Skip to content

VersaCross™ Steerable Access Solution VSTK0122Recalls & reported problems

Boston Scientific Corporation · Steerable Access Solution · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00685447019718), received between Aug 19, 2025 and Sep 9, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Failure to Advance” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Advance1
  • Material Perforation1
  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Cardiac Perforation1
  • Pericardial Effusion1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2026Material Perforation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50695240 ↗
  • MalfunctionFeb 3, 2026Nonstandard Device

    IT WAS REPORTED THAT A VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR USE IN A PAROXYSMAL ATRIAL FIBRILLATION PROCEDURE. DURING THE PROCEDURE, DILATOR WAS WITHDRAWN FROM THE SHEATH AND WIRE WAS THEN READVANCED, BUT DILATOR WAS UNABLE TO PASS BEYOND THE TIP OF THE SHEATH. IT APPEARS THAT WIRE ENTERED AN IRRIGATION HOLE AT THE DISTAL END OF THE SHEATH, AND DILATOR FOLLOWED INTO THE IRRIGATION PORT. SHEATH AND DILATOR WERE REMOVED FROM THE BODY, AND IT WAS FOUND THAT DISTAL IRRIGATION HOLE OF THE SHEATH WAS ENLARGED. AT THAT POINT, THE SHEATH WAS REPLACED AND PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATION WAS REPORTED. THE SHEATH IS NOT EXPECTED TO BE RETURN FOR ANALYSIS AS IT WAS DISPOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251039 ↗
  • InjuryAug 19, 2025Failure to Advance; Failure to Advance

    IT WAS REPORTED THAT A CARDIAC PERFORATION AND PERICARDIAL EFFUSION (PE) OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A ATRIAL FIBRILLATION ABLATION (PULMONARY VEIN ISOLATION) REDO PROCEDURE, A VERSACROSS STEERABLE ACCESS SOLUTION WAS SELECTED FOR USE. DURING TRANSSEPTAL, THERE WERE FIVE (5) ATTEMPTS WERE MADE USING RADIO FREQUENCY AT ONE SECOND CONSTANT SETTING TO CROSS THE SEPTUM. THIS WAS UNSUCCESSFUL. A PERICARDIAL EFFUSION WAS NOTED IMMEDIATELY AND THE PHYSICIAN PERFORMED AN EMERGENCY PERICARDIOCENTESIS. A 500 ML OF BRIGHT BLOOD WAS REMOVED FROM THE PERICARDIUM. THE PATIENT WAS NOT IN CARDIAC ARREST. ONCE STABILIZED, THE PATIENT WAS EXTUBATED AND TRANSFERRED TO THE CARDIAC INTENSIVE CARE UNIT WITH A PERICARDIAL DRAIN. THE FOLLOWING DAY, THE DRAIN WAS REMOVED AND FOLLOW-UP ECHOCARDIOGRAM WAS STABLE. THE PATIENT WAS DISCHARGED HOME AND ABLATION PROCEDURE WILL BE RESCHEDULED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED). IN THE PHYSICIAN'S OPINION, THE ACCESS SOLUTION DEVICES DID NOT CONTRIBUTED TO THE PATIENT COMPLICATIONS.

    Patient problems: Cardiac Perforation; Pericardial Effusion

    MAUDE report 22841675 ↗

FDA clearance and classification

The FDA regulates product code OFK (percutaneous atrial catheter kit) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VersaCross™ Steerable Access Solution VSTK0122

Has the VersaCross™ Steerable Access Solution VSTK0122 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00685447019718), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VersaCross™ Steerable Access Solution VSTK0122?

The FDA's MAUDE database lists 3 adverse event reports for this device, received between Aug 19, 2025 and Sep 9, 2026. In those reports, the most common specific device problems are “Failure to Advance” (1), “Material Perforation” (1) and “Nonstandard Device” (1); the most common patient problems are “Cardiac Perforation” (1) and “Pericardial Effusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VersaCross™ Steerable Access Solution VSTK0122?

None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VersaCross™ Steerable Access Solution VSTK0122, and is it still sold?

The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand VersaCross™ Steerable Access Solution, model VSTK0122). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Percutaneous atrial catheter kit devices

All Percutaneous atrial catheter kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page