Skip to content

Welch Allyn, Inc. 901060 (REF 68MXTX-B)Recalls & reported problems

Welch Allyn, Inc. · Welch Allyn Connex 6800 Vital Signs Monitor with Masimo SpO2, SureTemp Plus Thermometry, SureBP Non-invasive Blood Pressure, Pulse Rate, MAP, BP Averaging, Nurse Call, Four USB Ports for Accessories, Internal 802.11 a/b/g Wireless Radio; 100-240 V, 50-60 Hz AC, IEC Plug Type-B · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00732094150933), received between Dec 16, 2020 and Jan 23, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fire” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2025Class IIOpen, Classified

    There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

    Z-1558-2025 · Root cause: Process design · Recall details · FDA record ↗

  2. Mar 21, 2025Class IIOpen, Classified

    Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.

    Z-1564-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fire2
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2025Fire; Fire

    THE CUSTOMER REPORTED A CVSM CAUGHT FIRE IN THE PHYSICAL THERAPY ROOM. THERE WAS NO ADVERSE EVENT REPORTED. THIS INCIDENT WAS CAPTURED UNDER COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21216522 ↗
  • MalfunctionJul 24, 2023Overheating of Device; Overheating of Device

    THE CUSTOMER REPORTED THE POWER CORD WAS BAD. THIS EVENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17377968 ↗
  • MalfunctionDec 16, 2020Fire; Fire

    WELCH ALLYN RECEIVED A REPORT FROM THE ACCOUNT STATING THE CVSM DEVICE CAUGHT FIRE AFTER PLUGGING IN THE DEVICE AT SET UP. THE DEVICE WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 11018816 ↗

Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page