VersaCross™ Steerable Access Solution VSTK0116Recalls & reported problems

Boston Scientific Corporation · Steerable Access Solution · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00685447019664), received between Oct 17, 2025 and Feb 20, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Advance2
  • Material Integrity Problem2
  • Failure to Advance1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Pericardial Effusion2
  • Low Blood Pressure/ Hypotension1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 20, 2026Failure to Advance; Difficult to Advance

    IT WAS REPORTED THAT A PERICARDIAL EFFUSION (PE) OCCURRED. DURING A CAVOTRICUSPID ISTHMUS (CTI)/ PAROXYSMAL ATRIAL FIBRILLATION PROCEDURE, A VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR USE. THE PHYSICIAN COMPLETED RF CTI LINE WITH A NON-BOSTON SCIENTIFIC RF CATHETER. ROUGHLY 15-20 ABLATIONS NEEDED TO ACHIEVE CTI BLOCK. PHYSICIAN THEN ATTEMPTED TO CROSS THE SEPTUM WITH VERSACROSS STEERABLE THAT WAS USED FOR THE CTI ABLATION HOWEVER, WAS UNABLE TO CROSS THE SEPTUM WITH DILATOR AND SHEATH. IT WAS NOTED A BLOOD PRESSURE DROP TO WHICH THE PHYSICIAN USED INTRACARDIAC ECHOCARDIOGRAPHY (ICE) AND NOTED EFFUSION WAS PRESENT. A CARDIOLOGIST AND CARDIOTHORACIC (CT) WERE CONSULTED. A PERICARDIOCENTESIS WAS ATTEMPTED TO DRAW FLUID WITHOUT SUCCESS AND CT OPENED A WINDOW TO REMOVE BLOOD. THE PATIENT WAS INTUBATED AND MOVED TO INTENSIVE CARE UNIT FOR RECOVERY. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED). IT WAS FURTHER REPORTED THAT RF WIRE APPEARED TO CROSS THE SEPTUM, HOWEVER WIRE MANEUVERABILITY WAS LIMITED AFTER CROSSING DUE TO PATENT FORAMEN OVALE (PFO) CLOSURE BOVINE PLUG USED IN PREVIOUS PROCEDURE. PREVIOUS PFO CLOSURE USING A BOVINE PLUG WAS KNOWN ANATOMICAL

    Patient problems: Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 24401725 ↗
  • InjuryNov 4, 2025Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT A PERICARDIAL EFFUSION (PE) OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A WATCHMAN PROCEDURE, A VERSACROSS LARGE ACCESS SOLUTION AND A VERSACROSS ACCESS STEERABLE SOLUTION WERE SELECTED FOR USE. THE PHYSICIAN WAS ATTEMPTING TO GAIN TRANSEPTAL ACCESS WITH DIFFICULTY ADEQUATELY IMAGING THE TRANSEPTAL LOCATION VIA TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE). THE PHYSICIAN WAS UTILIZING VERSACROSS LARGE ACCESS INITIALLY THEN SWITCHED OUT TO VERSACROSS STEERABLE DUE TO HAVING ISSUES REACHING THE TRANSEPTAL LOCATION DUE TO PATIENT ANATOMY AND DEVICE REACH TO THE SEPTAL BY ADDING ADDITIONAL CURVE TO THE TRANSEPTAL SYSTEM. UPON OBTAINING A POSITION ON THE FOSSA VIA TEE, THE PHYSICIAN ADVISED TO ACTIVATE RF TO GO TRANSEPTAL. THE RF WIRE ADVANCED EASILY HOWEVER PHYSICIAN EXPERIENCED DIFFICULTY ADVANCING THE STEERABLE SHEATH ACROSS THE SEPTUM. THE PHYSICIAN VERIFIED TRANSEPTAL LOCATION AND INSERTED THE WATCHMAN TRUSTEER SHEATH. UPON INSERTING THE PIGTAIL CATHETER THROUGH THE SHEATH AND TAKING AN ANGIOGRAM, IT WAS NOTED CONTRAST STAINING IN THE PERICARDIAL SPACE. AT THAT TIME, IT WAS CONFIRMED VIA TEE THAT THEY WERE IN THE PERICARDIAL SPACE. THE PATIENT REMAINED STABLE AND A SM

    Patient problems: Pericardial Effusion

    MAUDE report 23468586 ↗
  • MalfunctionOct 17, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR USE. IT WAS NOTED THAT A PIECE OF VERSACROSS RF WIRE SHEARED OFF IN THE MIDDLE OF THE CASE. NO FURTHER DETAILS PROVIDED. NO PATIENT COMPLICATIONS REPORTED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS. MEDSUN# (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23329628 ↗
  • MalfunctionOct 17, 2025Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR USE. IT WAS NOTED THAT THE VERSACROSS WIRE SHEARED OFF. NO FURTHER DETAILS PROVIDED. NO PATIENT COMPLICATIONS REPORTED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS. MEDSUN (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23329052 ↗

FDA clearance and classification

The FDA regulates product code OFK (percutaneous atrial catheter kit) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VersaCross™ Steerable Access Solution VSTK0116

Has the VersaCross™ Steerable Access Solution VSTK0116 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00685447019664), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VersaCross™ Steerable Access Solution VSTK0116?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Oct 17, 2025 and Feb 20, 2026. In those reports, the most common specific device problems are “Difficult to Advance” (2), “Material Integrity Problem” (2) and “Failure to Advance” (1); the most common patient problems are “Pericardial Effusion” (2) and “Low Blood Pressure/ Hypotension” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VersaCross™ Steerable Access Solution VSTK0116?

None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VersaCross™ Steerable Access Solution VSTK0116, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand VersaCross™ Steerable Access Solution, model VSTK0116). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.