LIFEPAK® 500 Automated External Defibrillator 99401 (REF 99401-000210)Recalls & reported problems
Physio-Control, Inc. · LP500AAAAAZZZZZLP500EECGIAAAABAZZZZZ · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00721902172557), received between Mar 1, 2019 and May 7, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Sensing Problem1
- Electrical /Electronic Property Problem1
Patient problems
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 7, 2019Device Sensing Problem; Device Sensing Problem
THE CUSTOMER, A HEALTHCARE PROVIDER, CONTACTED PHYSIO-CONTROL TO REPORT THAT DURING A PATIENT EVENT THEIR DEVICE WAS UNABLE TO DETECT THE PATIENT THROUGH THE DEFIBRILLATION ELECTRODES. THE CUSTOMER ADVISED THAT THE DEVICE PROMPTED THEM TO ¿CONNECT ELECTRODES¿ WHEN THE DEFIBRILLATION ELECTRODES WERE CONNECTED TO THE PATIENT. THE SAME DEFIBRILLATION ELECTRODES WERE CONNECTED TO BACKUP AED THAT WAS ON THE SCENE, AND THE THAT DEVICE WORKED PROPERLY. IT IS UNKNOWN IF SKIN PREP WAS PERFORMED PRIOR TO APPLYING THE DEFIBRILLATION ELECTRODES. THE CUSTOMER ALSO ADVISED THAT THE DEFIBRILLATION ELECTRODES WERE NOT EXPIRED. THE PATIENT, A (B)(6) YEAR OLD MALE DID NOT SURVIVE. THE CUSTOMER, A HEALTH CARE PROVIDER ADVISED THAT THE DEVICE USE DID NOT CONTRIBUTE TO THE PATIENT OUTCOME, BUT WAS DUE TO THE PATIENT'S CONDITION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8589236 ↗MalfunctionMar 1, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE CUSTOMER CONTACTED PHYSIO-CONTROL TO REPORT THAT THE SERVICE INDICATOR ON THEIR DEVICE IS FLASHING AND A CALL SERVICE MESSAGE IS DISPLAYED. THE DEVICE ALSO LOGGED AN EVENT CODE THAT IS RELATED TO A POTENTIAL FAILURE OF THE DEVICE TO DELIVER DEFIBRILLATION ENERGY. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED EVENT.
Patient problems: No Patient Involvement
MAUDE report 8382920 ↗
Other Automated external defibrillators (non-wearable) devices
- NA 861290Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)12 deaths5,167 reports
- NA 861288Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)21 deaths5,015 reports
- NA 861289Philips Medical Systems Hsg · Automated External Defibrillators (Non-Wearable)7 deaths3,385 reports
- R SERIES 30001005201310012Zoll Medical Corporation · Automated external defibrillators (non-wearable)10 deaths1,720 reports
- PHILIPS HeartStart FRxPhilips Electronics North America Corporation · Automated external defibrillators (non-wearable)26 deaths1,586 reports
- LIFEPAK CR2 99512Physio-Control, Inc. · Automated external defibrillators (non-wearable)8 deaths1,502 reports
- Lifeline Configuration DDU-100ADefibtech, L.L.C. · Automated External Defibrillators (Non-Wearable)1 death1,285 reports
- PHILIPS HeartStart FR3Philips Electronics North America Corporation · Automated External Defibrillators (Non-Wearable)10 deaths1,188 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page