Skip to content

Welch Allyn, Inc. 901060 (REF 68NSXX-B)Recalls & reported problems

Welch Allyn, Inc. · Welch Allyn Connex 6800 Vital Signs Monitor with Nellcor SpO2, EarlySense Contact-Free Monitoring (Includes One Contact-Free Sensor #720011), SureBP Non-invasive Blood Pressure, Pulse Rate, MAP, BP Averaging, Nurse Call, Four USB Ports for Accessories, Internal 802.11 a/b/g Wireless Radio; 100-240 V, 50-60 Hz AC, IEC Plug Type-B · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Mar 24, 2025Class IIOpen, Classified

    There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

    Z-1558-2025 · Root cause: Process design · Recall details · FDA record ↗

Other MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices

All MONITOR,PHYSIOLOGICAL,PATIENT(WITHOUT ARRHYTHMIA DETECTION OR ALARMS) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page