Skip to content

VersaCross™ Access Solution VXSK0121Recalls & reported problems

Boston Scientific Corporation · Access Solution · Class 2

All models of VersaCross Access Solution (15)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00685447019817), received between Oct 23, 2025 and Aug 13, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 10, 2025Class IIOpen, Classified

    Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

    Z-2103-2025 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Integrity Problem3
  • Adverse Event Without Identified Device or Use Problem1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Pericardial Effusion1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 13, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that during Atrial Fibrillation procedure a FARAWAVE and VersaCross Access Solution were selected for use. When trying to perform basket in the RIPV, the spline became inverted. It was tried manually fixing the spline, but after going back into the left atrium, the catheter inverted again. Manually fixing the splines but did not work. Ablation was finished and it was noticed a pericardial effusion around 4 hours after the procedure. The catheter had some resistance with the wire when trying to take the catheter out of the body and the splines were inverted at that point. Patient was admitted to the hospital and a pericardiocentesis was performed. Patient has been discharged.

    Patient problems: Pericardial Effusion

    MAUDE report 50240115 ↗
  • MalfunctionMar 20, 2026Material Integrity Problem

    IT WAS REPORTED THAT DURING THE PULSE FIELD ABLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. IT HAS BEEN MENTIONED THAT THE COATING CAME OFF IN THE MIDDLE OF VERSACROSS WIRE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING A DIFFERENT DEVICE (SAME MODEL, BOSTON SCIENTIFIC PRODUCT). NO PATIENT COMPLICATIONS HAVE BEEN REPORTED. THE PRODUCT IS EXPECTED TO BE RETURNED. IT WAS FURTHER REPORTED, THE DAMAGE WAS OBSERVED AFTER THE WIRE AND DILATOR WERE REMOVED. THE WIRE WAS PULLED BACK INTO THE SHEATH/ DILATOR ONCE THE SHEATH WAS IN THE SUPERIOR VENA CAVA (SVC). THE WIRE WAS THEN PULLED BACK AGAIN WHEN REMOVED. THE DILATOR WAS ALREADY BACKLOADED ONTO THE WIRE/INSIDE THE PATIENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24650221 ↗
  • InjuryNov 13, 2025Material Frayed; Material Integrity Problem; Material Frayed; Material Integrity Problem

    IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION SELECTED FOR USE. DURING PREPARATION, RF WIRE WAS NOTED TO BE FRAYED WHEN OPENED THE PACKAGE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING DIFFERENT DEVICE, SAME MODEL. NO PATIENT COMPLICATION WAS REPORTED. THE DEVICE IS NOT EXPECTED TO RETURN FOR ANALYSIS AS DISPOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23544914 ↗
  • MalfunctionOct 23, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT DURING FARAPULSE PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. DURING PREPARATION, IT WAS NOTED THAT THE BACK OF RF WIRE WAS PEELING OFF. THUS, THE PHYSICIAN OPTED TO OPEN A REPLACEMENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS OCCURRED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23373148 ↗

FDA clearance and classification

The FDA regulates product code OFK (percutaneous atrial catheter kit) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VersaCross™ Access Solution VXSK0121

Has the VersaCross™ Access Solution VXSK0121 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2103-2025 (Class II), began on Jun 10, 2025: Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the VersaCross™ Access Solution VXSK0121?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Oct 23, 2025 and Aug 13, 2026. In those reports, the most common specific device problems are “Material Integrity Problem” (3) and “Material Frayed” (1); the most common patient problem is “Pericardial Effusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VersaCross™ Access Solution VXSK0121?

None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VersaCross™ Access Solution VXSK0121, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand VersaCross™ Access Solution, model VXSK0121). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Percutaneous atrial catheter kit devices

All Percutaneous atrial catheter kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page