AMSCO 7053HP (REF FH19043)Recalls & reported problems

Corporation Steris Canada · The AMSCO 7053HP Washer-Disinfector with Electric Heat delivers a validated and cleared process for cleaning reusable da Vinci EndoWrist X/Xi, S/Si, and Single-Site instruments and staplers. · Class 2

All models of AMSCO (25)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00724995176679), received between May 14, 2025 and Jul 30, 2026 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 19, 2025Class IIOpen, Classified

    Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).

    Z-0910-2026 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Leak/Splash1

Patient problems

  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 30, 2026Detachment of Device or Device Component

    The user facility reported while an employee was cleaning their AMSCO 7053HP Washer/Disinfector the panel that cover's the facility's plumbing/water inlet adjacent to the unit detached and "struck" the employee on the arm. It is unknown if medical treatment was sought or administered.

    Patient problems: Insufficient Information

    MAUDE report 50082608 ↗
  • MalfunctionMay 14, 2025Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THEIR AMSCO 7053HP WASHER WAS LEAKING WATER ONTO THE FLOOR. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22021695 ↗

FDA clearance and classification

The FDA regulates product code MEC (disinfector, medical devices) under 21 CFR 880.6992 as a Class II device. 21 CFR 880.6992 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Common questions about the AMSCO 7053HP

Has the AMSCO 7053HP been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0910-2026 (Class II), began on Nov 19, 2025: Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users). If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the AMSCO 7053HP?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between May 14, 2025 and Jul 30, 2026. In those reports, the most common specific device problems are “Detachment of Device or Device Component” (1) and “Leak/Splash” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AMSCO 7053HP?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AMSCO 7053HP FDA approved?

Devices of this type (product code MEC) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Corporation Steris Canada.

Who makes the AMSCO 7053HP, and is it still sold?

The FDA's device database lists Corporation Steris Canada as the company responsible for it (brand AMSCO, model 7053HP). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Disinfector, medical devices devices

All Disinfector, medical devices devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page