VersaCross™ Access Solution VXSK0101Recalls & reported problems
Boston Scientific Corporation · Access Solution · Class 2
All models of VersaCross Access Solution (15)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447019749), received between Feb 11, 2026 and Mar 30, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 10, 2025Class IIOpen, Classified
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Z-2103-2025 · Root cause: Packaging change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Difficult to Advance1
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Pericardial Effusion1
- Thrombosis/Thrombus1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Material Integrity Problem
IT WAS REPORTED THAT A VERSACROSS ACCESS SOLUTIONS KIT WAS SELECTED TO BE USED IN A PROCEDURE TO TREAT ATRIAL FIBRILLATION. JUST PRIOR TO INSERTION OF THE RF WIRE INTO THE PATIENT, AFTER THE WIRE HAD BEEN INSERTED INTO A SHORT SHEATH FOR SHEATH EXCHANGE, THE COATING WAS OBSERVED TO BE PEELING OFF THE DISTAL TIP. NO OTHER DAMAGE WAS NOTED TO ANY ACCESSORY DEVICES THAT COULD HAVE CONTRIBUTED TO THE COATING ISSUE. THE KIT WAS REPLACED WITH A SIMILAR KIT, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED. THE KIT WAS DISPOSED AND IS NOT EXPECTED TO RETURN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24720065 ↗InjuryFeb 11, 2026Difficult to Advance
IT WAS REPORTED THAT A THROMBOSIS AND PERICARDIAL EFFUSION (PE) OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A WATCHMAN LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THE PHYSICIAN WAS ATTEMPTING TO CROSS SEPTUM BUT WAS UNSUCCESSFUL THE FIRST FEW TIMES APPLYING ENERGY WITH ONE SECOND CONSTANT GENERATOR SETTING. ON THE FOURTH ATTEMPT, THE WIRE ADVANCED INTO THE LEFT ATRIUM (LA). THE PHYSICIAN FELT THAT THE WIRE CROSSED INTO THE LA THEN INTO PERICARDIUM. A PERICARDIAL EFFUSION (PE) WAS NOTED. THE IMPLANTER AND PHYSICIAN IMMEDIATELY SHIFTED TO OPENING A PERICARDIOCENTESIS TRAY AND TAPPING THE PATIENT. THE PATIENT WAS SENT TO SURGERY FOR WINDOW AND CLOT ASPIRATION. THE PATIENT WAS HOSPITALIZED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED). IT WAS FURTHER REPORTED THAT THE SEPTUM WAS ANEURYSMAL. PATIENT WAS SENT TO SURGERY AFTER STABILIZED TO CREATE A WINDOW TO REMOVE SOME RESIDUAL CLOT IN THE PERICARDIUM. NO MALFUNCTIONS OR CONTRIBUTIONS FROM VERSACROSS. THE PATIENT WAS LATER DISCHARGED.
Patient problems: Pericardial Effusion; Thrombosis/Thrombus
MAUDE report 24316708 ↗
FDA clearance and classification
The FDA regulates product code OFK (percutaneous atrial catheter kit) under 21 CFR 870.1250 as a Class II device. 21 CFR 870.1250 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VersaCross™ Access Solution VXSK0101
Has the VersaCross™ Access Solution VXSK0101 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2103-2025 (Class II), began on Jun 10, 2025: Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the VersaCross™ Access Solution VXSK0101?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 11, 2026 and Mar 30, 2026. In those reports, the most common specific device problems are “Difficult to Advance” (1) and “Material Integrity Problem” (1); the most common patient problems are “Pericardial Effusion” (1) and “Thrombosis/Thrombus” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VersaCross™ Access Solution VXSK0101?
None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VersaCross™ Access Solution VXSK0101, and is it still sold?
The FDA's device database lists Boston Scientific Corporation as the company responsible for it (brand VersaCross™ Access Solution, model VXSK0101). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Percutaneous atrial catheter kit devices
- VersaCross Connect Access Solution for FARADRIVE VXAK0041Boston Scientific Corporation · Percutaneous Atrial Catheter Kit3 deaths272 reports
- VersaCross Connect™ LAAC Access Solution VXAK0107Boston Scientific Corporation · Percutaneous atrial catheter kit3 deaths73 reports
- VersaCross™ Access Solution VXSK0122Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths18 reports
- VersaCross™ Access Solution VXSK0134Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall1 death11 reports
- VersaCross™ Access Solution VXSK0132Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall11 reports
- VersaCross™ Access Solution VXSK0131Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths10 reports
- VersaCross™ Access Solution VXSK0124Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall4 reports
- VersaCross™ Access Solution VXSK0121Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page