Skip to content

HarmonyAIR A-Series Lighting System (REF LAS24NC02)Recalls & reported problems

Steris Corporation · The A-Series Single Light, Single Monitor Mount 51-inch Single Monitor Mount, 43-inch Light, 20-inch Drop Tube, No Controller is used with the HarmonyAIR A-Series Lighting System. · Class 2

All models of HarmonyAIR (240)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00724995188269), received between May 7, 2026 and May 7, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 15, 2026Open, Classified

    There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.

    Z-2448-2026 · Root cause: Process change control · Recall details · FDA record ↗

  2. May 15, 2026Open, Classified

    The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.

    Z-2449-2026 · Root cause: Material/Component Contamination · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 7, 2026Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT MULTIPLE LIGHTHEADS TO THEIR HARMONYAIR A-SERIES LIGHTING SYSTEMS ARE "CRACKING" SUBSEQUENTLY CAUSING "PIECES" TO FALL ONTO SURGICAL TABLES. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25116009 ↗

Other Light, surgical, ceiling mounted devices

All Light, surgical, ceiling mounted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page