VersaCross™ Access Solution VXSK0122Recalls & reported problems

Boston Scientific Corporation · Access Solution · Class 2

All models of VersaCross Access Solution (15)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
18
Deaths reported
2
Injuries reported
10
Recalls naming it
1

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 00685447019824), received between May 20, 2025 and Sep 14, 2026 — classified as 2 deaths, 10 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 4 reportsJan 2026: 0 reportsFeb 2026: 3 reportsMar 2026: 3 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 10, 2025Class IIOpen, Classified

    Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

    Z-2103-2025 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem10
  • Material Integrity Problem3
  • Nonstandard Device3
  • Off-Label Use3
  • Device-Device Incompatibility2
  • Device Dislodged or Dislocated1
  • Difficult to Advance1
  • Failure to Advance1

Patient problems

  • Pericardial Effusion8
  • No Clinical Signs, Symptoms or Conditions6
  • Cardiac Perforation5
  • Low Blood Pressure/ Hypotension5
  • Cardiac Tamponade3
  • Non specific EKG/ECG Changes2
  • Asystole1
  • Bradycardia1

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that Thrombosis occurred.\\It was reported that a VersaCross RF Wire was selected to treat atrial fibrillation (Afib). During procedure, patients activated clotting times (ACTs) were not therapeutic and a clot formed on right side of the heart. They were 244 then down to 233 and then up to 360 at discovery of thrombus on wire in right atrium (RA). There were no device issues reported. Product return is not expected as it was disposed.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 50756610 ↗
  • DeathAug 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced cardiac tamponade, pericardial effusion, and perforation of the left atrial appendage. The patient died.During a pulse field ablation (PFA) procedure to treat paroxysmal atrial fibrillation a VersaCross Access Solution radiofrequency (RF) wire was selected for use. A VersaCross fixed sheath was used with the RF wire to perform the transseptal puncture (TSP). It was noted that the RF wire went across the septum without delivery of RF energy. The physician spent some time checking intracardiac echocardiography (ICE) imaging before dilating the septum with the VersaCross fixed sheath. It was noted that the RF wire did not appear as it normally did going out the left superior pulmonary vein (LSPV) on fluoroscopy, though it did not appear entirely abnormal either. The VersaCross fixed sheath was exchanged for the FARADRIVE. With the FARAWAVE catheter inserted past the sheath handle and into the clear shaft, the sheath was aspirated. Only air was aspirated from the sheath; the air was observed being drawn back from the tip of the FARAWAVE catheter. The FARAWAVE catheter was removed, but air continued to be aspirated. The sheath was exchanged,

    Patient problems: Cardiac Perforation; Cardiac Tamponade; Pericardial Effusion

    MAUDE report 50255873 ↗
  • InjuryJul 27, 2026Off-Label Use

    It was reported that a patient experienced a pericardial effusion growing to a cardiac tamponade following a cardiac perforation requiring a pericardiocentesis, exploratory open-heart surgery, medication, and extended hospitalization. The ablation procedure was unable to be completed due to this adverse event.\\A VersaCross Access Solutions kit was selected for use for the transseptal access portion of a FARAPULSE procedure to treat atrial fibrillation. Femoral access was achieved and a non-Boston Scientific (BSC) sheath and the Radio Frequency (RF) Pigtail wire from a fixed curve VersaCross kit were advanced into the Superior Vena Cava (SVC) with the use of Intracardiac Echocardiography (ICE). At this time, ICE showed a trace amount of effusion in the pericardial space. The non-BSC sheath was dropped down onto the atrial septum for transeptal access. Access was gained on the mid septum with veins in view. The RF Pigtail wire was advanced into the left superior pulmonary vein (LSPV) and the non-BSC sheath was advanced into the left atrium. The non-BSC dilator and RF Pigtail wire were removed. The non-BSC sheath was aspirated and flush. A Merit guidewire and non-BSC mapping catheter

    Patient problems: Cardiac Perforation; Cardiac Tamponade; Pericardial Effusion

    MAUDE report 50000315 ↗
  • InjuryJul 12, 2026Off-Label Use; Adverse Event Without Identified Device or Use Problem

