Skip to content

HarmonyAIR A-Series Lighting System (REF LAH11IN02)Recalls & reported problems

Steris Corporation · The A-Series High Density Single Light, Dual Monitor Mount 51-inch Dual Monitor Mount, 43-inch Light, 20-inch Drop Tube is used with the HarmonyAIR A-Series Lighting System. · Class 2

All models of HarmonyAIR (240)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00724995187637), received between Oct 21, 2022 and Jul 16, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component2
  • Power Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THEIR DFPM YOKE ASSEMBLY TO THEIR HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM DETACHED FROM THE SPRING ARM AND WAS HANGING FROM THE WIRES WITHIN THE YOKE ASSEMBLY. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22523334 ↗
  • MalfunctionJul 2, 2025Power Problem; Power Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE WALL CONTROL TO THEIR HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM DISPLAYED AN ERROR CODE AND THE LIGHT HEAD WAS FLICKERING RESULTING IN A PROCEDURE DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22385984 ↗
  • MalfunctionOct 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THEIR DUAL MONITOR YOKE ASSEMBLY DETACHED FROM THE SUSPENSION ARM OF THEIR HARMONYAIR A-SERIES SURGICAL LIGHTING SYSTEM AND WAS HANGING BY THE WIRING WITHIN THE LIGHT. THERE WAS NO ONE IN THE ROOM AT THE TIME OF THE REPORTED EVENT. NO REPORT OF INJURY OR PROCEDURAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15651222 ↗

Other Light, surgical, ceiling mounted devices

All Light, surgical, ceiling mounted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page