Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00733132615896), received between Jun 17, 2021 and Jun 17, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Obstruction of Flow” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Obstruction of Flow1
  • Partial Blockage1

Patient problems

  • Obstruction/Occlusion1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 17, 2021Partial Blockage; Obstruction of Flow; Partial Blockage; Obstruction of Flow

    THE FOLLOWING WAS REPORTED TO GORE: ON MARCH 17, 2021, THE PATIENT UNDERWENT A PROCEDURE TO REPAIR AN ARTERY IN THE TORSO AS PART OF A BLALOCK-TAUSSIG SHUNT (BTT SHUNT) UPSIZING PROCEDURE (ORIGINAL SHUNT PLACED ON (B)(6) 2020) UTILIZING A GORE-TEX® VASCULAR GRAFT CONFIGURED FOR PEDIATRIC SHUNT DEVICE. A THERAPEUTIC HEPARIN DRIP WAS UTILIZED POST-OPERATION. ON (B)(6) 2021, CTA FOR SHUNT EVALUATION WAS PERFORMED DUE TO HIGH OXYGEN REQUIREMENTS AND MARGINAL SATURATIONS. SHUNT WAS PATENT WITHOUT ASSOCIATED NARROWING. THE PATIENT CONTINUED TO HAVE EPISODES OF SIGNIFICANT DESATURATIONS POST-OPERATIVELY. ON (B)(6) 2021, AN ECHOCARDIOGRAM DEMONSTRATED PROXIMAL BTT SHUNT CONTINUOUS FLOW, BUT IT WAS DIFFICULT TO TRACE THE MORE DISTAL SHUNT TO THE PULMONARY ARTERIES. ON (B)(6) 2021 THE PATIENT UNDERWENT A CARDIAC CATHETERIZATION WITH STENT PLACEMENT IN THE SHUNT AFTER FINDING SEVERE NARROWING OF DISTAL BTT SHUNT MEASURING APPROXIMATELY 2.5MM. THERE WAS SIGNIFICANT IMPROVEMENT OF SYSTEMIC ARTERIAL OXYGEN SATURATION AFTER STENT PLACEMENT. ON (B)(6) 2021 THE PATIENT WAS DISCHARGED HOME ON LOVENOX AND ASPIRIN. ON (B)(6) 2021, THE PATIENT WAS RETURNED TO THE HOSPITAL WITH CPR IN PROGRESS AND EXTRA

    Patient problems: Obstruction/Occlusion; Thrombosis/Thrombus

    MAUDE report 12015819 ↗

Other PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices

All PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER devices →