    It was reported that during an Atrial Fibrillation Ablation an RF Wire was selected for use. During procedure, the Versacross wire was used to go transseptal through a non - Boston Scientific dilator and sheath. The dilator and wire were then removed; the sheath was aspirated then input the FARAWAVE catheter and aspirated again upon insertion of the FARAWAVE. Once the FARAWAVE was exposed on the left atrium, it was noted that the patient was experiencing ST Segment Elevation that was not present at baseline. A nitroglycerin dose was ordered and waited for a couple minutes prior to continuing. When the ST segment elevation had begun to decrease, the ablation procedure was continued. The patients ECG came back to baseline during the case, and the procedure was completed successfully.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 25791592 ↗
  • MalfunctionMar 25, 2026Failure to Advance

    THE RF WIRE WAS SELECTED FOR USE DURING THE PROCEDURE TO TREAT ATRIAL FIBRILLATION (A FIB). IT WAS REPORTED THAT WHEN ATTEMPTING TO GO TRANSEPTAL, THE WIRE WOULD NOT GO ACROSS AFTER RF ENERGY WAS DELIVERED. A NEW VERSACROSS WAS OPENED BUT ISSUE PERSISTED. REPLACING THE RF WIRE RESOLVED THE ISSUE. NO PATIENT COMPLICATIONS OCCURRED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE DEVICE HAS BEEN RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24686141 ↗
  • InjuryMar 11, 2026Difficult to Advance; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PERICARDIAL EFFUSION (PE) OCCURRED. IT WAS REPORTED THAT DURING AN ATRIAL FIBRILLATION PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THERE WAS A BASELINE SMALL EFFUSION SEEN ON TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE). THE PATIENT HAD TO BE SUPPORTED FOR LOW BLOOD PRESSURE ALL THROUGHOUT THE CASE WHERE ANESTHESIA HAD TO KEEP INCREASING THE PRESSERS. AFTER ABLATING THE RIGHT INFERIOR PULMONARY VEIN (RIPV), THE PHYSICIAN CHECKED ON INTRACARDIAC ECHOCARDIOGRAPHY (ICE) FOR PERICARDIAL EFFUSION AND SAW IT HAD GROWN SUBSTANTIALLY AS COMPARED TO BASELINE. 46 ABLATIONS HAD BEEN PERFORMED. A PERICARDIOCENTESIS WAS PERFORMED. THE PATIENT WAS ADMITTED TO INTENSIVE CARE UNIT AND EXPECTED TO FULLY RECOVER.

    Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 24571717 ↗
  • MalfunctionMar 9, 2026Material Integrity Problem

    IT WAS REPORTED THAT THE COATING ON THE WIRE APPEARED TO BE PEELING OFF. THE VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE FOR TRANSSEPTAL. THE DEVICE DID NOT ENTER THE BODY WITH THE WIRE. THE WIRE WAS REPLACED, AND THE CASE WAS SUCCESSFULLY COMPLETED. NO PATIENT COMPLICATIONS OCCURRED. THE DEVICE IS NOT RETURNING DUE TO DISPOSAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24544035 ↗
  • MalfunctionFeb 19, 2026Nonstandard Device

    IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. IT HAS BEEN MENTIONED THAT THE PHYSICIAN WAS ATTEMPTING TO INTRODUCE THE DILATOR AND SHEATH COMBINATION INTO THE PATIENT'S BODY BUT WAS MET WITH FORCE. WHEN THE DEVICE WAS REMOVED, IT WAS NOTED THAT THE DISTAL END OF THE SHEATH HAD FOLDED/CURVED BACK OVER THE DILATOR. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING A NEW VERSACROSS SYSTEM. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED. THE PRODUCT IS NOT EXPECTED TO BE RETURNING AS IT WAS DISPOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24390898 ↗
  • MalfunctionFeb 13, 2026Nonstandard Device

    IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. DURING THE PROCEDURE WHEN AN ATTEMPT WAS MADE FOR FEMORAL VEIN ACCESS FROM THE GROIN, THE DILATOR HAD RESISTANCE AND COULD NOT BE ADVANCED INTO THE VEIN. WHEN IT WAS WITHDRAWN, THE DISTAL END WAS FOUND TO BE ROLLED BACK AT THE END LIKE A SLEEVE. THE DILATOR WAS NOT FULLY INTRODUCED INTO THE VESSEL. THUS, THE DEVICE WAS REPLACED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATION WAS REPORTED. THE DEVICE IS NOT EXPECTED TO BE RETURN AS DISPOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24333558 ↗
  • InjuryFeb 2, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PERICARDIAL EFFUSION OCCURRED. DURING A PULMONARY VEIN ISOLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION, AN VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. AFTER COMPLETING THE TRANSSEPTAL PUNCTURE, THE PHYSICIAN ADVANCED THE VERSACROSS WIRE INTO THE LEFT ATRIUM, RESULTING IN A PERFORATION OF THE LEFT ATRIAL APPENDAGE. THE PATIENT'S BLOOD PRESSURE AND PULSE DROPPED SIGNIFICANTLY, AND A PERICARDIAL EFFUSION WAS CONFIRMED USING ICE AND FLUOROSCOPY. A PERICARDIOCENTESIS WAS PERFORMED, AND THE PATIENT WAS STABILIZED. THE PATIENT WAS ADMITTED TO THE INTENSIVE UNIT OF CARE (ICU). THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR INVESTIGATION, AS IT WAS DISPOSED OF AT THE FACILITY.

    Patient problems: Bradycardia; Cardiac Perforation; Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 24241550 ↗
  • InjuryDec 23, 2025Off-Label Use; Adverse Event Without Identified Device or Use Problem; Off-Label Use; Adverse Event Without Identified Device or Use Problem

    THE VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE DURING THE PULSE FIELD ABLATION PROCEDURE TO TREAT ATRIAL FIBRILLATION (A FIB) AND ATRIAL FLUTTER. IT WAS REPORTED THAT DURING THE TRANSEPTAL PUNCTURE, THE AORTA WAS PUNCTURED AND THE VERSACROSS WIRE WAS VISUALIZED IN THE AORTA ON ICE. THE PHYSICIAN CONTINUALLY VIEWED ICE IMAGES AND DETERMINED THERE WAS A SMALL AREA OF BLEEDING BACK FROM THE AORTA TO THE RIGHT ATRIUM, BUT NO AORTIC HEMATOMA DEVELOPED. THE SHEATH AND DILATOR WERE NEVER ADVANCED AND THEREFORE ONLY THE WIRE WENT INTO THE AORTA. AT THIS POINT, THE PHYSICIAN HAD THE HEPARIN INFUSION STOPPED AND GAVE THE PATIENT PROTAMINE. THE RF APPLICATION THROUGH A NON-BSC DEVICE WAS DELIVERED AT 12:05 AND THE WIRE WAS PULLED OUT OF THE AORTA AT 12:20. THE PATIENT REMAINED STABLE AND WAS MONITORED FOR AN HOUR PRIOR TO ENDING THE PROCEDURE AND CLOSING THE GROIN ACCESS SITE. THE PATIENT WAS SET TO STAY OVERNIGHT AND WAS ORDERED A LIMITED ECHOGRAM. THE PATIENT REMAINED STABLE THROUGHOUT THE MONITORING PERIOD AND THE PHYSICIAN PLANS TO BRING THE PATIENT BACK AT A LATER DATE TO COMPLETE THE ABLATION PROCEDURE. THE DEVICE HAS BEEN RETURNED FOR ANALYSIS.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Needle Stick/Puncture; Perforation of Vessels

    MAUDE report 23879424 ↗
  • InjuryDec 19, 2025Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Positioning Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Positioning Problem

