Skip to content

Rad-97 9738Recalls & reported problems

Masimo Corporation · Rad-97 · Class 2

Quick summary

This Masimo Rad-97 patient monitor has 224 adverse event reports, 222 of which describe malfunctions. No injuries are recorded, and 1 report lists a death. Many reports describe the monitor shutting down or resetting on its own, sensing readings intermittently, or not sounding an alarm when the battery is low. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
222
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 222 adverse event reports that list this exact device (DI 00843997012317), received between Nov 27, 2017 and Sep 25, 2026 — classified as 1 death, 221 malfunctions.

The most frequently reported specific device problem is “Unexpected Shutdown” (67 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 3 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 5 reportsJun 2024: 4 reportsJul 2024: 3 reportsAug 2024: 3 reportsSep 2024: 5 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 3 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 5 reportsJul 2025: 8 reportsAug 2025: 3 reportsSep 2025: 2 reportsOct 2025: 3 reportsNov 2025: 3 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 5 reportsSep 2026: 14 reportsOct 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unexpected Shutdown67
  • Sensing Intermittently43
  • Inappropriate or Unexpected Reset33
  • No Audible Alarm23
  • Incorrect Measurement21
  • Defective Alarm11
  • Device Alarm System11
  • Failure to Power Up6

Patient problems

  • No Clinical Signs, Symptoms or Conditions178
  • No Consequences Or Impact To Patient42
  • Brain Injury1
  • Death1

Recent adverse event reports

The 25 most recent of 222 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 25, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50902565 ↗
  • MalfunctionSep 19, 2026Defective Alarm; Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50832416 ↗
  • MalfunctionSep 19, 2026Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50832409 ↗
  • MalfunctionSep 15, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50766050 ↗
  • MalfunctionSep 7, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50645066 ↗
  • MalfunctionSep 7, 2026Defective Alarm; Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50645063 ↗
  • MalfunctionSep 4, 2026Incorrect Measurement; Sensing Intermittently

    The customer reported Intermittently cutting out with replace sensor on a known working setup. Found extra sensitive to interference and giving inaccurate SPO2% readings. There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50604258 ↗
  • MalfunctionSep 3, 2026Defective Alarm; Unexpected Shutdown

    The customer reported that the device is turning off and on without any error message or alarm. There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50578605 ↗
  • MalfunctionSep 3, 2026Sensing Intermittently

    The customer reported: "intermittent readings". There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50578538 ↗
  • MalfunctionSep 1, 2026Inappropriate or Unexpected Reset

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50537145 ↗
  • MalfunctionSep 1, 2026Incorrect Measurement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50537143 ↗
  • MalfunctionSep 1, 2026Defective Alarm; Unexpected Shutdown

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50536089 ↗
  • MalfunctionSep 1, 2026Unexpected Shutdown

    The customer reported that the device is turning on and off with no error messages or alarms. There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50535909 ↗
  • MalfunctionSep 1, 2026Defective Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50535628 ↗
  • MalfunctionAug 25, 2026Unexpected Shutdown

    The customer reported that "The power turns off by itself". There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50422927 ↗
  • MalfunctionAug 23, 2026Device Alarm System; No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50393444 ↗
  • MalfunctionAug 23, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50393440 ↗
  • MalfunctionAug 7, 2026Sensing Intermittently

    The customer reported "Returned from a patient because the connection between the RAD and the sensor is faulty. The device regularly loses data. Test performed with the M20 Tester; same problem - the RAD has great difficulty receiving data." There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50170081 ↗
  • MalfunctionAug 7, 2026Unexpected Shutdown

    The customer reported the device shuts down without any warning. There was no patient impact or consequence reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50170050 ↗
  • DeathJan 12, 2026Accessory Incompatible; Failure to Sense

    THE CUSTOMER REPORTED "ON (B)(6) 2025, PROMPT CARE WAS NOTIFIED THAT A PATIENT USING A SUNSET HEALTHCARE SOLUTIONS SPO2 SENSOR WITH THE RAD 97 MASIMO DIED AFTER GOING BRAIN DEAD BECAUSE THE SYSTEM DID NOT ALARM." BASED ON THE MEDICAL DEVICE PROBLEM CODES, IT WAS REPORTED THE ACCESSORY WAS INCOMPATIBLE AND A FAILURE TO SENSE.

    Patient problems: Brain Injury

    MAUDE report 24031600 ↗
  • MalfunctionDec 1, 2025Inappropriate or Unexpected Reset; Inappropriate or Unexpected Reset

    THE CUSTOMER REPORTED THAT THE DEVICE KEEPS TURNING ITSELF OFF AND ON AGAIN. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23689170 ↗
  • MalfunctionNov 11, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THAT THE UNIT POWERS OFF BY ITSELF AND IT BECOMES UNUSABLE. THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23529208 ↗
  • MalfunctionNov 7, 2025Unexpected Shutdown; Unexpected Shutdown

    THE CUSTOMER REPORTED THE RAD-97 "SHUT ITSELF OFF SHORTLY AFTER BEGINNING MONITORING." PER ADDITIONAL INFORMATION OBTAINED VIA CONVERSATION WITH THE CUSTOMER "THE DEVICE TURNS OFF BY ITSELF RANDOMLY. SOMETIMES RIGHT AWAY; SOMETIMES NOT." THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23503887 ↗
  • MalfunctionNov 6, 2025Device Alarm System; Device Alarm System

    THE CUSTOMER REPORTED "NO ALARM". THERE WAS NO PATIENT IMPACT OR CONSEQUENCE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23493843 ↗
  • MalfunctionOct 31, 2025Sensing Intermittently; Sensing Intermittently

    THE CUSTOMER REPORTED THE RAD-97 "CUTS OFF." PER ADDITIONAL INFORMATION FROM THE CUSTOMER, "THE DEVICE STOPPED PROVIDING READINGS DURING TESTING." THERE WAS NO PATIENT IMPACT OR CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23444160 ↗

Other Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) devices

All Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.