VersaCross™ Access Solution VXSK0123Recalls & reported problems
Boston Scientific Corporation · Access Solution · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00685447019831), received between Jan 29, 2026 and Jan 29, 2026 — classified as 1 malfunction.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Jun 10, 2025Open, Classified
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
Z-2103-2025 · Root cause: Packaging change control · FDA record ↗
Most reported problems
Device problems
- Difficult to Advance1
- Material Integrity Problem1
- Nonstandard Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2026Nonstandard Device; Difficult to Advance; Material Integrity Problem
IT WAS REPORTED THAT VERSACROSS ACCESS SOLUTION WAS SELECTED FOR A LEFT ATRIUM APPENDAGE (LAA) LEAK REPAIR PRIOR TO A LAA PERFORMED. AFTER RF WIRE WAS INSERTED, THE DILATOR WAS ATTEMPTED TO BE INSERTED OVER THE WIRE, HOWEVER THE OUTER LAYER OF THE WIRE SHEARD FROM THE INNER CORE. THE RF WIRE WAS REPLACED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATION WAS REPORTED. THE DEVICE HAS BEEN RECEIVED AT BOSTON SCIENTIFIC'S POST MARKET LABORATORY WHERE IT IS AWAITING ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24215992 ↗
Other Percutaneous atrial catheter kit devices
- VersaCross Connect Access Solution for FARADRIVE VXAK0041Boston Scientific Corporation · Percutaneous Atrial Catheter Kit3 deaths275 reports
- VersaCross Connect™ LAAC Access Solution VXAK0107Boston Scientific Corporation · Percutaneous atrial catheter kit3 deaths77 reports
- VersaCross™ Access Solution VXSK0122Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths21 reports
- VersaCross™ Access Solution VXSK0131Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall2 deaths12 reports
- VersaCross™ Access Solution VXSK0134Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall1 death11 reports
- VersaCross™ Access Solution VXSK0132Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall11 reports
- VersaCross™ Access Solution VXSK0121Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall6 reports
- VersaCross™ Access Solution VXSK0124Boston Scientific Corporation · Percutaneous atrial catheter kit1 recall5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.