIFIED AT/AF. THE LEAD REMAINS IN USE. NO PATIENT COMP LICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24375843 ↗
  • MalfunctionFeb 13, 2026High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24331947 ↗
  • MalfunctionJan 16, 2026Pacing Problem

    IT WAS REPORTED THAT THERE WAS ATRIAL EVENTS FALLING IN BLANKING/REFRACTORY RESULTING IN UNNECESSARY PACING AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24093471 ↗
  • MalfunctionNov 24, 2025Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23632658 ↗
  • MalfunctionNov 14, 2025Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES AND APPEARS TO RESULT IN INAPPROPRIATE ATRIAL PACING. THE RA LEAD REMAINS IN  USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23552522 ↗
  • InjuryNov 14, 2025Signal Artifact/Noise; Signal Artifact/Noise

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED NOISE. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23551985 ↗
  • MalfunctionSep 15, 2025Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING AND FAR FIELD R-WAVE (FFRW) OVERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23050507 ↗
  • MalfunctionSep 11, 2025Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA (AT)/ ATRIAL FIBRILLATION (AF) EVENT(S) AT TIMES AND APPEARS TO RESULT IN INAPPROPRIATE ATRIAL PACING AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23023858 ↗
  • MalfunctionAug 27, 2025Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD APPEARED TO BE UNDER SENSING, TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA (AT)/ATRIAL FIBRILLATION (AF) EVENT(S) AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22902190 ↗
  • InjuryAug 20, 2025Fracture; Appropriate Device Problem Term/Code Not Available; Fracture; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS CAPPED AND REPLACED FOR UNKNOWN REASONS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22847722 ↗
  • MalfunctionAug 14, 2025Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING NOTED ON STORED ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22802008 ↗
  • MalfunctionJul 9, 2025Over-Sensing; Impedance Problem; Over-Sensing; Impedance Problem

    IT WAS REPORTED THAT THE DEVICE APPEARS TO BE OVERSENSING ON THE RIGHT ATRIAL (RA) LEAD ON STORED ELECTROGRAMS (EGM) AND ALSO EXHIBITED ATRIAL UNIPOLAR LEAD IMPEDANCE WARNING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22456466 ↗
  • InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. IT WAS ALSO NOTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) HAD ERODED THROUGH THE SKIN. THE IPG SYSTEM WAS REMOVED AND A TEMPORARY LEAD WAS IMPLANTED WHICH WAS ATTACHED TO THE IPG EXTERNALLY UNTIL THE PATIENT HAD A NEW SYSTEM IMPLANTED A FEW DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pocket Erosion; Unspecified Infection

    MAUDE report 22369042 ↗
  • MalfunctionJun 9, 2025Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS AT TIMES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22166357 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page