VersaCross™ Access Solution VXSK0101Recalls & reported problems

Boston Scientific Corporation · Access Solution · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00685447019749), received between Feb 11, 2026 and Mar 30, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 10, 2025Open, Classified

    Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

    Z-2103-2025 · Root cause: Packaging change control · FDA record ↗

Most reported problems

Device problems

  • Difficult to Advance1
  • Material Integrity Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Pericardial Effusion1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 30, 2026Material Integrity Problem

    IT WAS REPORTED THAT A VERSACROSS ACCESS SOLUTIONS KIT WAS SELECTED TO BE USED IN A PROCEDURE TO TREAT ATRIAL FIBRILLATION. JUST PRIOR TO INSERTION OF THE RF WIRE INTO THE PATIENT, AFTER THE WIRE HAD BEEN INSERTED INTO A SHORT SHEATH FOR SHEATH EXCHANGE, THE COATING WAS OBSERVED TO BE PEELING OFF THE DISTAL TIP. NO OTHER DAMAGE WAS NOTED TO ANY ACCESSORY DEVICES THAT COULD HAVE CONTRIBUTED TO THE COATING ISSUE. THE KIT WAS REPLACED WITH A SIMILAR KIT, AND THE PROCEDURE WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED. THE KIT WAS DISPOSED AND IS NOT EXPECTED TO RETURN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24720065 ↗
  • InjuryFeb 11, 2026Difficult to Advance

    IT WAS REPORTED THAT A THROMBOSIS AND PERICARDIAL EFFUSION (PE) OCCURRED. THE PROCEDURE WAS CANCELLED. DURING A WATCHMAN LEFT ATRIAL APPENDAGE CLOSURE (LAAC) PROCEDURE, A VERSACROSS ACCESS SOLUTION WAS SELECTED FOR USE. THE PHYSICIAN WAS ATTEMPTING TO CROSS SEPTUM BUT WAS UNSUCCESSFUL THE FIRST FEW TIMES APPLYING ENERGY WITH ONE SECOND CONSTANT GENERATOR SETTING. ON THE FOURTH ATTEMPT, THE WIRE ADVANCED INTO THE LEFT ATRIUM (LA). THE PHYSICIAN FELT THAT THE WIRE CROSSED INTO THE LA THEN INTO PERICARDIUM. A PERICARDIAL EFFUSION (PE) WAS NOTED. THE IMPLANTER AND PHYSICIAN IMMEDIATELY SHIFTED TO OPENING A PERICARDIOCENTESIS TRAY AND TAPPING THE PATIENT. THE PATIENT WAS SENT TO SURGERY FOR WINDOW AND CLOT ASPIRATION. THE PATIENT WAS HOSPITALIZED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS (DISPOSED). IT WAS FURTHER REPORTED THAT THE SEPTUM WAS ANEURYSMAL. PATIENT WAS SENT TO SURGERY AFTER STABILIZED TO CREATE A WINDOW TO REMOVE SOME RESIDUAL CLOT IN THE PERICARDIUM. NO MALFUNCTIONS OR CONTRIBUTIONS FROM VERSACROSS. THE PATIENT WAS LATER DISCHARGED.

    Patient problems: Pericardial Effusion; Thrombosis/Thrombus

    MAUDE report 24316708 ↗

Other Percutaneous atrial catheter kit devices

All Percutaneous atrial catheter kit devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.