AMSCO 5052 (REF FH14043)Recalls & reported problems
Corporation Steris Canada · The AMSCO 5052 Single-Chamber Washer-Disinfector, double-door, electric unit is designed for use in the cleaning and intermediate-level disinfection of soiled reusable utensils, trays, bedpans and urinals. It can also be used for rubber and plastic goods, simple hard-surfaced rigid surgical instruments (such as forceps and clamps) and other similar and related articles found in healthcare facilities. · Class 2
All models of AMSCO (25)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00724995192969), received between May 26, 2022 and Jul 9, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Device Fell” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Fell1
- Difficult to Open or Close1
- Improper or Incorrect Procedure or Method1
- Insufficient Device Problem Information1
- Mechanical Jam1
- Sharp Edges1
Patient problems
- Laceration(s)3
- Bone Fracture(s)1
- Bruise/Contusion1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 9, 2026Improper or Incorrect Procedure or Method
THE USER FACILITY REPORTED THAT AN EMPLOYEE OBSERVED A BASKET HAD SHIFTED WHILE THE UNLOAD DOOR TO THEIR AMSCO 5052 WASHER/DISINFECTOR WAS CLOSING. THE EMPLOYEE MANUALLY REPOSITIONED THE BASKET BY PUSHING IT BACK TOWARD THE LOAD SIDE AND THE DOOR CONTINUED TO LOWER SUBSEQUENTLY CONTACTING THE EMPLOYEE'S HAND, CAUSING LACERATIONS TO THEIR FINGERS. MEDICAL TREATMENT WAS SOUGHT AND ADMINISTERED AT THE USER FACILITY'S EMERGENCY DEPARTMENT.
Patient problems: Laceration(s)
MAUDE report 25771011 ↗MalfunctionFeb 27, 2026Sharp Edges
THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A CUT ON THEIR FINGER FROM A SHARP EDGE ON THEIR AMSCO 5052 SINGLE-CHAMBER WASHER/DISINFECTOR. MEDICAL TREATMENT WAS SOUGHT AND ADMINISTERED (MEDICAL GRADE GLUE).
Patient problems: Laceration(s)
MAUDE report 24475470 ↗MalfunctionJan 31, 2025Difficult to Open or Close; Difficult to Open or Close
THE USER FACILITY REPORTED THAT AN EMPLOYEE WAS ATTEMPTING TO CLEAR AN OBSTRUCTED RACK FROM THE DOOR OF THEIR AMSCO 5052 WASHER/DISINFECTOR, WHEN THE DOOR CAME DOWN AND CONTACTED AN EMPLOYEE'S HAND. THE EMPLOYEE RECEIVED STITCHES AS A RESULT.
Patient problems: Laceration(s)
MAUDE report 21285343 ↗MalfunctionOct 29, 2024Device Fell; Device Fell
THE USER FACILITY REPORTED THAT AN EMPLOYEE EXPERIENCED A BRUISE ON HER ARMS AND A CONCUSSION AFTER THE PANEL OF THEIR AMSCO 5052 SINGLE-CHAMBER WASHER-DISINFECTOR FELL. THE EMPLOYEE WAS GIVEN ZOFRAN.
Patient problems: Bruise/Contusion
MAUDE report 20556337 ↗MalfunctionMay 26, 2022Mechanical Jam; Insufficient Device Problem Information; Mechanical Jam; Insufficient Device Problem Information
THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED AN INJURY TO THEIR HAND WHILE OPERATING THEIR AMSCO 5052 WASHER. MEDICAL TREATMENT WAS SOUGHT AND ADMINISTERED.
Patient problems: Bone Fracture(s)
MAUDE report 14518830 ↗
FDA clearance and classification
The FDA regulates product code MEC (disinfector, medical devices) under 21 CFR 880.6992 as a Class II device. 21 CFR 880.6992 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Common questions about the AMSCO 5052
Has the AMSCO 5052 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00724995192969), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the AMSCO 5052?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between May 26, 2022 and Jul 9, 2026. In those reports, the most common specific device problems are “Device Fell” (1), “Difficult to Open or Close” (1) and “Improper or Incorrect Procedure or Method” (1); the most common patient problems are “Laceration(s)” (3), “Bone Fracture(s)” (1) and “Bruise/Contusion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the AMSCO 5052?
None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the AMSCO 5052 FDA approved?
Devices of this type (product code MEC) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Corporation Steris Canada.
Who makes the AMSCO 5052, and is it still sold?
The FDA's device database lists Corporation Steris Canada as the company responsible for it (brand AMSCO, model 5052). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Disinfector, medical devices devices
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls6 reports
- GETINGE 88-SERIES 88-5Getinge Disinfection AB · Disinfector, Medical Devices3 recalls0 reports
- GETINGE CM320-SERIESGetinge Disinfection AB · Disinfector, Medical Devices2 recalls4 reports
- GETINGE CM320-SERIESGetinge Disinfection AB · Disinfector, Medical Devices2 recalls1 report
- GETINGE 86-SERIES S-8668Getinge Disinfection AB · Disinfector, Medical Devices1 recall36 reports
- Getinge Aquadis 56 56MGetinge Disinfection AB · Disinfector, Medical Devices1 recall11 reports
- AMSCO 7053HPCorporation Steris Canada · Disinfector, medical devices1 recall10 reports
- GETINGE 9100-SERIES 9128EGetinge Disinfection AB · Disinfector, Medical Devices1 recall10 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page