Skip to content

CO-SET+ 93505Recalls & reported problems

Edwards Lifesciences LLC · CO-SET HOUSING CASE OF 8 · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07460691949973), received between Apr 30, 2024 and Dec 5, 2024 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component5
  • Break3
  • Incorrect Measurement3
  • Appropriate Device Problem Term/Code Not Available2
  • Fluid/Blood Leak2

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Hemorrhage/Blood Loss/Bleeding2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, WHEN MONITORING THE PATIENT AS PART OF POSTOPERATIVE CARE, THIS CO-SET WAS CRACKED CAUSING BLOOD LEAKAGE SPILLED ON TO THE BED. THERE IS NO FURTHER INFORMATION AVAILABLE. AS PER PRODUCT EVALUATION FINDINGS, THE MANIFOLD OF THIS CO-SET WAS COMPLETELY BROKEN. PATIENT DEMOGRAPHICS WERE NOT AVAILABLE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20853239 ↗
  • MalfunctionSep 27, 2024Detachment of Device or Device Component; Appropriate Device Problem Term/Code Not Available; Detachment of Device or Device Component; Appropriate Device Problem Term/Code Not Available

    AS REPORTED, DURING USE IN PATIENT WITH THIS MANIFOLD, IT BROKE DURING THE MOBILIZATION PROCEDURE AND A BLOOD LEAKAGE WAS PRODUCED. TO SOLVE THE ISSUE THE DEVICE WAS REPLACED. AS PER THE HOSPITAL INFORMATION THE PHARMACEUTICAL SERVICE, CONFIRMED THAT THERE WAS AN EVIDENT LEAK THROUGH THE THERMISTOR CABLE WAS CONNECTED. THE AMOUNT OF BLOOD LOST BY THE PATIENT WAS NOT MEASURED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. IT WAS DISCARDED AS THE DEVICE WAS HEAVILY IMPREGNATED WITH THE PATIENT'S BLOOD.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20320087 ↗
  • MalfunctionMay 22, 2024Incorrect Measurement; Incorrect Measurement

    AS REPORTED, WHEN INJECTING THE BOLUS TO MEASURE CARDIAC OUTPUT, THE CO-SET MANIFOLD WAS FOUND TO HAVE A CRACK AT THE FEMALE LUER AND SALINE LEAKED. THEREFORE, CARDIAC OUTPUT VALUES WERE 8.5L/MIN, WHILE AS PER PATIENT'S CONDITION IT SHOULD BE 6L/MIN. REPLACING THE CO-SET MANIFOLD THE ISSUE WAS SOLVED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19368605 ↗
  • MalfunctionMay 21, 2024Fluid/Blood Leak; Incorrect Measurement; Fluid/Blood Leak; Incorrect Measurement

    AS REPORTED, WHEN INJECTING THE BOLUS TO MEASURE CARDIAC OUTPUT, THE CO-SET MANIFOLD WAS FOUND TO HAVE A CRACK AT THE FEMALE LUER AND SALINE LEAKED. THEREFORE, CARDIAC OUTPUT VALUES WERE 8.5L/MIN, WHILE AS PER PATIENT'S CONDITION IT SHOULD BE 6L/MIN. REPLACING THE CO-SET MANIFOLD THE ISSUE WAS SOLVED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19357254 ↗
  • MalfunctionMay 21, 2024Fluid/Blood Leak; Incorrect Measurement; Fluid/Blood Leak; Incorrect Measurement

    AS REPORTED, WHEN INJECTING THE BOLUS TO MEASURE CARDIAC OUTPUT, THE CO-SET MANIFOLD WAS FOUND TO HAVE A CRACK AT THE FEMALE LUER AND SALINE LEAKED. THEREFORE, CARDIAC OUTPUT VALUES WERE 8.5L/MIN, WHILE AS PER PATIENT'S CONDITION IT SHOULD BE 6L/MIN. REPLACING THE CO-SET MANIFOLD THE ISSUE WAS SOLVED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19357235 ↗
  • MalfunctionMay 21, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    AS REPORTED, WHEN INJECTING THE BOLUS TO MEASURE CARDIAC OUTPUT, THE CO-SET MANIFOLD WAS FOUND TO HAVE A CRACK AT THE FEMALE LUER AND SALINE LEAKED. THEREFORE, CARDIAC OUTPUT VALUES WERE 8.5L/MIN, WHILE AS PER PATIENT'S CONDITION IT SHOULD BE 6L/MIN. REPLACING THE CO-SET MANIFOLD THE ISSUE WAS SOLVED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19357074 ↗
  • MalfunctionApr 30, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    AS REPORTED, DURING USE IN PATIENT, THE MANIFOLD OF THIS CO-SET BROKE. THERE IS NO ALLEGATION OF PATIENT INJURY. PATIENT DEMOGRAPHICS WERE NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19213290 ↗
  • MalfunctionApr 30, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    AS REPORTED, IN SEVERAL OCCASION DURING USE IN PATIENT, THIS MANIFOLD BROKE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICES WERE NOT AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19213004 ↗
  • MalfunctionApr 30, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    AS REPORTED, DURING USE, SEVERAL TIMES IN THE PAST, THE MANIFOLD OF THIS CO-SET BROKE. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS NOT AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS WERE NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19212722 ↗

Other PROBE, THERMODILUTION devices

All PROBE, THERMODILUTION devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.