VersaCross™ Steerable Access Solution VSTK0115Recalls & reported problems
Boston Scientific Corporation · Steerable Access Solution · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00685447019657), received between Nov 24, 2025 and Sep 18, 2026 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Delivered as Unsterile Product1
- Fluid/Blood Leak1
- Material Integrity Problem1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2026Fluid/Blood Leak
It was reported that during a Pulsed Field Ablation (PFA) procedure for the treatment of atypical atrial flutter, a VersaCross Steerable Access Solution was selected for use. The dilator was inserted into the sheath without resistance, followed by advancement of the FARAPOINT catheter and ORION mapping catheter across the valve. No resistance was encountered during insertion of the ORION mapping catheter. During the procedure, the physician observed significant leakage from the sheath valve, described as a fast drip of saline during flushing. The syringe was securely attached to the handle luer, and no damage to the device or luer was noted. There was no air leakage or air observed in the patient, and no error messages were displayed. No RF generator was used. The procedure was successfully completed without troubleshooting. The patient remained stable, and no patient complications were reported as a result of the event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50819349 ↗MalfunctionApr 13, 2026Delivered as Unsterile Product; Unsealed Device Packaging
IT WAS REPORTED THAT PRIOR TO AN ATRIAL FIBRILLATION PROCEDURE, PULSE FIELD ABLATION PROCEDURE A VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR USE. UPON OPENING THE VERSACROSS PACKAGE, IT WAS NOTED TO HAVE A PARTIALLY OPEN SEAL. THE DEVICE WAS NOT STERILE AND DISCARDED. THE PROCEDURE WAS CANCELLED. NO PATIENT WAS NOTED TO BE PRESENT AT THE TIME OF THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865662 ↗MalfunctionApr 1, 2026Material Integrity Problem
IF WAS REPORTED THAT DURING A PULSED FIELD ABLATION (PFA) PROCEDURE WITH VARIPULSE TO TREAT AN ATRIAL FIBRILLATION, THE COATING OF THE VERSACROSS WIRE WAS COMING OFF AND EXPOSED AT PROXIMAL END OF THE WIRE. A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760571 ↗MalfunctionNov 24, 2025Break; Break
IT WAS REPORTED THAT DURING THE PULMONARY VEIN ISOLATION TO TREAT ATRIAL FIBRILLATION VERSACROSS STEERABLE ACCESS SOLUTION WERE SELECTED FOR USE. IT HAS BEEN MENTIONED THAT BEFORE THE INSERTION STOPCOCK BROKE OFF. THE PROCEDURE WAS COMPLETED USING DIFFERENT DEVICE (SAME MODEL). THE PRODUCT IS NOT EXPECTED TO RETURN AS DISPOSED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23638294 ↗
Other Percutaneous atrial catheter kit devices
- VersaCross Connect Access Solution for FARADRIVE VXAK0041Boston Scientific Corporation · Percutaneous Atrial Catheter Kit3 deaths275 reports
- VersaCross Connect™ LAAC Access Solution VXAK0107Boston Scientific Corporation · Percutaneous atrial catheter kit3 deaths77 reports
- VersaCross™ Access Solution VXSK0122Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths21 reports
- VersaCross™ Access Solution VXSK0131Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths12 reports
- VersaCross™ Access Solution VXSK0134Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall1 death11 reports
- VersaCross™ Access Solution VXSK0132Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall11 reports
- VersaCross™ Access Solution VXSK0121Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall6 reports
- VersaCross™ Access Solution VXSK0124Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.