    IT WAS REPORTED THAT A CARDIAC PERFORATION AND PERICARDIAL EFFUSION (PE) OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A CONCOMITANT ATRIAL FIBRILLATION ABLATION AND WATCHMAN PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THERE WAS A TRACE EFFUSION PRIOR TO THE PROCEDURE. THE ATRIAL FIBRILLATION ABLATION PERFORMED WITH A NON-BOSTON SCIENTIFIC (AFFERA) SYSTEM. THE ABLATION WAS COMPLETED WITHOUT COMPLICATION. AFTERWARD, THE PHYSICIAN ELECTED TO PERFORM A SHEATH EXCHANGE AND INITIALLY PLACED A SHORT 16F INTRODUCER SHEATH. DURING THIS PROCESS, THE WIRE WAS INADVERTENTLY WITHDRAWN FROM THE LEFT ATRIUM, RESULTING IN LOSS OF TRANSSEPTAL ACCESS. THE PHYSICIAN RE-CROSSED THE SEPTUM USING THE VERSACROSS FIXED-CURVE SYSTEM UNDER 2D INTRACARDIAC ECHOCARDIOGRAPHY (ICE) GUIDANCE. AS THE TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) PHYSICIAN HAD NOT YET ARRIVED, THE SHEATH EXCHANGE PROCEEDED USING ICE ALONE. THE WIRE APPEARED APPROPRIATELY POSITIONED IN THE LEFT SUPERIOR PULMONARY VEIN DURING THE EXCHANGE ACCORDING TO THE ICE VIEW, AFTER WHICH THE WIRE WAS REMOVED AND A NON-BOSTON SCIENTIFIC PIGTAIL CATHETER WAS INSERTED. AT NO POINT DID THE WATCHMAN DOUBLE-CURVE SHEATH ENTER THE LEFT ATRI

    Patient problems: Cardiac Perforation; Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 23854561 ↗
  • MalfunctionDec 15, 2025Nonstandard Device; Physical Resistance/Sticking; Nonstandard Device; Physical Resistance/Sticking

    DURING AN ATRIAL FIBRILLATION PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THE VERSACROSS ACCESS SOLUTION WAS ADVANCED INTO THE SUPERIOR VENA CAVA AND THE SHEATH/DILATOR WAS LOADED ONTO THE WIRE. THE PHYSICIAN WAS ABLE TO ADVANCE THE DILATOR INTO THE FEMORAL ACCESS SITE BUT GOT STUCK AT THE SHEATH/DILATOR TRANSITION AND COULD NOT ADVANCE FURTHER. THE SHEATH WAS REMOVED FROM THE WIRE AND IT WAS OBSERVED THAT THE TIP OF THE SHEATH WAS STRETCHED AND BUCKLING BACKWARDS FROM THE DILATOR. A NEW VERSACROSS KIT WAS OPENED AND THE NEW SHEATH WAS ABLE TO ADVANCE THROUGH THE ACCESS SITE EASILY. THE DEVICE WILL NOT BE AVAILABLE FOR RETURN DUE TO DISPOSAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23804064 ↗
  • MalfunctionDec 11, 2025Device-Device Incompatibility; Material Integrity Problem; Device-Device Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT THE RF WIRE COATING WAS FRAYED ON THE PROXIMAL END AND A SHEATH COULD NOT BE LOADED ONTO IT. DURING A PULSE FIELD ABLATION (PFA) PULMONARY VEIN ISOLATION (PVI) AND POSTERIOR WALL (PW) ABLATION PROCEDURE TO TREAT PERSISTENT ATRIAL FIBRILLATION A VERSACROSS RF WIRE WAS USED. THE RF WIRE WAS PREPARED AND A VERSACROSS FIXED SHEATH WAS ADVANCED INTO THE BODY OVER THE WIRE. THE TRANSSEPTAL PUNCTURE WAS PERFORMED SUCCESSFULLY. THE VERSACROSS SHEATH WAS BACKED OUT OF THE PATIENT OVER THE RF WIRE, AND AN ATTEMPT WAS MADE TO LOAD THE FARADRIVE SHEATH OVER THE WIRE, BUT THE FARADRIVE DILATOR WOULD NOT ENGAGE WITH THE WIRE. AFTER A SECOND ATTEMPT TO LOAD THE FARADRIVE SHEATH ONTO THE RF WIRE IT WAS NOTICED THAT THE PROXIMAL END APPEARED FRAYED, AND THAT A PIECE OF THE FRAYED COATING COULD BECOME DISLODGED AND EMBOLIZE IF ATTEMPTS IF A SHEATH WERE FORCED OVER IT. STERILE CLAMPS WERE USED ATTEMPTING TO REMOVE THE FRAYED COATING, BUT THIS WAS UNSUCCESSFUL AND ONLY FURTHER MANGLED THE COATING. RATHER THAN REMOVING THE RF WIRE AND LOSING TRANSSEPTAL ACCESS, A DECISION WAS MADE TO USE NON-STERILE WIRE CUTTERS TO CLIP THE FRAYED END OF THE WIRE OFF ENTIRELY. THE PROXIMAL END OF T

