PRESTIGE® Cervical Disc System 6961470Recalls & reported problems
Medtronic Sofamor Danek, Inc. · DISC 6961470 PRESTIGE ARTIFICIAL 7X14MM · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00721902723254), received between Mar 25, 2019 and Mar 25, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
6 other recalls cover devices cleared under the same FDA submission (P060018); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Cerebrospinal Fluid Leakage1
- Device Embedded In Tissue or Plaque1
- Host-Tissue Reaction1
- Weakness1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMar 25, 2019Break; Break
PRE-OPERATIVE DIAGNOSIS: SUPERFICIAL SIDEROSIS LEVEL IMPLANTED: C3-C4 IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR CERVICAL DISCECTOMY AND ARTHROPLASTY SURGERY. POST-OP, ONE SCREW OF THE DISC BROKE AND DISTAL TIP WAS NOT RETRIEVED. REMOVAL OF DEVICE AND FUSION OF INDEX LEVEL WAS A REVISION SURGERY THAT WAS PERFORMED. SURROUNDING PRODUCT TISSUE PRIOR AND AFTER TO EXPLANATION WAS BLACKENED. BROKEN FRAGMENT REMAINS IN THE PATIENT. GENERAL INCREASING WEAKNESS AND IMAGING CONFIRMED DIAGNOSIS OF SIDEROSIS AS A PATIENT COMPLICATION. PATIENT SYMPTOMS MAY BE CAUSED BY LEAKAGE OF CHRONIC CSF LEAK DURING THE PROCEDURE.
Patient problems: Cerebrospinal Fluid Leakage; Device Embedded In Tissue or Plaque; Host-Tissue Reaction; Weakness
MAUDE report 8450276 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P060018. One submission often covers many models, so check the product and lot details in each record.
- Z-0138-2008 · Sep 4, 2007 · TerminatedMedtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #: 6972226, SIZE 6 X 12, MATL: Stainless Steel/Titanium, Rx only, Manufactured at: Bartlett, Te
- Z-0140-2008 · Sep 4, 2007 · TerminatedMedtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #: 6972266, SIZE 6 X 16, MATL: Stainless Steel/Titanium, Rx only, Manufactured at: Bartlett, Te
- Z-0142-2008 · Sep 4, 2007 · TerminatedMedtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog: 6972267, SIZE 7 X 16, MATL: Stainless Steel/Titanium, Rx only, Manufactured at: Bartlett, Ten
- Z-0143-2008 · Sep 4, 2007 · TerminatedMedtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #: 6972287, SIZE 7 X 18, MATL: Stainless Steel/Titanium, Rx only, Manufactured at: Bartlett, T
- Z-0139-2008 · Sep 4, 2007 · TerminatedMedtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #: 6972246, SIZE 6 X 14, MATL: Stainless Steel/Titanium, Rx only, Manufactured at: Bartlett, T
- Z-0141-2008 · Sep 4, 2007 · TerminatedMedtronic Sofamor Danek, Prestige Cervical Disc System, Implant Trial, REF/Catalog #: 6972247, SIZE 7 X 14, MATL: Stainless Steel/Titanium, Rx only, Manufactured at: Bartlett, T
Other PROSTHESIS, INTERVERTEBRAL DISC devices
- Prodisc C SK PDSM5Centinel Spine, LLC · Prosthesis, Intervertebral Disc2 recalls4 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall12 reports
- Prodisc C 09.820.055SCentinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall9 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall4 reports
- Prodisc C SK PDSXL5Centinel Spine, LLC · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
- Simplify® Cervical Artificial DiscNuvasive, Inc. · Prosthesis, Intervertebral Disc1 recall3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page