Uretero1 URORD101Recalls & reported problems
Black Diamond Video, Inc. · The Uretero1 Reverse Deflection Disposable Ureteroscope is a single use flexible ureteroscope that is used to visualize and perform various diagnostic and therapeutic procedures in the urinary tract. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00724995209179), received between Mar 15, 2024 and Jan 15, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Poor Quality Image3
- Erratic or Intermittent Display2
- Crack1
- Defective Device1
- Difficult to Advance1
- Display or Visual Feedback Problem1
- Insufficient Device Problem Information1
- Signal Artifact/Noise1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2026Display or Visual Feedback Problem
THE USER FACILITY REPORTED THAT THE IMAGE TO THEIR URETERO1 REVERSE DEFLECTION DISPOSABLE URETEROSCOPE FROZE DURING A PATIENT PROCEDURE CREATING A PROCEDURE DELAY. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24086685 ↗MalfunctionJan 5, 2026Insufficient Device Problem Information
THE USER FACILITY REPORTED THAT THEIR URETERO1 REVERSE DEFLECTION DISPOSABLE URETEROSCOPE WAS NOT WORKING AND COULD NOT BE USED DURING A PATIENT PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER SCOPE. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23969702 ↗MalfunctionJan 5, 2026Difficult to Advance
THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THE DISTAL END OF THEIR URETERO1 REVERSE DEFLECTION DISPOSABLE URETEROSCOPE BECAME OBSTRUCTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING A DIFFERENT SCOPE. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23968918 ↗MalfunctionAug 15, 2025Poor Quality Image; Poor Quality Image
THE USER FACILITY REPORTED VIA MEDWATCH REPORT (B)(4) THAT "DOCTOR WAS USING THE URETERO1FLEXIBLE URETEROSCOPE AND HAD BEEN WORKED FINE. BUT SUDDENLY, THE IMAGE WAS LOST AND THE SCREEN SAID, "SCOPE NOT DETECTED." WE CHANGED TO ANOTHER MONITOR, BUT SAME MESSAGE SHOWED. THE PROCEDURE WAS ALMOST FINISHED AT THAT TIME; HE DIDN'T NEED ANOTHER URETERO1. THE PATIENT HAS NO INJURY."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22819610 ↗MalfunctionAug 7, 2025Poor Quality Image; Poor Quality Image
THE USER FACILITY REPORTED VIA MEDWATCH # (B)(4) THAT, "URETERO1, DISPOSABLE FLEXIBLE URETEROSCOPE IMAGE GETS CUT OFF, INTERMITTENTLY LOST AND KEEPS CUTTING OUT OR FLICKERING. OR TEAM FOUND A CRACK AT THE BASE OF THE SCOPE AT THE END OF THE PROCEDURE. THE SALES REPRESENTATIVE FOR THE PRODUCT WAS CONTACTED AND THE PLAN WILL BE TO RETURN THE SCOPE FOR INSPECTION."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22743499 ↗MalfunctionJul 25, 2025Poor Quality Image; Poor Quality Image
THE USER FACILITY REPORTED VIA MEDWATCH # (B)(4) THAT, "URETERO1, DISPOSABLE FLEXIBLE URETEROSCOPE IMAGE GETS CUT OFF, INTERMITTENTLY LOST AND KEEPS CUTTING OUT OR FLICKERING. OR TEAM FOUND A CRACK AT THE BASE OF THE SCOPE AT THE END OF THE PROCEDURE. THE SALES REPRESENTATIVE FOR THE PRODUCT WAS CONTACTED AND THE PLAN WILL BE TO RETURN THE SCOPE FOR INSPECTION."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22612789 ↗MalfunctionJul 17, 2025Erratic or Intermittent Display; Defective Device
DEFECTIVE URETERO1 FLEXIBLE URETEROSCOPE. DOCTOR WAS USING THE URETERO1FLEXIBLE URETEROSCOPE AND HAD BEEN WORKED FINE. BUT SUDDENLY, THE IMAGE WAS LOST AND THE SCREEN SAID, "SCOPE NOT DETECTED." WE CHANGED TO ANOTHER MONITOR, BUT SAME MESSAGE SHOWED. THE PROCEDURE WAS ALMOST FINISHED AT THAT TIME; HE DIDN'T NEED ANOTHER URETERO1. THE PATIENT HAS NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22530184 ↗MalfunctionJun 30, 2025Crack; Erratic or Intermittent Display
URETERO1, DISPOSABLE FLEXIBLE URETEROSCOPE IMAGE GETS CUT OFF, INTERMITTENTLY LOST AND KEEPS CUTTING OUT OR FLICKERING. OPERATING ROOM TEAM FOUND A CRACK AT THE BASE OF THE SCOPE AT THE END OF THE PROCEDURE. THE SALES REPRESENTATIVE FOR THE PRODUCT WAS CONTACTED AND THE PLAN WILL BE TO RETURN THE SCOPE FOR INSPECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22352625 ↗MalfunctionMar 15, 2024Signal Artifact/Noise; Signal Artifact/Noise
THE USER FACILITY REPORTED THAT DURING A PROCEDURE THERE WAS INTERMITTENT VIDEO INTERFERENCE TO THEIR URETERO1 REVERSE DEFLECTION DISPOSABLE URETEROSCOPE. THERE WAS A DELAY IN THE PROCEDURE TO SWITCH TO A DIFFERENT SCOPE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18918441 ↗
Other Ureteroscope and accessories, flexible/rigid devices
- ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-10Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall1,125 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls397 reports
- URETERO-RENO VIDEOSCOPE OLYMPUS URF-V2ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall305 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls190 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P6Olympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid2 recalls170 reports
- ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3DOlympus Medical Systems Corp. · Ureteroscope And Accessories, Flexible/Rigid186 reports
- URETERO-RENO FIBERSCOPE OLYMPUS URF-P7ROlympus Medical Systems Corp. · Ureteroscope and accessories, flexible/rigid1 recall121 reports
- n.a. 091279-01Karl Storz GmbH & Co. KG · Ureteroscope and accessories, flexible/rigid118 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page