VersaCross™ Large Access Solution VLAK0114Recalls & reported problems
Boston Scientific Corporation · Large Access Solution · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447019596), received between Aug 18, 2025 and Nov 4, 2025 — classified as 1 death, 1 injury.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance1
- Display or Visual Feedback Problem1
- Positioning Failure1
- Positioning Problem1
Patient problems
- Pericardial Effusion2
- Low Blood Pressure/ Hypotension1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 4, 2025Positioning Failure; Positioning Problem; Positioning Failure; Positioning Problem
IT WAS REPORTED THAT A PERICARDIAL EFFUSION (PE) OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A WATCHMAN PROCEDURE, A VERSACROSS LARGE ACCESS SOLUTION AND A VERSACROSS ACCESS STEERABLE SOLUTION WERE SELECTED FOR USE. THE PHYSICIAN WAS ATTEMPTING TO GAIN TRANSEPTAL ACCESS WITH DIFFICULTY ADEQUATELY IMAGING THE TRANSEPTAL LOCATION VIA TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE). THE PHYSICIAN WAS UTILIZING VERSACROSS LARGE ACCESS INITIALLY THEN SWITCHED OUT TO VERSACROSS STEERABLE DUE TO HAVING ISSUES REACHING THE TRANSEPTAL LOCATION DUE TO PATIENT ANATOMY AND DEVICE REACH TO THE SEPTAL BY ADDING ADDITIONAL CURVE TO THE TRANSEPTAL SYSTEM. UPON OBTAINING A POSITION ON THE FOSSA VIA TEE, THE PHYSICIAN ADVISED TO ACTIVATE RF TO GO TRANSEPTAL. THE RF WIRE ADVANCED EASILY HOWEVER PHYSICIAN EXPERIENCED DIFFICULTY ADVANCING THE STEERABLE SHEATH ACROSS THE SEPTUM. THE PHYSICIAN VERIFIED TRANSEPTAL LOCATION AND INSERTED THE WATCHMAN TRUSTEER SHEATH. UPON INSERTING THE PIGTAIL CATHETER THROUGH THE SHEATH AND TAKING AN ANGIOGRAM, IT WAS NOTED CONTRAST STAINING IN THE PERICARDIAL SPACE. AT THAT TIME, IT WAS CONFIRMED VIA TEE THAT THEY WERE IN THE PERICARDIAL SPACE. THE PATIENT REMAINED STABLE AND A SM
Patient problems: Pericardial Effusion
MAUDE report 23468584 ↗DeathAug 18, 2025Display or Visual Feedback Problem; Difficult to Advance; Display or Visual Feedback Problem; Difficult to Advance
IT WAS REPORTED THAT A PATIENT DEATH OCCURRED. DURING A CONCOMITANT PULSE FIELD ABLATION PROCEDURE, A VERSACROSS LARGE ACCESS WAS SELECTED FOR USE. A BASELINE EFFUSION WAS NOTED PRIOR TO TRANSSEPTAL. ONCE GROIN ACCESS WAS OBTAINED, THERE WAS DIFFICULTY ASCENDING THE FEMORAL VEIN AND CROSSING TRANSEPTAL. TRANSEPTAL WAS SUDDEN AND VISIBILITY WAS POOR. THE PHYSICIAN FLOSSED WITH THE VERSACROSS LARGE ACCESS. HOWEVER, THE FARADRIVE SHEATH COULD NOT GO ACROSS. SO, IT WAS ATTEMPTED WITH THE FARADRIVE DILATOR AND TRUSTEER DILATOR, THEN WAS ABLE TO CROSS WITH FARADRIVE. THE PRESSURE DROPPED A FEW TIMES AT TRANSEPTAL DURING FLOSSING. THE PRESSURE WAS TREATED. THE PHYSICIAN USED INTRACARDIAC ECHOCARDIOGRAPHY (ICE) TO WATCH FOR EFFUSION WHILE ABLATION. IT STARTED GROWING WHEN ABLATING RIGHT VEINS. A PERICARDIOCENTESIS WAS DONE AFTER FINISHING ABLATION. ONE LITER WAS DRAWN AND PATIENT WAS STABILIZED. LATER, THE PATIENT PASSED AWAY IN THE INTENSIVE CARE UNIT (ICU) FROM COMPLICATIONS. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED).
Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 22832497 ↗
Other Percutaneous atrial catheter kit devices
- VersaCross Connect Access Solution for FARADRIVE VXAK0041Boston Scientific Corporation · Percutaneous Atrial Catheter Kit3 deaths275 reports
- VersaCross Connect™ LAAC Access Solution VXAK0107Boston Scientific Corporation · Percutaneous atrial catheter kit3 deaths77 reports
- VersaCross™ Access Solution VXSK0122Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths21 reports
- VersaCross™ Access Solution VXSK0131Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths12 reports
- VersaCross™ Access Solution VXSK0132Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall11 reports
- VersaCross™ Access Solution VXSK0134Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall1 death11 reports
- VersaCross™ Access Solution VXSK0121Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall6 reports
- VersaCross™ Access Solution VXSK0124Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page