Skip to content

AMSCO 7053HP (REF FH19043)Recalls & reported problems

Corporation Steris Canada · The AMSCO 7053HP Washer-Disinfector with Electric Heat delivers a validated and cleared process for cleaning reusable da Vinci EndoWrist X/Xi, S/Si, and Single-Site instruments and staplers. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00724995176679), received between May 14, 2025 and Jul 30, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 19, 2025Open, Classified

    Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

    Z-0910-2026 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Leak/Splash1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 30, 2026Detachment of Device or Device Component

    The user facility reported while an employee was cleaning their AMSCO 7053HP Washer/Disinfector the panel that cover's the facility's plumbing/water inlet adjacent to the unit detached and "struck" the employee on the arm. It is unknown if medical treatment was sought or administered.

    Patient problems: Insufficient Information

    MAUDE report 50082608 ↗
  • MalfunctionMay 14, 2025Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THEIR AMSCO 7053HP WASHER WAS LEAKING WATER ONTO THE FLOOR. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22021695 ↗

Other Disinfector, medical devices devices

All Disinfector, medical devices devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.