VersaCross™ Access Solution VXSK0131Recalls & reported problems
Boston Scientific Corporation · Access Solution · Class 2
All models of VersaCross Access Solution (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 10
- Deaths reported
- 2
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00685447019862), received between Oct 22, 2025 and Sep 4, 2026 — classified as 2 deaths, 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Nonstandard Device” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 10, 2025Class IIOpen, Classified
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Z-2103-2025 · Root cause: Packaging change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Nonstandard Device2
- Device Sensing Problem1
- Difficult to Advance1
- Failure to Advance1
- Material Integrity Problem1
- Material Perforation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Low Blood Pressure/ Hypotension4
- Pericardial Effusion4
- Cardiac Tamponade2
- Thrombosis/Thrombus2
- Cardiac Perforation1
- Hematoma1
- Perforation of Vessels1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 4, 2026Nonstandard Device
It was reported that the device tip dent occurred.During a pulse field ablation procedure for the treatment of paroxysmal atrial fibrillation, Versacross Access Solutions was selected for transeptal access. While preparing to perform the transseptal puncture, the physician attempted to insert the device through the groin access site but noted that it was not advancing properly. The device was withdrawn prior to insertion into the patient and inspected. Upon inspection, a dent was observed on the dilator tip of the VersaCross Access Solution. The issue was identified immediately before use when the device could not be inserted as expected. Versacross Access Solution was replaced with a new Versacross Access Solution to continue procedure without further issues. No patient complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50602484 ↗DeathJun 25, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT DEATH HAS OCCURRED. DURING FLX PRO PROCEDURE VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. TENTING WAS DONE ON THE SEPTUM INFERIORLY AND POSTERIORLY AND THE VERSACROSS CROSSED WITHOUT ANY RF BEING APPLIED. VERSACROSS PIGTAIL WIRE WENT INTO THE APPENDAGE. THE WIRE WAS PULLED BACK SUCCESSFULLY INTO THE VEIN TO BRING THE VERSACROSS FIXED SHEATH ACROSS. AT THIS TIME THE PHYSICIAN NOTICED A LARGE PERICARDIAL EFFUSION AND BLOOD PRESSURES WERE DROPPING. PERICARDIOCENTESIS TRAY WAS USED TO TAP THE PERICARDIAL SPACE AND CALLED A CODE. A FEW MINUTES LATER SURGICAL BACK UP WAS IN THE ROOM. CHEST COMPRESSIONS WERE DONE FOR ABOUT 45 MINUTES, AND BLOOD TRANSFUSION WAS DONE, BUT THE PATIENT'S BLOOD PRESSURE NEVER STABILIZED. THE OFFICIAL CAUSE OF DEATH WAS CARDIAC TAMPONADE. THE DEVICE IS NOT EXPECTED TO BE RETURNING AS IT HAS BEEN DISPOSED.
Patient problems: Cardiac Perforation; Cardiac Tamponade; Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 25627534 ↗MalfunctionMay 25, 2026Failure to Advance
IT WAS REPORTED THAT DURING A PROCEDURE USING A RF WIRE FOR TRANSSEPTAL PUNCTURE THE WIRE WAS UNABLE TO CROSS; NO ERRORS WERE OBSERVED ON THE GENERATOR DURING USE OF THE VERSACROSS WIRE, APPROXIMATELY 2 MM OF THE WIRE WAS EXPOSED OUTSIDE THE DILATOR PRIOR TO THE ATTEMPTED CROSSING, TWO CROSSING ATTEMPTS WERE MADE WITH RF APPLIED EACH TIME, AND NO DAMAGE WAS NOTED ON THE DISTAL END OF THE WIRE UPON REMOVAL FROM THE BODY. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF SAME DEVICE. NO PATIENT COMPLICATIONS OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25267603 ↗MalfunctionMay 7, 2026Material Perforation
IT WAS REPORTED THAT DURING PREPARATION/ PRIOR TO PROCEDURE OF FLX PRO PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. IT WAS NOTICED THAT THE TIP OF THE DILATOR WAS DAMAGED. THE ISSUE WAS NOTED BEFORE INSERTING INTO THE BODY WHERE THE TIP PIECE APPEARED FRAYED. THERE WERE NO PATIENT COMPLICATIONS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ALTERNATIVE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25118957 ↗InjuryMay 6, 2026Difficult to Advance
IT WAS REPORTED THAT PERICARDIAL EFFUSION (PE) OCCURRED. A LEFT ATRIAL APPENDAGE (LAA) CLOSURE PROCEDURE WAS BEING PERFORMED UNDER INTRACARDIAC ECHOCARDIOGRAPHY (ICE) GUIDANCE. BASELINE IMAGING DEMONSTRATED NO PE. THE INTERATRIAL SEPTUM WAS NOTED TO BE THICK, AND TRANSSEPTAL PUNCTURE (TSP) WAS PERFORMED WITH A VERSACROSS FIXED CURVE TRANSSEPTAL SYSTEM IN AN INFERIOR/MID LOCATION. DIFFICULTY WAS ENCOUNTERED ADVANCING THE VERSACROSS SYSTEM INTO THE LEFT ATRIUM, BUT ACCESS WAS EVENTUALLY ACHIEVED. ATTEMPTS WERE THEN MADE TO ADVANCE A WATCHMAN TRUSTEER ACCESS SYSTEM (WAS) ACROSS THE SEPTUM HOWEVER THE WAS WOULD NOT CROSS. A SECOND TSP WAS PERFORMED IN A SIMILAR LOCATION TO IMPROVE ACCESS. FOLLOWING THESE ATTEMPTS, A PE WAS IDENTIFIED NEAR THE RIGHT ATRIUM, ACCOMPANIED BY A MILD DECREASE IN BLOOD PRESSURE. THE PROCEDURE WAS ABORTED AND PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT WAS ADMITTED OVERNIGHT. IT WAS FURTHER REPORTED THAT THE PATIENT UNDERWENT SURGERY AND RECOVERED WITHOUT FURTHER COMPLICATION. SEVERAL DAYS LATER, THE PATIENT WAS DISCHARGED.
Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 25094002 ↗DeathApr 16, 2026Adverse Event Without Identified Device or Use Problem
IIT WAS REPORTED THAT AN AORTIC ROOT PERFORATION, PERICARDIAL EFFUSION (PE) AND CARDIAC TAMPONADE OCCURRED. THE PROCEDURE WAS CANCELLED. A LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE WAS BEING PERFORMED. A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THERE WAS NO EFFUSION NOTED DURING PRE IMAGING. DUE TO CHALLENGING ANATOMY, TORTUROUS VESSELS AND A DILATED RIGHT ATRIUM, THE PHYSICIAN UTILIZED MULTIPLE TRANSEPTAL DEVICES TO ATTEMPT TO GAIN ACCESS. PHYSICIAN ATTEMPTED TRANSSEPTAL PUNCTURE USING VERSACROSS CONNECT LAAC, VERSACROSS ACCESS SOLUTION AND NRG TRANSSEPTAL NEEDLE. WHEN THE PHYSICIAN WAS ATTEMPTING TO CROSS WITH THE NRG, PATIENT BLOOD PRESSURE DROPPED. TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) IMAGING WAS UTILIZED TO EVALUATE FOR EFFUSION. IT WAS IMMEDIATELY NOTED FLUID COMPRESSING THE LEFT ATRIUM. ARTERIAL ACCESS WAS OBTAINED, FLOUROSCOPY IMAGING WAS PREFORMED WHICH REVEALED A PERFORATION IN THE AORTIC ROOT. THE PATIENT HAD PREVIOUS CORONARY ARTERY BYPASS GRAFTING (CABG), SO THIS WAS NOT A PERICARDIAL EFFUSION, IT WAS AN EFFUSION CAUSING EXTRINSIC COMPRESSION OF THE LEFT ATRIUM. THE EFFUSION WAS EVENTUALLY LARGE ENOUGH TO CAUSE COMPLETE CHAMBER COLLAPSE. THE HOSPITAL INITI
Patient problems: Cardiac Tamponade; Hematoma; Low Blood Pressure/ Hypotension; Perforation of Vessels; Pericardial Effusion; Thrombosis/Thrombus
MAUDE report 24901359 ↗MalfunctionApr 14, 2026Nonstandard Device
IT WAS REPORTED THAT PRIOR TO MITRAL CLIP PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. IT WAS NOTED THE ONE END OF THE RF WIRE WAS TOO ROUGH AND 'CHEWED UP' TO BE USED. THERE WAS NO PACKAGING DAMAGE. THUS, THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24878753 ↗MalfunctionFeb 19, 2026Material Integrity Problem
IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. IT WAS MENTIONED THAT THE RF WIRE WAS NOTED SHREDDED DURING PREPARATION FOR THE PROCEDURE. THUS, THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATION WAS REPORTED. THE DEVICE IS NOT EXPECTED TO BE RETURN FOR ANALYSIS AS IT WAS RETAINED BY CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24389232 ↗InjuryFeb 2, 2026Device Sensing Problem
THE PATIENT EXPERIENCED A PERICARDIAL EFFUSION. IT WAS REPORTED THAT A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE DURING A FARAPULSE ABLATION TO TREAT PERSISTENT ATRIAL FIBRILLATION. INITIALLY, THE PHYSICIAN GAINED VASCULAR ACCESS WITHOUT DIFFICULTY AND SUCCESSFULLY INSERTED A ICE CATHETER AND A NON-BOSTON SCIENTIFIC CORONARY SINUS CATHETER. HENCE, THE VERSACROSS DEVICES WERE INSERTED, WITH NO RESISTANCE AND POSITIONED ON THE SEPTUM. THE TRANSSEPTAL WAS THEN COMPLETED, HOWEVER THE PHYSICIAN WAS UNABLE TO VISUALIZE THE RF WIRE LOCATION AND THEREFORE RETRACTED IT BACK INTO THE RIGHT ATRIUM. THEN, A SECOND TRANSSEPTAL PUNCTURE WAS PERFORMED USING THE SAME SYSTEM, SUCCESSFULLY ADVANCING THE THE VERSACROSS DEVICES INTO THE LEFT ATRIUM, AND THE HAVE BEEN REPLACED BY THE FARADRIVE IN THE LEFT ATRIUM. THE VERSACROSS RF WIRE WAS REPLACED BY THE NON-BOSTON SCIENTIFIC MAPPING CATHETER. UPON COMPLETION OF THE PRE-MAPPING, THE CATHETER WAS THEN EXCHANGED FOR THE FARAWAVE CATHETER SUCCESSFULLY. APPROXIMATELY 5-7 MINUTES INTO ABLATION, THE ANESTHESIOLOGIST REPORTED A DROP IN BLOOD PRESSURE. USING THE ICE CATHETER, THE PHYSICIAN IDENTIFIED A PERICARDIAL EFFUSION. A PERICARDIOCENTESIS WAS IMME
Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion
MAUDE report 24240527 ↗InjuryOct 22, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THROMBUS OCCURRED. PROCEDURE WAS ABORTED. IT WAS REPORTED THAT VERSACROSS CONNECT LAAC ACCESS SOLUTION SELECTED FOR USE. HEPARIN WAS ADMINISTRATED RIGHT BEFORE TRANSSEPTAL PUNCTURE. DURING PROCEDURE, INTRACARDIAC THROMBUS WAS DISCOVERED IN THE LEFT ATRIAL APPENDAGE AFTER THE CONTRAST SHOT WAS TAKEN. PIGTAIL CATHETER WAS IN THE APPENDAGE. PHYSICIAN WENT THROUGH THE PRE-IMAGING. THEY CLEARED FOR INTRA-ATRIAL THROMBUS. THERE WAS NO SPONTANEOUS ECHO CONTRAST OR ANYTHING THAT INDICATED A CLOT. THERE WAS NO INDEPENDENT MOVEMENT AND HAD VERY CLEAR IMAGING OF THE APPENDAGE IN 0,45,90, AND 135. THE PROCEDURE WENT NORMALLY. PHYSICIAN GAINED FEMORAL VEIN ACCESS AND USED THE VERSACROSS FIXED KIT TO GET LA ACCESS WITH A LOW-MID STICK. THE TRANSSEPTAL SHEATH WAS EXCHANGED FOR THE WATCHMAN TRUSTEER SHEATH. THE PIGTAIL CATHETER WAS LOADED OVER THE VERSACROSS WIRE AND INTO THE LEFT ATRIAL APPENDAGE. AROUND THIS TIME, WE GOT AN ACT OF 319. THE PIGTAIL WAS NAVIGATED INSIDE THE APPENDAGE, AND A CONTRAST SHOT WAS GIVEN. AFTER THE FLUORO, WE OPENED AND PREPARED THE 27MM WATCHMAN. AFTER WE FINISHED FLUSHING, THE DOCTORS MOVED TO SWAP THE PIGTAIL CATHETER OUT WITH THE DEVICE. PHYSICIA
Patient problems: Thrombosis/Thrombus
MAUDE report 23359015 ↗
FDA clearance and classification
The FDA regulates product code OFK (percutaneous atrial catheter kit) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VersaCross™ Access Solution VXSK0131
Has the VersaCross™ Access Solution VXSK0131 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2103-2025 (Class II), began on Jun 10, 2025: Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the VersaCross™ Access Solution VXSK0131?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Oct 22, 2025 and Sep 4, 2026. In those reports, the most common specific device problems are “Nonstandard Device” (2), “Device Sensing Problem” (1) and “Difficult to Advance” (1); the most common patient problems are “Low Blood Pressure/ Hypotension” (4), “Pericardial Effusion” (4) and “Cardiac Tamponade” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VersaCross™ Access Solution VXSK0131?
2 reports are classified as deaths, 3 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VersaCross™ Access Solution VXSK0131, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand VersaCross™ Access Solution, model VXSK0131). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Percutaneous atrial catheter kit devices
- VersaCross Connect Access Solution for FARADRIVE VXAK0041Boston Scientific Corporation · Percutaneous Atrial Catheter Kit3 deaths272 reports
- VersaCross Connect™ LAAC Access Solution VXAK0107Boston Scientific Corporation · Percutaneous atrial catheter kit3 deaths73 reports
- VersaCross™ Access Solution VXSK0122Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths18 reports
- VersaCross™ Access Solution VXSK0134Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall1 death11 reports
- VersaCross™ Access Solution VXSK0132Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall11 reports
- VersaCross™ Access Solution VXSK0124Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall4 reports
- VersaCross™ Access Solution VXSK0121Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall4 reports
- VersaCross™ Access Solution VXSK0101Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page