Skip to content

Welch Allyn, Inc. 901058 (REF 75MT-B)Recalls & reported problems

Welch Allyn, Inc. · Welch Allyn Connex Spot Monitor with SureBP Non-invasive Blood Pressure, Masimo SpO2, SureTemp Plus Thermometer · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00732094209372), received between Apr 29, 2022 and Feb 26, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Fire” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 21, 2025Class IIOpen, Classified

    Undeclared natural rubber band within the packaging of certain configurations of reusable blood pressure cuffs.

    Z-1563-2025 · Root cause: Device Design · Recall details · FDA record ↗

  2. Aug 10, 2023Class IIOpen, Classified

    Product is being recalled due to the improper placement of a copper tape on the back of the liquid-crystal display.

    Z-2581-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fire2
  • Material Split, Cut or Torn1
  • Overheating of Device1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 26, 2026Material Split, Cut or Torn

    THE CUSTOMER REPORTED THE POWER CORD WAS FRAYED. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24464769 ↗
  • MalfunctionApr 18, 2024Fire; Fire

    THE CUSTOMER REPORTED THAT THE CSM POWER CORD CAUGHT FIRE AND HAD CHARRING AT THE PLUG. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19137863 ↗
  • MalfunctionJun 20, 2023Fire; Fire

    THE CUSTOMER REPORTED THE POWER SUPPLY CAUGHT FIRE. THERE WERE NO ALLEGATIONS OF INJURY. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17161971 ↗
  • MalfunctionSep 8, 2022Smoking; Smoking

    THE CUSTOMER REPORTED THAT THE POWER CORD "SMOKED" WHEN THE CSM UNIT WAS PLUGGED INTO THE POWER OUTLET AND THE UNIT WOULD NOT TAKE A BP READING. THIS EVENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15378058 ↗
  • MalfunctionApr 29, 2022Overheating of Device; Overheating of Device

    THE CUSTOMER REPORTED THAT THE BACK HOUSING WAS CHARRED. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14252092 ↗

Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page