Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00690103187014), received between Sep 9, 2024 and Jan 9, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Fluid/Blood Leak1
Patient problems
- Hemorrhage/Blood Loss/Bleeding3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING USE IN A PATIENT IN THE NEUROLOGICAL ICU (INTENSIVE CARE UNIT), WHEN HANDLING THE MANIFOLD OF THIS VOLUMEVIEW SENSOR, IT BROKE INTO TWO PIECES. THE DEVICE WAS REPLACED BY A NEW ONE. THE BLOOD LOSS WAS MINIMAL. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION SINCE IT WAS DISCARDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21112232 ↗MalfunctionJan 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING USE IN A PATIENT IN THE CARDIAC ICU (INTENSIVE CARE UNIT), WHEN HANDLING THE MANIFOLD OF THIS VOLUME VIEW SENSOR, IT BROKE INTO TWO PIECES (COMPLAINT (B)(4) / MFR 2015691-2025-00170). THE DEVICE WAS REPLACED BY A NEW ONE AND, AFTER TWO DAYS OF USE, THE MANIFOLD ALSO BROKE INTO TWO PIECES (COMPLAINT (B)(4). A THIRD DEVICE WAS USED. THE BLOOD LOSS WAS MINIMAL. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICES WERE NOT AVAILABLE FOR EVALUATION SINCE THEY WERE DISCARDED; HOWEVER, AN IMAGE OF THE DEVICE INVOLVED IN THE COMPLAINT (B)(4) WAS AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21111941 ↗MalfunctionJan 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, DURING USE IN A PATIENT IN THE CARDIAC ICU (INTENSIVE CARE UNIT), WHEN HANDLING THE MANIFOLD OF THIS VOLUMEVIEW SENSOR, IT BROKE INTO TWO PIECES (COMPLAINT (B)(4)). THE DEVICE WAS REPLACED BY A NEW ONE AND, AFTER TWO DAYS OF USE, THE MANIFOLD ALSO BROKE INTO TWO PIECES (COMPLAINT (B)(4)). A THIRD DEVICE WAS USED. THE BLOOD LOSS WAS MINIMAL. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICES WERE NOT AVAILABLE FOR EVALUATION SINCE THEY WERE DISCARDED; HOWEVER, AN IMAGE OF THE DEVICE INVOLVED IN THE COMPLAINT (B)(4) WAS AVAILABLE.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21111872 ↗MalfunctionSep 9, 2024Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED, DURING USE IN PATIENT OF THIS VOLUME VIEW SENSOR, AN UNKNOWN QUANTITY OF MEDICATION (METHYLENE BLUE) LEAKED FROM THE MANIFOLD. NO ACTION REQUIRED DUE TO MEDICATION LOSS. THERE WAS NO ALLEGATION OF PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20167344 ↗
Other PROBE, THERMODILUTION devices
- PiCCO Catheter PV2015L20-APulsion Medical Systems SE · Probe, Thermodilution24 reports
- PiCCO Catheter PV2014L16-APulsion Medical Systems SE · Probe, Thermodilution16 reports
- CO-SET+ 93505Edwards Lifesciences LLC · PROBE, THERMODILUTION9 reports
- PiCCO Catheter PV2014L22-APulsion Medical Systems SE · Probe, Thermodilution8 reports
- CO-SET+ 93522Edwards Lifesciences LLC · PROBE, THERMODILUTION3 reports
- PiCCO Catheter PV2014L16NPulsion Medical Systems SE · Probe, Thermodilution3 reports
- PiCCO Catheter PV2015L20NPulsion Medical Systems SE · Probe, Thermodilution1 report
- CO-SET+ 93600Edwards Lifesciences LLC · PROBE, THERMODILUTION1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.