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23777690 ↗
  • InjuryOct 28, 2025Device-Device Incompatibility; Material Integrity Problem; Device-Device Incompatibility; Material Integrity Problem

    IT WAS REPORTED THAT ST SEGMENT ELEVATION OCCURRED. DURING AN ATRIAL FIBRILLATION/ATRIAL FLUTTER PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THE PHYSICIAN PERFORMED TRANSSEPTAL PUNCTURE, FARADRIVE SHEATH EXCHANGE TO ADVANCE OVER LEFT ATRIUM WAS DIFFICULT BECAUSE THE VERSACROSS WIRE INSULATION FROM PROXIMAL END WAS FRAYED. FARAPULSE PULSE FIELD ABLATION WAS APPLIED USING FARAWAVE NAV TO COMPLETE PULMONARY VEIN ISOLATION AND POSTERIOR WALL ISOLATION. AN ATRIAL TACHYCARDIA WAS ABLATED WITH THE FARAWAVE ANTERIOR AND INFERIOR TO THE LEFT INFERIOR PULMONARY VEIN (LIPV), NEAR THE MITRAL VALVE ISTHMUS AND LIKELY ADJACENT TO THE CIRCUMFLEX CORONARY ARTERY. PATIENT'S ECG DID NOT HAVE NOTICEABLE ST SEGMENT CHANGES AFTER ABLATION DELIVERY. PATIENT WAS ATRIAL-PACED TO INDUCE ATRIAL TACHYCARDIA. ISOPREL WAS INFUSED AND PATIENT WAS CONTINUALLY PACED. ISOPREL INFUSION WAS INCREASED, PATIENT WAS PACED, AND PHYSICIAN NOTICED ST SEGMENT ELEVATION. ISUPREL AND PACING WERE TERMINATED AND A 100MCG NITROGLYCERINE WAS ADMINISTERED THROUGH THE FARADRIVE SHEATH AND A SUBSEQUENT 100MCG WERE ADMINISTERED VIA IV. ST SEGMENT ELEVATION RESOLVED AFTER THE SECOND DOSE OF NITROGLYCERINE. THE PATIE

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 23408354 ↗
  • InjuryOct 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS PERICARDIAL EFFUSION OCCURRED. THE PROCEDURE WAS ABORTED. IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION WAS SELECTED FOR LAAC PROCEDURE. DURING THE PROCEDURE, THE PATIENT WAS PLACED UNDER ANESTHESIA, AND THE PROCEDURE WAS BEGUN. THE PHYSICIAN WAS GOT AN ACCESS AND ASKED FOR 5,000 UNITS OF HEPARIN. TRANSEPTAL POSITION WAS VIEWED IN SHORT AXIS AND BICAVAL VIEWS AND TRANSEPTAL WAS PERFORMED IN A MID/ INFERIOR POSITION USING THE VERSACROSS TRANSEPTAL SHEATH AND WIRE. POST TRANSEPTAL, THE PATIENTS EXPERIENCED A DROP IN BLOOD PRESSURE WITH PERICARDIAL EFFUSION NOTED BY CARDIAC IMAGER. PATIENT WAS MONITORED BY ANESTHESIA, THE PHYSICIAN AND CARDIAC IMAGER. HEPARIN WAS REVERSED WITH PROTAMINE FOR THAT PHYSICIAN DECIDED AGAINST DOING A PERICARDIOCENTESIS. THE PROCEDURE WAS ABORTED AND PATIENT DID NOT RECEIVE THE WATCHMAN IMPLANT. THE PATIENT IS EXPECTED TO FULLY BE RECOVERED AS PATIENT STAYED OVERNIGHT AS PLANNED TO DISCHARGE ON (B)(6) 2025. THERE WAS A MINOR PERICARDIAL EFFUSION NOTED PRIOR, NOTED UNDER THE RIGHT VENTRICLE ON TEE. THE EFFUSION INCREASED IN SIZE AFTER TRANSEPTAL PUNCTURE. THE DEVICE IS NOT EXPECTED TO RETURN AS DISCARDED. IT WAS FURTHER REPORTED THAT THE VERSACROSS

    Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 23294735 ↗
  • DeathJun 2, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PATIENT DEATH OCCURRED. DURING A PULMONARY VEIN ISOLATION ABLATION PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. THE PHYSICIAN OBTAINED ACCESS IN BOTH RIGHT AND LEFT FEMORAL VEIN. INTRACARDIAC ECHOCARDIOGRAPHY (ICE) IMAGINING WAS INSERTED IN LEFT FEMORAL VEIN. THE GUIDEWIRE WAS ADVANCED TO SUPERIOR VENA CAVA (SVC) AND THEN THE VERSACROSS TRANSSEPTAL SHEATH AS WELL AS THE RF WIRE. THE TRANSSEPTAL PUNCTURE (TSP) AND THE RF WIRE WERE ADVANCED IN THE LEFT ATRIUM. THERE WAS NO DIFFICULTY NOTED WITH THE TSP WORKFLOW. THE PATIENT HAD AN INCREASED HEART RATE OF 215 BPM THAT SUSTAINED FOR A MINUTE. THEN PERICARDIAL EFFUSION (PE) WAS NOTED ON ICE IMAGING AND A PERICARDIOCENTESIS BEGAN. CARDIOPULMONARY RESUSCITATION (CPR) WAS STARTED AND CODE BLUE ALARM WAS CALLED. THE PATIENT WAS ASYSTOLE. THE PHYSICIAN OPENED PATIENT CHEST, AND BLOOD TRANSFUSION BEGAN. CARDIAC MESSAGE WAS PERFORMED. SODIUM BICARB WAS GIVEN THROUGH THE SHEATH. VASOPRESSORS GIVEN BY IC PHYSICIAN. AORTIC CANNULATION WAS ADMINISTERED AND VENOUS PERFUSION BEGAN. THE BYPASS SURGERY WAS INITIATED. A 90 VOLT SHOCK ADMINISTERED TO HEART, HOWEVER LATER THE PATIENT WAS DECLARED DEAD. THE AUTOPSY H

    Patient problems: Asystole; Cardiac Arrest; Cardiac Perforation; Pericardial Effusion; Tachycardia

    MAUDE report 22123765 ↗
  • InjuryMay 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A CARDIAC TAMPONADE OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A WATCHMAN LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. THE PHYSICIAN USING SOLO OPERATOR TECHNIQUE USING TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE). THE STAFF WAS STILL NOT PREPARED TO INITIATE THE ENERGY FOR TRANSSEPTAL PUNCTURE AS THEY STILL HAD TO APPLY GROUNDING PAD. THE PHYSICIAN LOST VISUALIZATION OF THE INTRA ATRIAL SEPTUM, THEN GOT POOR QUALITY IMAGES AND USED ENERGY. THE WIRE DID NOT LOOK NORMAL THUS, IT WAS REMOVED. IT WAS NOTED THE ARTERIAL LINE PRESSURES DECREASED AND EFFUSION WAS PRESENT IN THE LEFT VENTRICULAR (THE INOCULATED BEHIND LEFT ATRIUM). THE PATIENT WAS TAPPED FOR TAMPONADE AND 400 ML TAKEN OFF AND NO MORE ACCUMULATION. THE PATIENT WAS FULLY RECOVERED AND WAS DISCHARGED NEXT DAY FROM THE HOSPITAL (B)(6) 2025). THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED).

    Patient problems: Cardiac Tamponade; Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 22052961 ↗

FDA clearance and classification

The FDA regulates product code OFK (percutaneous atrial catheter kit) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VersaCross™ Access Solution VXSK0122

Has the VersaCross™ Access Solution VXSK0122 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2103-2025 (Class II), began on Jun 10, 2025: Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the VersaCross™ Access Solution VXSK0122?

The FDA's MAUDE database lists 18 adverse event reports for this device, received between May 20, 2025 and Sep 14, 2026. In those reports, the most common specific device problems are “Material Integrity Problem” (3), “Nonstandard Device” (3) and “Off-Label Use” (3); the most common patient problems are “Pericardial Effusion” (8), “Cardiac Perforation” (5) and “Low Blood Pressure/ Hypotension” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VersaCross™ Access Solution VXSK0122?

2 reports are classified as deaths, 10 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VersaCross™ Access Solution VXSK0122, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand VersaCross™ Access Solution, model VXSK0122). Its record shows it as in commercial distribution.